Technical Note: Unified imaging and robotic couch quality assurance.

Molly C Cook1, Justin Roper2, Eric S Elder2, Eduard Schreibmann2.   

Abstract

PURPOSE: To introduce a simplified quality assurance (QA) procedure that integrates tests for the linac's imaging components and the robotic couch. Current QA procedures for evaluating the alignment of the imaging system and linac require careful positioning of a phantom at isocenter before image acquisition and analysis. A complementary procedure for the robotic couch requires an initial displacement of the phantom and then evaluates the accuracy of repositioning the phantom at isocenter. We propose a two-in-one procedure that introduces a custom software module and incorporates both checks into one motion for increased efficiency.
METHODS: The phantom was manually set with random translational and rotational shifts, imaged with the in-room imaging system, and then registered to the isocenter using a custom software module. The software measured positioning accuracy by comparing the location of the repositioned phantom with a CAD model of the phantom at isocenter, which is physically verified using the MV port graticule. Repeatability of the custom software was tested by an assessment of internal marker location extraction on a series of scans taken over differing kV and CBCT acquisition parameters.
RESULTS: The proposed method was able to correctly position the phantom at isocenter within acceptable 1 mm and 1° SRS tolerances, verified by both physical inspection and the custom software. Residual errors for mechanical accuracy were 0.26 mm vertically, 0.21 mm longitudinally, 0.55 mm laterally, 0.21° in pitch, 0.1° in roll, and 0.67° in yaw. The software module was shown to be robust across various scan acquisition parameters, detecting markers within 0.15 mm translationally in kV acquisitions and within 0.5 mm translationally and 0.3° rotationally across CBCT acquisitions with significant variations in voxel size. Agreement with vendor registration methods was well within 0.5 mm; differences were not statistically significant.
CONCLUSIONS: As compared to the current two-step approach, the proposed QA procedure streamlines the workflow, accounts for rotational errors in imaging alignment, and simulates a broad range of variations in setup errors seen in clinical practice.

Entities:  

Mesh:

Year:  2016        PMID: 27587038     DOI: 10.1118/1.4960369

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Efficient quality assurance method with automated data acquisition of a single phantom setup to determine radiation and imaging isocenter congruence.

Authors:  Hyejoo Kang; Rakesh Patel; John C Roeske
Journal:  J Appl Clin Med Phys       Date:  2019-09-19       Impact factor: 2.102

Review 2.  The impact of technology on the changing practice of lung SBRT.

Authors:  Marianne Camille Aznar; Samantha Warren; Mischa Hoogeman; Mirjana Josipovic
Journal:  Phys Med       Date:  2018-01-10       Impact factor: 2.685

3.  Development and Longitudinal Analysis of Plan-Based Streamlined Quality Assurance on Multiple Positioning Guidance Systems With Single Phantom Setup.

Authors:  Shun Zhou; Junyu Li; Yi Du; Songmao Yu; Meijiao Wang; Hao Wu; Haizhen Yue
Journal:  Front Oncol       Date:  2021-06-16       Impact factor: 6.244

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.