Reproducibility, interchangeability of measures, time to measure stabilization, and reference values of two tissue oximeters in healthy volunteers.

Alexis Cournoyer1, André Denault2, Sylvie Cossette3, Annik Fortier4, Raoul Daoust1, Massimiliano Iseppon1, Jean-Marc Chauny1, Eric Notebaert1.   

Abstract

This study aimed to compare two tissue oximeters, the INVOS 5100c and the Equanox 7600, in terms of their reproducibility and the interchangeability of their measures. In a randomized order, three measurements were taken at six different sites on both sides of the body in 53 healthy volunteers. Intraclass correlation coefficients (ICC) and within-subject standard deviation (Sw) were calculated for each device. The ICCs were compared using Fisher r-to-z transformation and the Sw were compared using paired-sample t-tests. We found no difference between the reproducibility of the INVOS {ICC=0.92 [95% confidence interval (CI) 0.90 to 0.93]} and Equanox [ICC=0.90 (95% CI 0.88 to 0.93)] in terms of ICCs (p=0.06). However, the Equanox [Sw=1.96 (95% CI 1.91 to 2.02)] showed a better Sw than the INVOS [Sw=2.11 (95% CI 2.05 to 2.17)] (p=0.019). Also, when compared directly to stable condition, the readings produced by the two oximeters varied considerably [ICC 0.43 (95% CI 0.36 to 0.49)]. When taken individually, both tissue oximeters displayed good reproducibility, the Equanox being slightly better than the INVOS in terms of absolute reproducibility. However, when compared, the oximeters showed poor interdevices agreement. Reference values were also described.

Mesh:

Year:  2016        PMID: 27637007     DOI: 10.1117/1.JBO.21.9.097003

Source DB:  PubMed          Journal:  J Biomed Opt        ISSN: 1083-3668            Impact factor:   3.170


  3 in total

1.  Four-wavelength near-infrared peripheral oximetry in cardiac surgery patients: a comparison between EQUANOX and O3.

Authors:  Arnaud Ferraris; Matthias Jacquet-Lagrèze; Jean-Luc Fellahi
Journal:  J Clin Monit Comput       Date:  2017-05-02       Impact factor: 2.502

2.  Combined cerebral and somatic near-infrared spectroscopy oximetry monitoring during liver surgery: an observational and non-interventional study.

Authors:  Yves Collin; Tina Hu; André Denault; Annik Fortier; William Beaubien-Souligny; Réal Lapointe; Franck Vandenbroucke-Menu
Journal:  Korean J Anesthesiol       Date:  2022-01-20

3.  Comparison of basic regional cerebral oxygen saturation values in patients of different ages: a pilot study.

Authors:  Chaohui Lian; Peng Li; Ning Wang; Yi Lu; Wangning Shangguan
Journal:  J Int Med Res       Date:  2020-08       Impact factor: 1.671

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.