Computerized artifact detection for ventilatory inductance plethysmographic apnea monitors.

K A East1, T D East, V J Mathews, B T Waterfall.   

Abstract

Ventilatory inductive plethysmography allows noninvasive monitoring of patient ventilation. Patient movements unrelated to breathing introduce severe errors in ventilator inductive plethysmographic measurements and restrict its usefulness. The purpose of this research was to develop and test a microprocessor-based real-time digital signal processor that uses an adaptive filter to detect patient movements unrelated to breathing. The adaptive filter processor was tested for retrospective identification of artifacts in 20 male volunteers who performed the following specific movements between epochs of quiet, supine breathing: raising arms and legs (slowly, quickly, once, and several times), sitting up, breathing deeply and rapidly, and rolling from a supine to a lateral decubitus position. Flow was simultaneously measured directly with a pneumotachography attached to a mouthpiece. A multilinear regression was used to continuously calculate the calibration constants that relate the pneumotachographic and ventilatory inductive plethysmographic signals. Ventilatory inductive plethysmographic data were then processed, and results scored. There were a total of 166 movements. The calibration coefficients changed dramatically in 146 (88%) of the 166 movements. These movements would have significant errors on ventilatory inductive plethysmographic flow calculation. The changes lasted for the duration of the movements and returned to baseline within two to three breaths. The changes in the coefficients were five or more times larger than the variability around baseline during quiet, supine breathing. All of the total body movements and changes in breathing patterns were detected accurately. The filter detected 46 of 53 upper body movements, 34 of 36 lower body movements, 38 of 38 total body movements, and 19 of 19 breathing pattern changes where the calibration changed.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1989        PMID: 2769315     DOI: 10.1007/BF01627449

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  20 in total

1.  Measurement of the separate volume changes of rib cage and abdomen during breathing.

Authors:  K Konno; J Mead
Journal:  J Appl Physiol       Date:  1967-03       Impact factor: 3.531

2.  Changes in tidal volume, frequency, and ventilation induced by their measurement.

Authors:  R Gilbert; J H Auchincloss; J Brodsky; W Boden
Journal:  J Appl Physiol       Date:  1972-08       Impact factor: 3.531

3.  Influence of the respiratory route on the resting breathing pattern in humans.

Authors:  D O Rodenstein; C Mercenier; D C Stănescu
Journal:  Am Rev Respir Dis       Date:  1985-01

4.  The respiratory inductive plethysmograph: a new non-invasive monitor of respiration.

Authors:  M A Cohn; A S Rao; M Broudy; S Birch; H Watson; N Atkins; B Davis; F D Stott; M A Sackner
Journal:  Bull Eur Physiopathol Respir       Date:  1982 Jul-Aug

5.  Validation of respiratory inductive plethysmography using different calibration procedures.

Authors:  T S Chadha; H Watson; S Birch; G A Jenouri; A W Schneider; M A Cohn; M A Sackner
Journal:  Am Rev Respir Dis       Date:  1982-06

6.  Measurement of ventilation in children using the respiratory inductive plethysmograph.

Authors:  E Tabachnik; N Muller; B Toye; H Levison
Journal:  J Pediatr       Date:  1981-12       Impact factor: 4.406

7.  Non-invasive measurement of ventilation during exercise using a respiratory inductive plethysmograph. I.

Authors:  J D Sackner; A J Nixon; B Davis; N Atkins; M A Sackner
Journal:  Am Rev Respir Dis       Date:  1980-12

8.  Breathing patterns. 2. Diseased subjects.

Authors:  M J Tobin; T S Chadha; G Jenouri; S J Birch; H B Gazeroglu; M A Sackner
Journal:  Chest       Date:  1983-09       Impact factor: 9.410

9.  Effect of breathing route on ventilation and ventilatory drive.

Authors:  N J Douglas; D P White; J V Weil; C W Zwillich
Journal:  Respir Physiol       Date:  1983-02

10.  Effects of a face mask and pneumotachograph on breathing in sleeping infants.

Authors:  T Dolfin; P Duffty; D Wilkes; S England; H Bryan
Journal:  Am Rev Respir Dis       Date:  1983-12
View more
  4 in total

1.  Rib cage and abdominal piezoelectric film belts to measure ventilatory airflow.

Authors:  B E Pennock
Journal:  J Clin Monit       Date:  1990-10

2.  Evaluation of a parallel implementation of the learning portion of the backward error propagation neural network: experiments in artifact identification.

Authors:  D F Sittig; J A Orr
Journal:  Proc Annu Symp Comput Appl Med Care       Date:  1991

3.  Automated anesthesia artifact analysis: can machines be trained to take out the garbage?

Authors:  Allan F Simpao; Olivia Nelson; Luis M Ahumada
Journal:  J Clin Monit Comput       Date:  2020-09-12       Impact factor: 2.502

4.  Tissue artifact removal from respiratory signals based on empirical mode decomposition.

Authors:  Shaopeng Liu; Robert X Gao; Dinesh John; John Staudenmayer; Patty Freedson
Journal:  Ann Biomed Eng       Date:  2013-01-17       Impact factor: 3.934

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.