Literature DB >> 27608534

Accuracy of both virtual and printed 3-dimensional models for volumetric measurement of alveolar clefts before grafting with alveolar bone compared with a validated algorithm: a preliminary investigation.

C P Kasaven1, G T McIntyre2, P A Mossey1.   

Abstract

Our objective was to assess the accuracy of virtual and printed 3-dimensional models derived from cone-beam computed tomographic (CT) scans to measure the volume of alveolar clefts before bone grafting. Fifteen subjects with unilateral cleft lip and palate had i-CAT cone-beam CT scans recorded at 0.2mm voxel and sectioned transversely into slices 0.2mm thick using i-CAT Vision. Volumes of alveolar clefts were calculated using first a validated algorithm; secondly, commercially-available virtual 3-dimensional model software; and finally 3-dimensional printed models, which were scanned with microCT and analysed using 3-dimensional software. For inter-observer reliability, a two-way mixed model intraclass correlation coefficient (ICC) was used to evaluate the reproducibility of identification of the cranial and caudal limits of the clefts among three observers. We used a Friedman test to assess the significance of differences among the methods, and probabilities of less than 0.05 were accepted as significant. Inter-observer reliability was almost perfect (ICC=0.987). There were no significant differences among the three methods. Virtual and printed 3-dimensional models were as precise as the validated computer algorithm in the calculation of volumes of the alveolar cleft before bone grafting, but virtual 3-dimensional models were the most accurate with the smallest 95% CI and, subject to further investigation, could be a useful adjunct in clinical practice.
Copyright © 2016 The British Association of Oral and Maxillofacial Surgeons. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  3D; CAD-CAM models; bone grafting; micro-CT; volumetric assessment

Mesh:

Year:  2016        PMID: 27608534     DOI: 10.1016/j.bjoms.2016.08.016

Source DB:  PubMed          Journal:  Br J Oral Maxillofac Surg        ISSN: 0266-4356            Impact factor:   1.651


  2 in total

1.  Intra- and Interobserver Reliability of Bone Volume Estimation Using OsiriX Software in Patients with Cleft Lip and Palate Using Cone Beam Computed Tomography.

Authors:  Anuraj Singh Kochhar; Maninder Singh Sidhu; Mona Prabhakar; Ritasha Bhasin; Gulsheen Kaur Kochhar; Himanshu Dadlani; Gianrico Spagnuolo; Viral Vijay Mehta
Journal:  Dent J (Basel)       Date:  2021-01-22

2.  Comparative Volume Analysis of Alveolar Defects by 3D Simulation.

Authors:  Pang-Yun Chou; Rafael Denadai; Rami R Hallac; Sarayuth Dumrongwongsiri; Wei-Chuan Hsieh; Betty Cj Pai; Lun-Jou Lo
Journal:  J Clin Med       Date:  2019-09-06       Impact factor: 4.241