Heat Generation During Bone Drilling: A Comparison Between Industrial and Orthopaedic Drill Bits.

Christopher Hein1, Serkan Inceoglu, David Juma, Lee Zuckerman.   

Abstract

OBJECTIVES: Cortical bone drilling for preparation of screw placement is common in multiple surgical fields. The heat generated while drilling may reach thresholds high enough to cause osteonecrosis. This can compromise implant stability. Orthopaedic drill bits are several orders more expensive than their similarly sized, publicly available industrial counterparts. We hypothesize that an industrial bit will generate less heat during drilling, and the bits will not generate more heat after multiple cortical passes.
METHODS: We compared 4 4.0 mm orthopaedic and 1 3.97 mm industrial drill bits. Three types of each bit were drilled into porcine femoral cortices 20 times. The temperature of the bone was measured with thermocouple transducers. The heat generated during the first 5 drill cycles for each bit was compared to the last 5 cycles. These data were analyzed with analysis of covariance.
RESULTS: The industrial drill bit generated the smallest mean increase in temperature (2.8 ± 0.29°C) P < 0.0001. No significant difference was identified comparing the first 5 cortices drilled to the last 5 cortices drilled for each bit. The P-values are as follows: Bosch (P = 0.73), Emerge (P = 0.09), Smith & Nephew (P = 0.08), Stryker (P = 0.086), and Synthes (P = 0.16). The industrial bit generated less heat during drilling than its orthopaedic counterparts. The bits maintained their performance after 20 drill cycles.
CONCLUSIONS: Consideration should be given by manufacturers to design differences that may contribute to a more efficient cutting bit. Further investigation into the reuse of these drill bits may be warranted, as our data suggest their efficiency is maintained after multiple uses.

Entities:  

Mesh:

Year:  2017        PMID: 27682019     DOI: 10.1097/BOT.0000000000000723

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  4 in total

Review 1.  Surgical Drill Bit Design and Thermomechanical Damage in Bone Drilling: A Review.

Authors:  Mohd Faizal Ali Akhbar; Akmal Wani Sulong
Journal:  Ann Biomed Eng       Date:  2020-08-28       Impact factor: 3.934

2.  Biomechanical Evaluation of Temperature Rising and Applied Force in Controlled Cortical Bone Drilling: an Animal in Vitro Study.

Authors:  Mohammad Javad Ein-Afshar; Mostafa Shahrezaee; Mohammad Hossein Shahrezaee; Seyed Reza Sharifzadeh
Journal:  Arch Bone Jt Surg       Date:  2020-09

3.  Comparative Study of Different Drills for Bone Drilling: A Systematic Approach.

Authors:  O Pazarci; Y Torun; A Ozturk; Z Oztemur
Journal:  Malays Orthop J       Date:  2020-07

4.  Influence of Drilling Technique on the Radiographic, Thermographic, and Geomorphometric Effects of Dental Implant Drills and Osteotomy Site Preparations.

Authors:  Lara Fraguas de San José; Filippo Maria Ruggeri; Roberta Rucco; Álvaro Zubizarreta-Macho; Jorge Alonso Pérez-Barquero; Elena Riad Deglow; Sofía Hernández Montero
Journal:  J Clin Med       Date:  2020-11-11       Impact factor: 4.241

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.