Accelerometer-Based, Portable Navigation (KneeAlign) vs Conventional Instrumentation for Total Knee Arthroplasty: A Prospective Randomized Comparative Trial.

Monther A Gharaibeh1, Gandhi N Solayar1, Ian A Harris2, Darren B Chen1, Samuel J MacDessi1.   

Abstract

BACKGROUND: Accelerometer-based, portable navigation devices have been introduced as a less invasive and simpler technique to perform navigated surgical implantation of knee prostheses. They have been postulated to have better accuracy than conventional instruments in restoration of alignment in total knee arthroplasty.
METHODS: A total of 190 patients were enrolled in this prospective, randomized controlled trial and underwent total knee arthroplasty using either the KneeAlign or conventional guides. Multiplanar alignment was evaluated with a CT imaging protocol.
RESULTS: A total of 86.5% of portable navigation device and 82.2% of conventional group had a postoperative hip-knee angle within 3° of neutral alignment (P = .54). There was no significant difference between the 2 groups for component coronal and sagittal plane alignment. Portable navigation device did not significantly increase the time to perform the surgery.
CONCLUSION: Portable navigation device demonstrates accurate restoration of alignment; however, there was no statistically significant difference when compared with conventional guides.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  KneeAlign; accelerometer; computer-assisted surgery; conventional guides; knee arthroplasty

Mesh:

Year:  2016        PMID: 27667534     DOI: 10.1016/j.arth.2016.08.025

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  7 in total

1.  Augmented marker tracking for peri-acetabular osteotomy surgery.

Authors:  Silvio Pflugi; Rakesh Vasireddy; Till Lerch; Timo M Ecker; Moritz Tannast; Nane Boemke; Klaus Siebenrock; Guoyan Zheng
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-11-29       Impact factor: 2.924

2.  Accelerometer-based, hand-held navigation for improved knee alignment in total knee arthroplasty: An observational study.

Authors:  Arun M Swamy; Archit Goyal; Vijay Kumar Digge; Vikrant Manhas; Deepak Gautam; Rajesh Malhotra
Journal:  J Clin Orthop Trauma       Date:  2021-11-10

Review 3.  Clinical outcomes associated with robotic and computer-navigated total knee arthroplasty: a machine learning-augmented systematic review.

Authors:  Quinlan D Buchlak; Joe Clair; Nazanin Esmaili; Arshad Barmare; Siva Chandrasekaran
Journal:  Eur J Orthop Surg Traumatol       Date:  2021-06-25

4.  Total knee arthroplasty in the next decade: is navigation necessary?

Authors:  Graham S Goh; Ming Han Lincoln Liow
Journal:  Ann Transl Med       Date:  2020-06

5.  Fragility Index as a Measure of Randomized Clinical Trial Quality in Adult Reconstruction: A Systematic Review.

Authors:  Carl L Herndon; Kyle L McCormick; Anastasia Gazgalis; Elise C Bixby; Matthew M Levitsky; Alexander L Neuwirth
Journal:  Arthroplast Today       Date:  2021-10-11

Review 6.  Accelerometer-based navigation vs. conventional techniques for total knee arthroplasty (TKA): a systematic review and meta-analysis of randomized controlled trials.

Authors:  Juntan Li; Yuqi Zhang; Xiang Gao; Tianxu Dou; Xu Li
Journal:  Arthroplasty       Date:  2022-09-02

7.  Are There Differences in Accuracy or Outcomes Scores Among Navigated, Robotic, Patient-specific Instruments or Standard Cutting Guides in TKA? A Network Meta-analysis.

Authors:  Pierre-Alban Bouché; Simon Corsia; Agnès Dechartres; Matthieu Resche-Rigon; Rémy Nizard
Journal:  Clin Orthop Relat Res       Date:  2020-09       Impact factor: 4.755

  7 in total