| Literature DB >> 27698664 |
David Friedel1, Rani Modayil1, Stavros Stavropoulos1.
Abstract
Colon capsule endoscopy utilizing PillCam COLON 2 capsule allows for visualization potentially of the entire colon and is currently approved for patients who cannot withstand the rigors of traditional optical colonoscopy (OC) and associated sedation as well as those that had an OC that was incomplete for technical reasons other than a poor preparation. We will then describe the prior experience and current status of colon capsule endoscopy.Entities:
Year: 2016 PMID: 27698664 PMCID: PMC5028851 DOI: 10.1155/2016/9643162
Source DB: PubMed Journal: Gastroenterol Res Pract ISSN: 1687-6121 Impact factor: 2.260
Figure 1PillCam 2 COLON capsule.
PillCam 2 properties.
| Battery life | 10 hours |
| Cameras | 2-one at each end |
| Lenses | 3 |
| Angle view | 172 degrees |
| Esophageal-gastric frame rate | 14/sec |
| Small bowel colon | AFR-4/s(stationary) 32–35/s (motion) |
Figure 2(a) Transverse colon. (b) Ascending colon. (c) Diverticulum. (d) Crohn's disease. (e) Colon polyp. (f) Colon cancer.
PillCam COLON 2 in colon polyp and cancer detection.
| Lead author/ | Results and comments |
|---|---|
| Akyuz/28/2016/[ | Only 28 of 62 subjects in this study focusing on prep had tandem colonoscopy. 6/28 subjects had polyps in 5–10 mm range. Sensitivity, specificity, PPV, and NPV were 100%, 92%, 93%, 100% |
|
| |
| Morgan/50/2016/[ | Of the 50 patients who had tandem capsule and OC 30% and 14% had polyps ≥6 and 10 mm, respectively. For lesions ≥10 mm on OC, capsule sensitivity was 100% and specificity 100%. For polyps ≥6 mm, capsule sensitivity was 93% and specificity 80.0%. The excretion rate was 65%–61% of studies considered adequate cleansing |
|
| |
| Hollerhan/62/2014/[ | 62 subjects with (+) FIT had tandem capsule/colonoscopy. Sensitivity/specificity for polyps >6 and 10 mm, respectively, was 95/65% and 89/96%, respectively. 92% had adequate cleaning with 73% excretion rate |
|
| |
| Hagel/24/2014/[ | Sensitivity/specificity for polyps ≥6 mm and 10 mm was 72/91 and 75/100%, respectively. Adequate cleansing 90%. Excretion rate 71%. Extracolonic GI pathology >50% |
|
| |
| Rex/884/2013/[ | Largest series. Sensitivity/specificity for polyps ≥6 and 10 mm was 88/82 and 92/95%, respectively. AC 80%. ER 91% |
|
| |
| Spada/117/2015/[ | 8 European centers: 109 subjects analyzed. Sensitivity/specificity for polyps ≥6 and 10 mm was 84/64 and 88/95%, respectively. AC 85%. ER 81% |
|
| |
| Eliakim/104/2009/[ | Initial validating series for CCE-2. Data from 98 subjects. Sensitivity/specificity for polyps ≥6 and 10 mm was 89/76 and 88/89%, respectively. AC 78%. ER 81% |