Literature DB >> 27642727

Monte Carlo based investigations of electron contamination from telecobalt unit head in build up region and its impact on surface dose.

A S Jagtap1, T Palani Selvam2, B J Patil3, S T Chavan4, S N Pethe4, Gauri Kulkarni5, S S Dahiwale5, V N Bhoraskar5, S D Dhole6.   

Abstract

A Telecobalt unit has wide range of applications in cancer treatments and is used widely in many countries all around the world. Estimation of surface dose in Cobalt-60 teletherapy machine becomes important since clinically useful photon beam consist of contaminated electrons during the patient treatment. EGSnrc along with the BEAMnrc user code was used to model the Theratron 780E telecobalt unit. Central axis depth dose profiles including surface doses have been estimated for the field sizes of 0×0, 6×6, 10×10, 15×15, 20×20, 25×25, 30×30cm2 and at Source-to-surface distance (SSD) of 60 and 80cm. Surface dose was measured experimentally by the Gafchromic RTQA2 films and are in good agreement with the simulation results. The central axis depth dose data are compared with the data available from the British Journal of Radiology report no. 25. Contribution of contaminated electrons has also been calculated using Monte Carlo simulation by the different parts of the Cobalt-60 head for different field size and SSD's. Moreover, depth dose curve in zero area field size is calculated by extrapolation method and compared with the already published data. They are found in good agreement.
Copyright © 2016 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Gafchromic film; Monte Carlo simulation; Percentage Depth Dose; Secondary electrons; Surface dose; Telecobalt unit

Mesh:

Substances:

Year:  2016        PMID: 27642727     DOI: 10.1016/j.apradiso.2016.09.012

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  2 in total

1.  Monte Carlo study on the secondary cancer risk estimations for patients undergoing prostate radiotherapy: A humanoid phantom study.

Authors:  Amir Ghasemi-Jangjoo; Hosein Ghiasi
Journal:  Rep Pract Oncol Radiother       Date:  2020-01-10

2.  Characterization of GafChromic EBT2 film dose measurements using a tissue-equivalent water phantom for a Theratron® Equinox Cobalt-60 teletherapy machine.

Authors:  Daniel Akwei Addo; Elsie Effah Kaufmann; Samuel Nii Tagoe; Augustine Kwame Kyere
Journal:  PLoS One       Date:  2022-08-19       Impact factor: 3.752

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.