EGS_cbct: Simulation of a fan beam CT and RMI phantom for measured HU verification.

Dete van Eeden1, Freek du Plessis2.   

Abstract

INTRODUCTION: A mathematical 3D model of an existing computed tomography (CT) scanner was created and used in the EGSnrc-based BEAMnrc and egs_cbct Monte Carlo codes. Simulated transmission dose profiles of a RMI-465 phantom were analysed to verify Hounsfield numbers against measured data obtained from the CT scanner. METHODS AND MATERIALS: The modelled CT unit is based on the design of a Toshiba Aquilion 16 LB CT scanner. As a first step, BEAMnrc simulated the X-ray tube, filters, and secondary collimation to obtain phase space data of the X-ray beam. A bowtie filter was included to create a more uniform beam intensity and to remove the beam hardening effects. In a second step the Interactive Data Language (IDL) code was used to build an EGSPHANT file that contained the RMI phantom which was used in egs_cbct simulations. After simulation a series of profiles were sampled from the detector model and the Feldkamp-Davis-Kress (FDK) algorithm was used to reconstruct transversal images. The results were tested against measured data obtained from CT scans.
RESULTS: The egs_cbct code can be used for the simulation of a fan beam CT unit. The calculated bowtie filter ensured a uniform flux on the detectors. Good correlation between measured and simulated CT numbers was obtained.
CONCLUSIONS: In principle, Monte Carlo codes such as egs_cbct can model a fan beam CT unit. After reconstruction, the images contained Hounsfield values comparable to measured data.
Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Keywords:  Computed tomography; Monte Carlo; Simulation

Mesh:

Year:  2016        PMID: 27682511     DOI: 10.1016/j.ejmp.2016.09.011

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Set-up error validation with EPID images: Measurements vs Egs_cbct simulation.

Authors:  D van Eeden; F H J O'Reilly; F C P du Plessis
Journal:  Rep Pract Oncol Radiother       Date:  2019-10-21

2.  Multi-Energy Computed Tomography Breast Imaging with Monte Carlo Simulations: Contrast-to-Noise-Based Image Weighting.

Authors:  Déte Van Eeden; Freek C P Du Plessis
Journal:  J Med Phys       Date:  2019 Apr-Jun
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.