Literature DB >> 27591469

Inappropriate fixation of an endotracheal tube causing cuff malfunction resulting in difficult extubation.

Deb Sanjay Nag1, Devi Prasad Samaddar2.   

Abstract

We discuss a case of difficult extubation, due to inadequate deflation of the tracheal tube cuff, despite collapse of the pilot balloon, on its aspiration. This was caused by inadvertent kinking of the pilot balloon tubing due to inappropriate tape fixation of the endotracheal tube.
Copyright © 2013 Sociedade Brasileira de Anestesiologia. Published by Elsevier Editora Ltda. All rights reserved.

Keywords:  Cuff; Difficult extubation; Endotracheal; Endotraqueal; Extubação difícil; Fixation; Fixação; Manguito; Tube; Tubo

Mesh:

Year:  2013        PMID: 27591469     DOI: 10.1016/j.bjane.2013.04.009

Source DB:  PubMed          Journal:  Braz J Anesthesiol


  1 in total