| Literature DB >> 27676097 |
Daniel F Andrade1, Edenir Rodrigues Pereira-Filho1.
Abstract
Contaminants (Cd, Cr, and Pb) as well as minor (B, Cu, Mn, Na, and Zn) and major (Ca and Mg) elements were directly determined in solid fertilizer samples using laser-induced breakdown spectroscopy (LIBS). Factorial designs were used to define the most appropriate LIBS parameters and pellet pressure on solid fertilizers. Emission lines for all of the analytes were collected and employed 12 signal normalization modes. The best results were obtained using a laser energy of 75 mJ, a spot size of 50 μm, a pressure of 10 t/in., and a delay of 2.0 μs. Good correlation was obtained between the calibration model's prediction using the proposed LIBS method and the reference values obtained with ICP-OES. The limits of detection (LOD) for the proposed method varied from 2 mg/kg (for Cd) to 1% (for Zn).Entities:
Keywords: chemometrics; fertilizers; laser-induced breakdown spectroscopy (LIBS); solid sample analysis,
Year: 2016 PMID: 27676097 DOI: 10.1021/acs.jafc.6b04028
Source DB: PubMed Journal: J Agric Food Chem ISSN: 0021-8561 Impact factor: 5.279