| Literature DB >> 27669267 |
Dongmin Seo1, Sung-Ho Paek2, Sangwoo Oh3,4, Sungkyu Seo5, Se-Hwan Paek6.
Abstract
The incidence of diabetes is continually increasing, and by 2030, it is expected to have increased by 69% and 20% in underdeveloped and developed countries, respectively. Therefore, glucose sensors are likely to remain in high demand in medical device markets. For the current study, we developed a needle-type bio-layer interference (BLI) sensor that can continuously monitor glucose levels. Using dialysis procedures, we were able to obtain hypoglycemic samples from commercial human serum. These dialysis-derived samples, alongside samples of normal human serum were used to evaluate the utility of the sensor for the detection of the clinical interest range of glucose concentrations (70-200 mg/dL), revealing high system performance for a wide glycemic state range (45-500 mg/dL). Reversibility and reproducibility were also tested over a range of time spans. Combined with existing BLI system technology, this sensor holds great promise for use as a wearable online continuous glucose monitoring system for patients in a hospital setting.Entities:
Keywords: bio-layer interferometry; competitive assay; continuous glucose monitoring; glucose dialysis method; needle-type sensor
Year: 2016 PMID: 27669267 PMCID: PMC5087370 DOI: 10.3390/s16101581
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Needle-type sensor performance and the influence of membrane size on the detection of glucose concentrations in defined buffer. (a) Schematic of the needle-type sensor. A syringe needle was modified and covered with a semi-permeable membrane to fabricate the needle-type sensor. Bovine serum albumin (BSA)-ligand conjugate was kept outside of the semi-permeable membrane; (b) A glucose sensing profile generated by the sensor. The wavelength shift increases with time as the sensor is exposed to the buffer containing the BSA-ligand conjugates, whereas the shift sharply decreases as the sensor is exposed to a solution of 500 mg/dL glucose (with BSA-ligand conjugate); (c) Membrane pore-size performance. The response of the sensor increased as the pore size increased from 50 nm to 200 nm. Con A: concanavalin A.
Figure 2Schematic of the dialysis method and the performance of the sensor for the detection of a broad range of glucose concentrations. (a) Schematic of the dialysis procedure. A dialysis bag containing a mixture of glucose oxidase and human serum was used. This dialysis bag was kept in the container of human serum to reduce the glucose concentration of the serum. The time vs. glucose concentration is the glucose attrition of the serum measured with a standard strip sensor; (b) Performance of the needle-type sensor. The sensor shows a peak at 1800 s for the batch 1 samples (dialyzed human serum: 45 mL/dL, 100 mL/dL, 250 mL/dL, and 500 mL/dL), which have glucose concentrations comparable to that observed in hypoglycemia, whereas this peak is not observed in the profile for batch 2 (normal human serum: 100 mL/dL, 250 mL/dL, and 500 mL/dL). Averaged standard deviation for a constant glucose concentration (i.e., 100 mg/dL and t = 1800–3600 s of batch 2) was measured as 0.005496 nm, corresponding to 2.77% of the wavelength shift; (c) Repeatability test results (n = 3) for glucose concentration range of 100–500 mg/dL.
Figure 3Comparison of the needle-type sensor with the standard bio-layer interference (BLI) sensor. Nearly identical profiles were observed when tested with a wide range of glucose concentrations. The profiles indicate that the reproducibility of the needle-type sensor is more stable than the standard BLI sensor.