Hang Ren1, Joseph L Bull2, Mark E Meyerhoff1. 1. Department of Chemistry, University of Michigan , 930 North University, Ann Arbor, Michigan 48109, United States. 2. Department of Biomedical Engineering, University of Michigan , 2200 Bonisteel Boulevard, Ann Arbor, Michigan 48109, United States.
Abstract
Nitric oxide (NO) releasing polymers are promising in improving the biocompatibility of medical devices. Polyurethanes are commonly used to prepare/fabricate many devices (e.g., catheters); however, the transport properties of NO within different polyurethanes are less studied, creating a gap in the rational design of new NO releasing devices involving polyurethane materials. Herein, we study the diffusion and partitioning of NO in different biomedical polyurethanes via the time-lag method. The diffusion of NO is positively correlated with the PDMS content within the polyurethanes, which can be rationalized by effective media theory considering various microphase morphologies. Using catheters as a model device, the effect of these transport properties on the NO release profiles and the distribution around an asymmetric dual lumen catheter are simulated using finite element analysis and validated experimentally. This method can be readily applied in studying other NO release medical devices with different configurations.
Nitric oxide (NO) releasing polymers are promising in improving the biocompatibility of medical devices. Polyurethanes are commonly used to prepare/fabricate many devices (e.g., catheters); however, the transport properties of NO within different polyurethanes are less studied, creating a gap in the rational design of new NO releasing devices involving polyurethane materials. Herein, we study the diffusion and partitioning of NO in different biomedical polyurethanes via the time-lag method. The diffusion of NO is positively correlated with the PDMS content within the polyurethanes, which can be rationalized by effective media theory considering various microphase morphologies. Using catheters as a model device, the effect of these transport properties on the NO release profiles and the distribution around an asymmetric dual lumen catheter are simulated using finite element analysis and validated experimentally. This method can be readily applied in studying other NO release medical devices with different configurations.
Entities:
Keywords:
distribution; finite element analysis; mass transport; medical devices; nitric oxide; polyurethanes
Biocompatibility
is central to the design and performance of medical
devices. The issues of biocompatibility include clot formation on
blood contacting devices,[1] foreign body
response on subcutaneous implants,[2] biofilm
formation, and microbial infections on all types of invasive devices.[3] One promising means to combat all of these issues
with a single agent is to employ nitric oxide (NO), which is endogenously
produced in the body. In the bloodstream, NO can prevent thrombosis
by inhibiting platelet adhesion and activation, while in soft tissues,
NO can promote wound healing.[4−7] It is also a potent antimicrobial/antibiofilm agent.
These versatile properties are desirable for medical devices, for
example, catheters. However, because NO is a gas molecule and is relatively
reactive, the use of NO donors together with proper release methods
are necessary to deliver NO locally to be effective clinically.[8] Nitric oxide donors can either be covalently
attached to, or noncovalently incorporated within polymers,[9,10] and such polymers are then applied either as coatings for devices,
including intravascular sensors, or as the bulk material of the devices
(e.g., catheters, intravascular sensors, and stents).[11−13]Two types of biomedical polymers, silicone rubber[14−17] and polyurethanes,[18−21] are extensively used in creating NO releasing materials because
of their innate compatibility with NO release chemistry, appropriate
mechanical properties and high stability in vivo.[22] However, the transport properties of such polymers with
respect to NO diffusion rates, partition coefficients, etc., have
been less studied. Such transport properties not only can significantly
impact the NO release profile, including surface flux of NO, the longevity
of release, and time required to reach steady-state release, but also
can affect the distribution of NO release around the surface of actual
devices. On the other hand, such transport properties, once known,
can help predict the NO release profile and NO distribution around
the devices made of certain polymers, ultimately guiding the optimal
design/configuration of devices for biomedical applications.Mowery et al. examined the transport of NO in polymers and obtained
apparent diffusion coefficients for NO in PVC, silicone rubber, and
two aliphaticpolyether polyurethanes.[23] Although useful in predicting the average flux at steady-state,
transient processes, local flux distribution, and transport processes
coupled with chemical reactions cannot be accurately described using
apparent diffusion coefficient values alone. This is because this
apparent diffusion coefficient represents the mixed processes of diffusion
and partitioning. Polymers with same apparent diffusion coefficient
can display very different NO release profiles and distributions.
Moreover, the transport properties of NO in newly developed polyurethanes
that have been shown to exhibit improved stability and biocompatibility
in vivo have not been reported.[24]To better understand the time-dependent NO transport and release
processes in biomedical polyurethanes and to improve the design of
NO releasing scaffolds and devices, in this study, the transport properties
of NO in different polyurethanes, including classic aliphatic, aromatic
polyether polyurethanes, as well as novel silicone and polycarbonate
containing polyurethanes, are examined. The true diffusion coefficients
of NO in these polymers are separated from the partition process.
Finally, the effects of varied diffusion and partitioning on the NO
release profile as well as the distribution of NO at the surface of
multilumen catheters are simulated using finite element analysis and
compared with experimental results.
Theory
Gas Transport in Polymers:
Diffusion and Solution
Transport
of gas through polymers can be described by two different processes—diffusion
and solution. Here, solution is the process of gas partitioning between
the polymer phase and the gas or liquid phase, and therefore the term
partition and solution are used interchangeably in this paper. Figure summarizes a general
1D model for NO permeation from one stream (upstream) into another
stream (downstream) through a polymer layer, which can describe the
diffusion experiments conducted in this study, as well as NO release
process from a NO reservoir inside a polymeric membrane, e.g., electrochemical
(e-chem) NO release system.[24−26]
Figure 1
Schematic for NO transport through a polymer
film/wall with a thickness
of L. Dashed line indicates the concentration profile of NO.
Schematic for NO transport through a polymer
film/wall with a thickness
of L. Dashed line indicates the concentration profile of NO.Suppose gas molecules dissolved
in a solution (Domain I) with bulk
concentration C* permeate through a polymeric membrane
(Domain II) to the downstream solution on the other side of the membrane
(Domain III) because of the concentration gradient (see Figure ). First, the dissolved gas
in Domain I is transported to close proximity to the polymer/solution
interface (Boundary I) by diffusion (diffusion coefficient D) and convection. Second, the gas molecules close to the
interface (Boundary I) partition between the polymer and solution
phases. This partitioning process can be viewed as two elementary steps: adsorption
(transfer from solution into polymer, with a heterogeneous rate constant ka), and desorption (transfer from polymer into
solution, with a heterogeneous rate constant kd). The ratio of ka/kd determines the partition coefficient K between the two phases. In the third step, the gas molecules that
have partitioned into the polymer membrane (Domain II) are transported
to the other interface of the membrane (at x = L) by diffusion (diffusion coefficient D̅; the bar denotes the membrane phase). The fourth step involves again
the partition between polymer/solution interface (Boundary II) with
the same ka and kd as in the second step. Finally, the gas molecules that have
entered the receiving solution (Domain III) will be transported away.
The driving force for all these processes is the chemical potential
gradient, or equivalently, the concentration gradient.Several
assumptions are made to simplify the mathematical model
for the diffusion process shown in Figure . First, we assume that all the diffusion
processes are Fickian (i.e., magnitude of the flux is proportional
to the concentration gradient, and diffusion coefficient is concentration
independent). This assumption is valid as the gas molecule is small
and the concentration of gas in the experiment is low, ensuring no
significant change of the polymer, structurally or dynamically, by
the presence of the diffusing gas molecules (i.e., no swelling or
plasticization). Second, the membrane is isotropic and homogeneous.
This should be macroscopically true for the polymers under study.
On the basis of these assumptions, diffusion equations in the three
domains coupled by partitioning in 1D as well as the corresponding
initial values and boundary conditions are shown belowDomain
IDomain II:Domain III:Equations , 4, and 8 are diffusion equations in the different domains, and the other
equations are initial values and boundary conditions. Equations and 5, and eqs and 9 couple the diffusions in Domains I, II and Domains
II, III, respectively, by enforcing continuous fluxes across the boundaries.For the diffusion experiment, assuming stirring provides sufficient
mixing in Domain I, the concentration of NO in solution can be treated
as homogeneous. The large amount of solution volume in Domain I as
compared to the membrane volume (Domain II) ensures that bulk concentration
of NO in Domain I remains essentially constant during a typical diffusion
experiment. The homogeneity and time invariance of NO concentration
simplifies eqs and 2 toAssuming
fast equilibrium of partition at
the interface, eqs and 5 can be reduced toSimilarly, because NO in the downstream (Domain
III) is rapidly pumped into the detector for measuring the rate of
NO transport, or scavenged very quickly by oxyhemoglobin when practically
using NO release devices in contact with blood, the concentration
of NO in Domain III is essentially zero at all times. eqs and 10 are
therefore reduced toAnd by
analogy for fast equilibria, eqs and 9 becomeUnder
these conditions, the complicated coupled
diffusion in three domains is now simplified into the diffusion process
in Domain II alone, with governing eq eq and boundary conditions provided by eqs and 14 and
initial value, eq .Such diffusion equations can be solved exactly, and integrating
the flux at x = L over time yieldswhere
Q(L, t) is the accumulated amount
of gas that penetrates the membrane per
unit area at time t. When t →
∞, the exponential term is droppedTherefore,
plot of the amount of permeated
gas versus time will approach a linear asymptote as t→ ∞, with an intercept at the x-axisτ
is defined as the time lag. Once the
membrane thickness L and time lag τ are obtained,
the diffusion coefficient can be calculated from eq .The first derivative of eq with respect to time
givesThe first term of eq corresponds to the steady-state portion
of the flux, whereas the second term corresponds to the transient
portion of the flux. As t → ∞, the
flux reaches steady-state:From the steady-state flux Jss, the product
of D̅K can be obtained.
Together with the D̅ obtained from the transient
time lag, the partition coefficient K can be derived.
This forms the basis for measurement of both the diffusion coefficients
and partition coefficients of NO in this study for a variety of biomedical
polymers.Under other conditions, for example, electrochemical
NO generation
from catheters,[26,27] the assumption of effective stirring
in Domain I is no longer valid and a significant diffusion layer will
exist within the inner source solution phase. To solve this more complicated
problem, diffusion equations in Domain I and II need to be coupled
together. Distribution of NO around an actual catheter surface is
also of interest during the application of these type of devices.
For this purpose, however, the 1D model is not sufficient to describe
the distribution process. This complicated case can be solved in higher
dimension numerically using finite element analysis, which is described
in the next section.
Experimental Section
Materials
and Instrument
Sodium nitrite (99.99%), potassium
iodide, and sulfuric acid were purchased from Sigma-Aldrich (St. Louis,
MO) and used as received. All the solutions were prepared with deionized
water from a Milli-Q system (18 MΩ cm–1; Millipore
Corp., Billerica, MA).Carbosil 20 80A and Bionate 80A were
from DSM (Heerlen Netherlands), while Elast-Eon 5–325 80A (E5–325)
was from AorTech International plc (Weybridge, UK). Tecoflex SG-80A
and Pellethane 80AE were gifts from Lubrizol (Cleveland, Ohio). Silicone
rubber sealant (RTV-3140) was a product of Dow Corning (Midland, MI).
Membrane Preparation
All the polyurethane films were
prepared by casting a 10 wt % solution of the polymer in THF in a
6 cm diameter glass O-ring on a glass slide. The slide was left to
dry in a fume hood for 24 h and then placed under vacuum for further
drying for another 4 h period. Siliconerubber films were prepared
by casting a suspension of RTV sealant in THF within a 6 cm Teflon
O-ring on a Teflon slide with 48 h drying under ambient conditions.
The chemical composition selective and properties of all the polymers
used in this study are listed in Table S1.
NO Transport Measurements
In a homemade diffusion cell,
a membrane was clamped in between the two parts of the cell (see Figure S1 for the experimental setup). The temperature
of the cell was controlled by a water bath. The NO was generated from
one part of the cell reproducibly via quantitative conversion of nitrite
to NO in the presence of acid and reducing agent (e.g., 2 NO2– + 2 I– + 4 H+ →
2 NO + 2 H2O + I2) as reported previously.[28] Briefly, solutions of 0.1 M H2SO4 and 5% KI were prepurged with N2 for 20 min to
remove O2. A 3.5 mL aliquot of each solution was then added
to the left side of the diffusion cell. Each side of the cell was
purged thoroughly with N2 again for another 20 min to eliminate
O2. The solution on the left side was then vigorously stirred
throughout the entire time of the experiment. Then, the time was recorded
when a 50 μL aliquot of a NaNO2 standard solution
(5 mM) was injected into the 7 mL solution on the left side of the
cell. The NO flux was measured in real time with chemiluminescence
using a nitric oxide analyzer (NOA) (Sievers 280i, GE Analytics, Boulder
CO) until the steady-state flux was reached.Alternatively,
0.1 M H2SO4 and 5% freshly made ascorbic acid
can also be used to generate NO from nitrite solutions when I2 adsorption was found to be significant on the polymer surface
(for Carbosil 20 80A).
Fabrication of Electrochemical NO Releasing
Silicone and Polyurethane
Catheters
Standard silicone tubing was purchased from VWR.
Polyurethane catheter tubing was prepared by dip coating of a 15 wt
% of Tecoflex SG80A in THF on a 2.0 mm diameter straight stainless
steel mandrel (McMaster-Carr, IL). The tubing obtained had a wall
thickness of ∼0.3 mm and were cut off for the fabrication of
the electrochemical NO releasing catheters. The procedures for the
fabrication of the electrochemical NO releasing catheters were similar
as previously reported.[24,27] Briefly, the catheter
tubing was filled with an aqueous solution of 0.4 M NaNO2, 1 mM copper(II)- 1,4,7-trimethyl-1,4,7-triazacyclononane, 0.15
M NaCl, and 0.5 M HEPES buffer (pH 7.3) using a microsyringe after
one end of the catheter was sealed using polyurethane/THF solution
(15 wt %). A Teflon-coated Pt wire and a Teflon-coated Ag/AgCl wires
were exposed 3 and 6 cm at the tip, respectively. The Ag/AgCl wire
was coiled onto the Teflon coating of the Pt wire, and both wires
were inserted into the catheter. The other opening of the catheter
was then sealed with the same polyurethane/THF solution (15 wt %).
Measurement of Asymmetric Release of NO by Agar Immobilization
An electrochemical NO releasing catheter was mounted on a glass
slide with tape. A heat melted 1% agar solution was poured onto a
glass slide with four additional glass coverslips forming a container,
and the agar incorporated the catheter when cold. The agar was further
cut so that the cross-section geometry was rectangular (6 cm ×
1 cm, W × H) with the catheter in the center. The NO release
was then turned on, and the cumulative amount of NO at each side of
the catheter was quantified by measuring the nitrite content using
chemiluminescence. Briefly, a piece of the agar sample (6 cm ×
1 cm × 2 cm, W × H × L) was dissolved in DI water,
and aliquots of the solution were injected into a cell containing
degassed 5% KI and 0.1 M H2SO4 solution to convert
nitrite to NO. The cell was connected to a nitric oxide analyzer for
NO measurement via chemiluminescence. The data at each time point
are triplicate.
Simulation Methods
Finite element
analysis via Comsol
Multiphysics (5.0b) was used to simulate the effect of partition coefficient
and diffusion coefficient on the NO release profiles when using single
lumen catheters and the NO distribution when employing multilumen
catheters.For response time estimation, a 2-D model of the
cross-section of a single lumen catheter was implemented. Similar
2-D equations for diffusion coupled by partition as described in eqs (4) (8) and (3) (5) (6) (9) are the governing equations for this transport study, except the
Neumann boundary condition of constant flux at the electrode surface
was used instead of constant concentration. In Domain III, an NO sink
at 0.2 mm away from the catheter surface was purposely set to mimic
the fact that NO is removed from the surface of the catheter very
fast, either into the NOA by purging, or reacting with oxyhemoglobin
when placed within the bloodstream in vivo. For simulation of the
distribution of NO around the outer surface of multilumen catheters,
a similar model was used except that the cross-section geometry was
different. The cross-section geometries of the different catheter
models used are shown in Figure , and the related parameters are reported in Table S2.
Figure 2
Cross-section geometries for (A) single
lumen, (B) commercial dual
lumen, and (C) proposed triple lumen catheter studied by finite element
analysis.
Cross-section geometries for (A) single
lumen, (B) commercial dual
lumen, and (C) proposed triple lumen catheter studied by finite element
analysis.
Results and Discussion
Diffusion
Studies of NO Through Polyurethanes
A typical
plot of NO flux (J)
and the accumulative amount of permeated NO (Q) vs time for NO diffusion through an E5–325
polyurethane membrane is shown in Figure . From the steady-state flux, together with
membrane thickness, D̅K can be obtained according
to eq . The accumulated
amount of the permeated NO (Q) vs time can be used to estimate time lag (τ) and calculate
the diffusion coefficient. Similar experiments were performed for
the other polyurethanes and siliconerubber films (structure description
and selected physical properties of all the polymers under investigation
are listed in Table S1) at both 25 and
37 °C. The measured diffusion coefficient and partition coefficients
of NO in these polymers are summarized in Table .
Figure 3
Typical NO flux profile (J, black) and amount of permeated NO (QNO, blue) vs time for a 368 μm E5–325
polyurethane
film at 25 °C in a diffusion experiment. Red dotted line denotes
the asymptote used for deriving the time lag (τ).
Table 1
Summary of Diffusion Coefficients
(D̅) and Partition Coefficients (K) of NO in Various Polymers at 25 and 37 °C Obtained in This
Studya
polymer
description
thickness
(μm)
D̅25 °C (× 10–6 cm2 s–1)
D̅25 °C (× 10–6 cm2 s–1)
K25 °C
K37 °C
Silicone
silicone rubber
1107 ± 16
16 ± 2
20 ± 4
5 ± 1
4 ± 1
Elast-Eon 5–325
silicone polyurethane
621 ± 10
6 ± 1
8.5 ± 0.8
3.8 ± 0.5
3.3 ± 0.3
Tecoflex SG80A
polyether polyurethane
437 ± 10
1.2 ± 0.5
1.9 ± 0.5
2.5 ± 0.8
2.2 ± 0.3
Carbosil 20 80A
silicone polycarbonate polyurethane
442 ± 8
0.7 ± 0.1
1.4 ± 0.5
1.2 ± 0.4
0.3 ± 0.1
Pellethane 80A
polyether polyurethane
128 ± 4
0.5 ± 0.1
0.8 ± 0.3
1.5 ± 0.4
1.1 ± 0.2
Bionate 80A
polycarbonate polyurethane
174 ± 4
0.3 ± 0.1
0.7 ± 0.1
0.7 ± 0.2
0.6 ± 0.1
n = 3 replicates.
Results are reported as mean ± standard deviations.
Typical NO flux profile (J, black) and amount of permeated NO (QNO, blue) vs time for a 368 μm E5–325
polyurethane
film at 25 °C in a diffusion experiment. Red dotted line denotes
the asymptote used for deriving the time lag (τ).n = 3 replicates.
Results are reported as mean ± standard deviations.The diffusion coefficient of NO
(D̅) determined
in the different polymers is negatively correlated to the density
(ρ) of the polymer (see Table S1).
This can be understood by the free volume theory,[30] in which the polymer is considered to be a combination
of hard and soft segments. The transport through the soft segment
is a dominant pathway for the gas to diffuse through the polymer.
Some authors have referred to the regions of soft segment within the
hard segment structures as holes, and describe diffusion in a polymer
as “hopping” between the “holes” of free
volume,[31] with the relationshipwhere A and B are constants related to the size
of the penetrants and the hole
size of the polymer matrix, respectively, and fv is the fraction of the free volume. Assuming similar densities
of different polymer chains (occupied volumes) in this study, the
specific density of a polymer is negatively proportional to fv (Figure S2). In
fact, a direct relationship between fv and D̅ has been established experimentally
by diffusion and positron annihilation lifetime experiments.[31]From the diffusion study, silicone rubber
has the highest D̅ for NO, and all the polyurethanes
containing PDMS
segments examined in this study exhibit significantly higher diffusion
coefficients for NO than similar polyurethanes without the PDMS segments
(e.g., E5–325 vs Pellethane, Carbosil vs Bionate, see the detailed
structure information in Table S1). This
is because that the PDMS chain is flexible with a low rotation barrier,
which allows for a large free volume and therefore fast diffusion
for penetrants like NO. The large free volume of PDMS chain is also
indicated by the low glass transition temperature (Tg) of neat PDMS (−112 °C).[32]The relatively slow diffusion of NO in Bionate and
Carbosil can
be attributed to the interaction between the polycarbonate chain and
the urethane moiety. Although the polycarbonate chains, (poly(1,6-hexyl
1,2-ethyl carbonate) (PHEC), exhibit a relatively low Tg of −70 °C in the neat form, they can form
strong hydrogen bonds with the urethane moieties in polyurethanes.[32] Such hydrogen bonding significantly decreases
the free volume by a closer packing of the polymer chains and a reduction
in chain mobility, which is also indicated by the much higher Tg (−7 °C) of the PHEC segments in
polycarbonate copolymers of polyurethanes.[32] This reduced free volume of PHEC chains in polyurethanes and decrease
the diffusion of NO in these polymers.Tecoflex SG80A and Pellethane
80A are both polyurethanes with polyether
chains as soft segments but exhibit quite different diffusion properties
for NO (see Table ). Tecoflex SG80A contains urethane segments from 4,4′-methylene-bis(cyclohexyl
isocyanate) (HMDI) and 1,4-butanediol (BDO), whereas Pellethane 80A
contains urethane segments from 4,4′-methylene-bis(phenyl isocyanate)
(MDI) and BDO.[33−35] As the aromatic MDI is more rigid than the aliphatic
HMDI, polyurethanes derived from MDI segments have a higher barrier
for chain rotation and therefore lower free volume. Therefore, Tecoflex
SG80A exhibits a higher diffusion coefficient for NO than Pellethane.In addition to the chain flexibility and free volume, polymer morphology
in block copolymers also significantly affects the diffusion of gas
molecules.[36] This is especially true for
polyurethanes as they tend to microphase separate into different morphologies,
including a continuous phase with spheres, cylinders, and lamellae.[37] The exact morphology depends on the miscibility
of the blocks of the copolymer, stoichiometry of the blocks, and the
preparation conditions (including thermal history).[38] In polyurethanes with PDMS chain, such as E5–325
and Carbosil, the PDMS segments can undergo phase separation of nearly
100%,[32,39] which allows the fraction of the PDMS phase
to be simply represented by the stoichiometry. Effective media theory
has been proposed to model diffusion within polymer blends with different
morphology, including spheres, cylinders, and lamellae.[40] Such methods were first developed for solving
conductivity of composite materials, and the derived equations can
be directly applied for diffusion by changing the conductivity into
permeability (see the derived equations in the Supporting Information). The diffusion of gas in polyurethanes
can be modeled by considering the compositions of cells with morphologies
as shown in Figure A in random orientations.
Figure 4
Effect of volume fraction of PDMS in polyurethanes
on the diffusion
coefficient of NO from experiment and modeling. A) Cells of typical
morphology (lamellae, spheres, and cylinders) for block copolymers
with phase separation used for modeling; Predicted (lines) and experimentally
obtained (triangles, error bars denote standard deviations from n = 3) diffusion coefficient of NO (D) vs volume fraction of PDMS (XA) based on different morphologies for B) silicone polyurethanes and
(C) silicone polycarbonate polyurethanes.
Effect of volume fraction of PDMS in polyurethanes
on the diffusion
coefficient of NO from experiment and modeling. A) Cells of typical
morphology (lamellae, spheres, and cylinders) for block copolymers
with phase separation used for modeling; Predicted (lines) and experimentally
obtained (triangles, error bars denote standard deviations from n = 3) diffusion coefficient of NO (D) vs volume fraction of PDMS (XA) based on different morphologies for B) siliconepolyurethanes and
(C) silicone polycarbonate polyurethanes.E5–325 possesses segments of PDMS and MDI:BDO with
the PDMS
content as high as 66%. This micromorphology can be described as two
phases comprising a continuous PDMS phase (66%) and discontinuous
hard segment phase. From Figure B, the point for E5–325 lies close to the predicted
line for a cylinder of a hard phase in PDMS. However, a volume fraction
of 66% is likely to be near the percolation threshold.[41] Predictions near the percolation threshold using
this method often involves large error, because of uncertainty whether
a percolating network spans the entire material.[42] In the finite element model, the availability of percolating
network through the wall of the catheter depnds on morphologies of
the mode (Figure A),
which are idealized. The idealized morphology could lead to erroneous
model predictions near the percolation threshold, but it is accurate
for other volume fractions because the morphology does not affect
the availability of a percolation network.For Carbosil-20,
the PDMS chains phase separate completely because
of low compatibility with other chains. From percolation theory, PDMS,
with a volume fraction of 20%, is less likely to form a continuous
phase. Significant mixing of PHEC and the hard segment occurs, and
only 10–15% of hard segment phase is separated (∼4%
of total volume).[32] The small amount of
separated hard segment phase can be neglected to simplify the description
as a two-phase model. The morphology is, therefore, described as a
continuous PHEC/hard segment mixed phase with islands of PDMS. From Figure C, Carbosil-20 is
likely to contain a lamellae structure of the PDMS phase and the PHEC/hard
segment mixed phase.In this study, diffusion coefficients and
partition coefficients
of NO were obtained only at 25 and 37 °C as these temperatures
are most relevant for biomedical applications. As expected, the diffusion
coefficients of NO increase for all the polymers tested when the temperature
increases (see Table ). However, we did not attempt to derive an apparent Arrhenius activation
energy from the two temperature data points. Such an activation energy
would contain contributions from both the diffusant (NO) and the polymer
and is less useful considering the complex compositions and phase
transitions of the different polymers at different temperatures.Errors in the diffusion experiments stem from many factors with
major ones being the measurement of the thickness of the membranes,
the uncertainty from the response time of the NO measurement (∼2
s), and the effectiveness of convection by stirring the NO generating
solution. During the diffusion experiment, asymmetry exists as one
side of the membrane is contacting an aqueous phase (Domain I in Figure ), whereas the other
side is contacting a gas phase (Domain III in Figure ). Such asymmetry could potentially affect
the adsorption and desorption processes (by changing the local structure
near the interfaces). In this experiment, the adsorption and desorption
processes have been assumed to occur very fast for both the water/polymer
interface and the gas/polymer interface and, therefore, can be negligible.
Indeed, diffusion experiments with different thickness of membranes
yield very similar diffusion coefficients, suggesting that the surface
process is not rate limiting. The purpose of having a gas phase instead
of a solution on the other side is to minimize the response time for
the NO detection and to ensure an effective zero concentration of
NO in the by quickly forcing the arriving NO into the detector.
Effect of Diffusion and Partition on Response Time of NO Release
Partition and diffusion describe NO transport in polymers and therefore
affect the NO release profile of some NO release materials/systems.
For example, the duration for the NO release to reach steady-state
release, defined as response time, is of importance as it can affect
the performance and the use of NO releasing devices. Catheters made
from silicone and Tecoflex SG 80A with electrochemical NO release
system are used in this work to demonstrate such an effect. The electrochemical
NO generation system can produce a steady flux of NO from the electrode
surface with constant voltages, which can release from the catheter
surface via diffusion through the inner filling solution and the wall
of the catheter. Such NO releasing catheters made from different materials
with the same geometry show very different release profiles. As seen
in Figure A, for a
silicone catheter, the surface flux of NO reaches a steady state within
5 min after the NO release is turned on, whereas for a catheter made
from Tecoflex SG80A, it takes more than 60 min to reach the steady-state
release. This is explained by the fact that the diffusion coefficient
of NO in silicone is 13 times greater than that in Tecoflex SG80A
(see Table ).
Figure 5
NO release
profile for electrochemical NO releasing catheters (0.3
mm wall thickness, 2.6 mm o.d.). A voltage of −0.3 V was applied
at time 0 to generate a steady flux of NO from the electrode surface.
A) Experimental data from single-lumen catheters made from silicone
(black) and Tecoflex SG80A (red); B) Simulation of the effect of diffusion
and partition on NO release profile from electrochemical NO releasing
catheters.
NO release
profile for electrochemical NO releasing catheters (0.3
mm wall thickness, 2.6 mm o.d.). A voltage of −0.3 V was applied
at time 0 to generate a steady flux of NO from the electrode surface.
A) Experimental data from single-lumen catheters made from silicone
(black) and Tecoflex SG80A (red); B) Simulation of the effect of diffusion
and partition on NO release profile from electrochemical NO releasing
catheters.To obtain the more general effects
of diffusion and partition on
NO release from single lumen catheters, we applied finite element
analysis (using Comsol Multiphysics) to simulate the surface NO flux
after NO is turned “on” for the electrochemical NO releasing
catheters. From the simulation, the response time of NO release becomes
smaller as D̅ and K increase
(see Figure B). This
simulation result together with D̅ and K obtained in the diffusion study (Table ) immediately suggests that silicone and
E5–325 is the preferred material for electrochemical NO releasing
catheters in terms of response time. For a typical single lumen catheter
with 0.3 mm wall thickness, the response time for silicone and E5–325
will be ≤20 min.The other polyurethanes with low D̅ and K values, although not ideal
for direct application in electrochemical
NO release catheter, could be potentially used as release barriers
to slow down the NO release and mitigate burst release in many NO
release materials, and to prolong the lifetime NO release. Indeed,
drug delivery using nanoparticles, where prolonged release is often
hard to achieve, calls for materials with low permeability as the
outer barrier.[43]
Effect of Diffusion and
Partition on Distribution of NO for
Multi-Lumen Catheters
Another aspect of NO release is the
distribution of NO release on the surface of the devices. For example,
in clinical practice, the application of electrochemical NO release
in catheters requires the use of multilumen catheters because one
of the lumens needs to be dedicated for the electrochemical NO generation
system. However, as most commercial multilumen catheters are not centrosymmetric,
NO distribution around the outer surfaces of such catheters can be
asymmetric. Depending on the severity, such asymmetry can potentially
create problems (e.g., one side of the device being less biocompatible
than the other). On the other hand, polymers with a high partition
coefficient could serve as a reservoir for NO release and could promote
a more symmetrical distribution of NO when coupled with a high diffusion
coefficient. Both D̅ and K, besides the geometry of the inherently asymmetric device, could
impact the distribution of NO.To study such asymmetry effect,
as a first step, a commercial dual lumen silicone catheter was simulated
using D̅ and K values obtained
in this study with NO release electrode in the right lumen (cross-section
geometry of this catheter is shown in Figure A). Such a catheter has already been used
in antithrombotic, antimicrobial studies as well as in preparing the
new NO release PO2 sensing catheters.[26,27,29] The local surface flux distribution
on the catheter surface over time with 0% O2 (no sink for
NO) is plotted in Figure B. It can be seen that a significant asymmetry exists initially
(e.g., at 10 min), which can be also observed in the concentration
color map in Figure S3. To quantify the
worst-case scenario of the asymmetry, we used the ratio of the highest
to the lowest local fluxes on the surface and termed it as the maximum
flux ratio thereafter. For this dual lumen silicone catheter, the
maximum surface flux ratio decreases to <2 after 20 min in the
absence of O2 (Figure D). With ambient O2 reacting with NO (as
a sink for NO), the surface flux reaches a steady-state after 30 min
(Figure C), and the
maximum ratio of surface flux reaches 2.5 (Figure D). Such asymmetry was also probed experimentally
by measuring the cumulative nitrite content at different sides of
the catheter in an agar gel (released NO reacts with O2 to form nitrite that stays in the agar) into which electrochemical
NO releasing catheter was placed. A schematic for the experimental
measurement of the asymmetry is shown in Figure A. As shown in Figure B, the cumulative NO from the nitrite measurement
agrees well with that from simulation with a correlation coefficient
(Pearson’s R2) of 0.960.
Figure 6
Local flux
of NO on surface of a 7-Fr Cook dual lumen silicone
catheter at room temperature in N2 and in air. A) illustration
of the polar angle for the dual lumen catheter; B) polar graph showing
the local surface flux vs time with 0% and C) 21% O2; D)
Maximum surface flux ratio vs time under 0 and 21% O2.
Figure 7
Experiment probing the distribution of NO by
around a dual lumen
catheter. A) Experimental setup of catheter immobilization using agar;
B) cumulative NO in the left and right domain of the catheter as measured
by the agar immobilization experiment (dot) and simulation (line).
The error bars indicate standard deviations from n = 3 samples. Pearson’s R2 is 0.960 for red curve
and 0.957 for blue curve.
Local flux
of NO on surface of a 7-Fr Cook dual lumen silicone
catheter at room temperature in N2 and in air. A) illustration
of the polar angle for the dual lumen catheter; B) polar graph showing
the local surface flux vs time with 0% and C) 21% O2; D)
Maximum surface flux ratio vs time under 0 and 21% O2.Experiment probing the distribution of NO by
around a dual lumen
catheter. A) Experimental setup of catheter immobilization using agar;
B) cumulative NO in the left and right domain of the catheter as measured
by the agar immobilization experiment (dot) and simulation (line).
The error bars indicate standard deviations from n = 3 samples. Pearson’s R2 is 0.960 for red curve
and 0.957 for blue curve.To continue using this commercial dual-lumen catheter as
an example,
the more general effect of diffusion and partition (in the relevant
range) on the asymmetry of NO distribution was further studied by
simulation. The lowest diffusion coefficient in the simulation (D̅ = 5 × 10–7 cm2 s–1) shows the most asymmetric distribution of
surface NO flux, with a maximum surface flux ratio of >10 (see Figure A). Such a large
asymmetry is partially a transient effect as the low diffusion coefficient
cause the NO to permeate one side of the catheter (the right side
of the catheter in Figure A) before a significant amount appears on the other side.
The symmetry is enhanced over time, but the enhancement is much faster
for larger diffusion coefficients (see Figure B). Overall, the symmetry is always significantly
better when diffusion coefficients are larger. The same trends are
also true when the reaction of NO with O2 is also considered,
although the exact symmetry is worse compared to that without the
reaction (see Figure C, D).
Figure 8
Effect of different diffusion coefficients (D̅) on the distribution/asymmetry of NO on a dual lumen catheter (same
configuration and polar angle as shown in Figure A) over time with K = 2.5.
A) Polar graph showing the distribution of local NO flux at the catheter
surface at 0% O2; B) maximum flux ratio vs time with 0%
O2; C) Polar graph showing the distribution of local NO
flux at the catheter surface with 21% O2; D) maximum flux
ratio vs time with 21% O2.
Effect of different diffusion coefficients (D̅) on the distribution/asymmetry of NO on a dual lumen catheter (same
configuration and polar angle as shown in Figure A) over time with K = 2.5.
A) Polar graph showing the distribution of local NO flux at the catheter
surface at 0% O2; B) maximum flux ratio vs time with 0%
O2; C) Polar graph showing the distribution of local NO
flux at the catheter surface with 21% O2; D) maximum flux
ratio vs time with 21% O2.The effect of partition coefficient on the distribution is
more
complex and is time-dependent. Without O2, the largest
partition coefficient (K = 10; NO partitions favorably
in the polymer) yields the lowest average flux of NO and worst distribution
at the 20 min (see Figure A, B). This is likely because that, with a high partition
coefficient, more NO needs to be dissolved in the polymer phase before
being released. A partition coefficient of 2 shows the best distribution
at all time. With 21% O2 present, the average surface flux
increases as K increases, but the trend for asymmetry
vs K changes with time (Figure C, D). Within the first 18 min K = 2 offers the best distribution, whereas K = 5
shows the best distribution between 18 and 36 min. After 36 min, a K = 10 exhibits the best distribution. The symmetry stops
to change after 70 min for all K, with K = 10 giving the best maximum flux ratio of 3.6 (Figure D).
Figure 9
Effect of partition coefficient
(K) on the distribution
of NO for a 7-Fr Cook dual lumen catheter at 20 min in 0% O2 and 21% O2. Diffusion coefficient D̅ = 2 × 10–6 cm2 s–1. A) Polar graph showing the distribution of NO flux at catheter
surface without O2; B) maximum flux ratio vs time without
O2; C) Polar graph of showing the distribution of NO flux
at catheter surface with 21%; D) maximum flux ratio vs time with 21%
O2.
Effect of partition coefficient
(K) on the distribution
of NO for a 7-Fr Cook dual lumen catheter at 20 min in 0% O2 and 21% O2. Diffusion coefficient D̅ = 2 × 10–6 cm2 s–1. A) Polar graph showing the distribution of NO flux at catheter
surface without O2; B) maximum flux ratio vs time without
O2; C) Polar graph of showing the distribution of NO flux
at catheter surface with 21%; D) maximum flux ratio vs time with 21%
O2.For even better distribution
of NO release, a more symmetric design
for multilumen catheters is also proposed and investigated by simulation
(cross-section geometry is shown in Figure C). The distribution of NO around the outer
surface is indeed more equal for this more symmetric design; the maximum
flux ratio falls below 2 after 20 min for all D̅ from 5.0 × 10–7 cm2 s–1 to 5.0 × 10–5 cm2 s–1 at a constant partition coefficient (see Figure A). Similarly, the maximum ratio of local
surface flux also falls below 2 after 20 min for K ranging from 0.5 to 10 at a constant diffusion coefficient (see Figure B).
Figure 10
Effect of
diffusion coefficient (D̅) and
partition coefficient (K) of NO in the polymer on
the maximum flux ratio of an electrochemical NO releasing catheters
with a more symmetric design as shown in Figure C (NO releasing electrode in the middle lumen).
(A) Maximum flux ratio with various D̅ at K = 2; (B) Maximum flux ratio with various K at D̅ = 2 × 10–6 cm2 s–1.
Effect of
diffusion coefficient (D̅) and
partition coefficient (K) of NO in the polymer on
the maximum flux ratio of an electrochemical NO releasing catheters
with a more symmetric design as shown in Figure C (NO releasing electrode in the middle lumen).
(A) Maximum flux ratio with various D̅ at K = 2; (B) Maximum flux ratio with various K at D̅ = 2 × 10–6 cm2 s–1.
Conclusions
Diffusion and partition coefficients for
NO in silicone and different
biomedical polyurethanes have measured using the time-lag method.
The diffusion coefficients for block bi- and tripolymerpolyurethanes
containing PDMS segments are explained by effective medium theory
with different microphase morphologies. The effect of D̅ and K on NO release response time as well as asymmetric
NO distribution at the outermost surfaces of single and dual-lumen
catheters have been examined both experimentally and by simulation.
Polymers with large D̅ and K for NO exhibit faster response times and catheters made from polymers
with large D̅ yield enhanced symmetry of NO
distribution on the outer surface of the catheter. A more symmetric
design for a multilumen catheter is proposed, and the NO distribution
is indeed more symmetric for all the partition coefficients and diffusion
coefficients being modeled. It should be noted that the transport
properties of NO in these biomedical polymers (diffusion and partition)
obtained here along with the simulation method demonstrated can be
readily applied for the simulation of the more general NO release
materials/devices not limited to catheters, by simply inputting the
geometry of the devices and the mass transport properties of the materials.
Overall, this combination provides a powerful means to design, study,
and understand novel NO releasing devices, including the new type
of electrochemically generated NO release catheters recently reported
by our group.[28,43]
Authors: Huiping Zhang; Gail M Annich; Judiann Miskulin; Kathryn Osterholzer; Scott I Merz; Robert H Bartlett; Mark E Meyerhoff Journal: Biomaterials Date: 2002-03 Impact factor: 12.479
Authors: Melissa M Reynolds; Joseph E Saavedra; Brett M Showalter; Carlos A Valdez; Anna P Shanklin; Bong K Oh; Larry K Keefer; Mark E Meyerhoff Journal: J Mater Chem Date: 2010-01-01
Authors: R E Solís-Correa; R Vargas-Coronado; M Aguilar-Vega; J V Cauich-Rodríguez; J San Román; A Marcos Journal: J Biomater Sci Polym Ed Date: 2007 Impact factor: 3.517
Authors: Hang Ren; Megan A Coughlin; Terry C Major; Salvatore Aiello; Alvaro Rojas Pena; Robert H Bartlett; Mark E Meyerhoff Journal: Anal Chem Date: 2015-08-05 Impact factor: 6.986
Authors: Hang Ren; Jianfeng Wu; Chuanwu Xi; Nicolai Lehnert; Terry Major; Robert H Bartlett; Mark E Meyerhoff Journal: ACS Appl Mater Interfaces Date: 2014-03-14 Impact factor: 9.229