Heterogeneity and scatter effects on Ir-192 brachytherapy dose distribution.

Alexander F I Osman1, Nabil Maalej2, Khateeb Ul-Rahman1, Wamied Abdel Rahman3.   

Abstract

We studied the effects of the presence of an air cavity and scatter due to patient size on dose distribution near an Iriduim-192 brachytherapy source (Ir-192). The source was modeled using Monte Carlo (MC) code MCNP5. The Radial dose, gL(r), and the anisotropy function, F(r,θ) specified by the AAPM TG-43 have been determined and compared with the consensus data (AAPM report No. 229). We compared our MC results to the measured dose distribution using an EBT3 Gafchromic® film measurement. The dose was determined in the presence of an air cavity of 3, 5, and 7mm diameters located at 2mm distance from Ir-192. The dose was also determined for Ir-192 centered in 30×30×30cm3 and 80×80×80cm3 water phantoms. The MC results of gL(r) and F(r,θ) agreed with the consensus data to within 2% and 3%, respectively. The MC and the measured dose distributions agreed well with a maximum difference of 8.2% at the periphery of the film. The dose at 10cm from the Ir-192 source with a full scattering medium (80×80×80cm3) was 7% higher compared to the dose in (30×30×30cm3) water phantom. The dose to water in the presence of a 3, 5, and 7mm diameter air cavity increased by an average of 3%, 6%, and 9%, respectively, compared to the dose with no air cavity. Ignoring scatter effects and the heterogeneity correction in the presence of an air cavity can lead to significant errors in dose delivered to patients.
Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Air cavity; Brachytherapy; Dosimetry; Film; Monte Carlo

Mesh:

Substances:

Year:  2016        PMID: 27623695     DOI: 10.1016/j.ejmp.2016.09.004

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  1 in total

Review 1.  A review of dosimetric impact of implementation of model-based dose calculation algorithms (MBDCAs) for HDR brachytherapy.

Authors:  Yousif A M Yousif; Alexander F I Osman; Mohammed A Halato
Journal:  Phys Eng Sci Med       Date:  2021-06-17
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.