Literature DB >> 27612183

Design and performance of a large lumen glaucoma drainage device.

J L Olson1, S Groman-Lupa1.   

Abstract

PurposeWe report the in vivo testing of a large-lumen glaucoma drainage (LL-GDD) device equipped with a flow regulator. The device's membrane can be non-invasively opened with laser in the postoperative period to adjust aqueous flow and intraocular pressure.MethodsThe initial LL-GDD prototypes were constructed using 22 G silicone angiocatheters cut down to size. A 10 nm PVDF membrane was then affixed to the end using cyanoacrylate. The LL-GDD was tested first in a model eye equipped with ports for infusion and pressure measurement and in New Zealand rabbits.ResultsNew Zealand white satin cross rabbits were used, two eyes receiving the LL-GDD and the two fellow eyes serving as the control group with no intervention performed. After the procedure, the IOP in the LL-GGD surgical group dropped an average of 5.5 mm Hg (P=0.001), which was maintained until the membrane laser procedure at week 5 resulting in an average IOP reduction of 1.8 mm Hg. At week 7, the average IOP in the surgical group was 11 mm Hg compared with 18 mm Hg in the control group (P<0.001). A second laser procedure was done to completely open the membrane face, which resulted in an immediate drop in the average IOP of the surgical group by another 2.7 mm Hg, which was maintained until the study termination at day 55.ConclusionsThe large-lumen glaucoma drainage device demonstrated an ability both to prevent immediate postoperative hypotony and to allow progressively lower IOP on demand in this proof-of-concept study.

Entities:  

Mesh:

Year:  2016        PMID: 27612183      PMCID: PMC5233926          DOI: 10.1038/eye.2016.182

Source DB:  PubMed          Journal:  Eye (Lond)        ISSN: 0950-222X            Impact factor:   3.775


  12 in total

1.  Application of Poiseuille's law to aqueous outflow.

Authors:  W K McEWEN
Journal:  AMA Arch Ophthalmol       Date:  1958-08

2.  Patterns of care for open-angle glaucoma in managed care.

Authors:  Allen M Fremont; Paul P Lee; Carol M Mangione; Kanika Kapur; John L Adams; Steven L Wickstrom; José J Escarce
Journal:  Arch Ophthalmol       Date:  2003-06

3.  Five-year treatment outcomes in the Ahmed Baerveldt comparison study.

Authors:  Donald L Budenz; Keith Barton; Steven J Gedde; William J Feuer; Joyce Schiffman; Vital P Costa; David G Godfrey; Yvonne M Buys
Journal:  Ophthalmology       Date:  2014-10-17       Impact factor: 12.079

Review 4.  Complications of tube implants and their management.

Authors:  Andrew K Bailey; Steven R Sarkisian
Journal:  Curr Opin Ophthalmol       Date:  2014-03       Impact factor: 3.761

5.  Blindness and glaucoma: a comparison of patients progressing to blindness from glaucoma with patients maintaining vision.

Authors:  Jessica E Oliver; Matthew G Hattenhauer; David Herman; David O Hodge; Robert Kennedy; Michael Fang-Yen; Douglas H Johnson
Journal:  Am J Ophthalmol       Date:  2002-06       Impact factor: 5.258

6.  Surgical complications in the Tube Versus Trabeculectomy Study during the first year of follow-up.

Authors:  Steven J Gedde; Leon W Herndon; James D Brandt; Donald L Budenz; William J Feuer; Joyce C Schiffman
Journal:  Am J Ophthalmol       Date:  2006-09-01       Impact factor: 5.258

7.  Randomized evaluation of the trabecular micro-bypass stent with phacoemulsification in patients with glaucoma and cataract.

Authors:  Thomas W Samuelson; L Jay Katz; Jeffrey M Wells; Yi-Jing Duh; Jane Ellen Giamporcaro
Journal:  Ophthalmology       Date:  2010-09-15       Impact factor: 12.079

Review 8.  Comparative effectiveness of treatments for open-angle glaucoma: a systematic review for the U.S. Preventive Services Task Force.

Authors:  Michael V Boland; Ann-Margret Ervin; David S Friedman; Henry D Jampel; Barbara S Hawkins; Daniela Vollenweider; Yohalakshmi Chelladurai; Darcy Ward; Catalina Suarez-Cuervo; Karen A Robinson
Journal:  Ann Intern Med       Date:  2013-02-19       Impact factor: 25.391

Review 9.  Non-penetrating filtration surgery versus trabeculectomy for open-angle glaucoma.

Authors:  Mohamed A Eldaly; Catey Bunce; Ola Z Elsheikha; Richard Wormald
Journal:  Cochrane Database Syst Rev       Date:  2014-02-15

10.  Laser Activated Flow Regulator for Glaucoma Drainage Devices.

Authors:  Jeffrey L Olson; Raul Velez-Montoya; Ramanath Bhandari
Journal:  Transl Vis Sci Technol       Date:  2014-11-03       Impact factor: 3.283

View more