Literature DB >> 27605368

Clinical Evaluation of the Immune Colloidal Gold Method for Rapid Qualitative and Quantitative Measurement of Thyroid-Stimulating Hormone as an Assay for Hypothyroidism.

Tingting Wang1, Shiwei Sheng1, Meifang Ruan2, Jing Yan2, Jianying Gu1, Yumin Jiang1, Yunchao Gao1, Hankui Lu3.   

Abstract

INTRODUCTION: The immune colloidal gold (ICG) method of measuring thyroid-stimulating hormone (TSH) is a rapid and easy-to-perform test, allowing off-site measurements. This study compared the clinical utility of the first ICG-based qualitative and quantitative TSH test methods in China with the third-generation serum TSH assay used worldwide.
METHODS: Fingertip and venous blood was collected within 30 min from 283 patients initially suspected of hypothyroidism. TSH was measured in fingertip blood using ICG-based qualitative quantitative tests. Serum TSH in venous blood was tested using the third-generation serum TSH assay. Correlations between systems were tested by kappa or Spearman correlation coefficients.
RESULTS: Compared with the third-generation serum TSH assay, the ICG-qualitative TSH test kit had a kappa coefficient of 0.86, a sensitivity of 85.00%, and a specificity of 99.38% in screening for hypothyroidism. The percentages of false negatives and false positives among all subjects were 6.38% and 0.35% respectively; the total consistency rate of the two methods was 93.26%. When compared with the third-generation serum TSH assay, the ICG-quantitative TSH analysis system had a Spearman correlation coefficient of 0.91, a sensitivity of 88.43%, and a specificity of 98.77%. The percentages of false negatives and false positives among all subjects were 4.95% and 0.71%, respectively; the total consistency rate of the two methods was 94.35%.
CONCLUSION: Both ICG-based assays are easier and faster to perform than the third-generation, laboratory-based serum TSH assay method. The ICG-based methods showed acceptable performance in the simplified screening for hypothyroidism. TRIAL REGISTRATION: ClinicalTrials.gov identifier, NCT01921452. FUNDING: Merck Serono Co., Ltd.

Entities:  

Keywords:  Hypothyroidism; Immune colloidal gold; Thyroid-stimulating hormone; Thyrotropin

Mesh:

Substances:

Year:  2016        PMID: 27605368     DOI: 10.1007/s12325-016-0401-y

Source DB:  PubMed          Journal:  Adv Ther        ISSN: 0741-238X            Impact factor:   3.845


  1 in total

1.  A Rapid Intraoperative Parathyroid Hormone Assay Based on the Immune Colloidal Gold Technique for Parathyroid Identification in Thyroid Surgery.

Authors:  Wenfei Xia; Jinjun Zhang; Wenzhuang Shen; Zhi Zhu; Zhifang Yang; Xingrui Li
Journal:  Front Endocrinol (Lausanne)       Date:  2021-04-22       Impact factor: 5.555