In-vitro impact of insertion angle on primary stability of miniscrews.

Mehdi Araghbidikashani1, Amin Golshah2, Nafiseh Nikkerdar3, Majid Rezaei4.   

Abstract

INTRODUCTION: The use of miniscrews as anchorage has become increasingly popular, but there is no consensus on the best insertion angle. The purpose of this study was to determine the impact of insertion angle on implant primary stability.
METHODS: Sixty-four miniscrews, 6 × 1.6 mm, were placed in 64 fresh bone segments from 32 sheep mandibles. The miniscrews were inserted into the bone surface at 4 angles: 30°, 45°, 60°, and 90°. The maximum resistance forces of miniscrews in pullout and shear tests were determined as primary stability using a universal testing machine.
RESULTS: In the pullout tests, the 90° insertion angle had the highest primary stability. Oblique angulations of the miniscrews resulted in lower resistance. In the shear tests, angulations of the miniscrews from 90° to 45° led to increased primary stability. A more oblique insertion (30°) resulted in reduced primary stability.
CONCLUSIONS: The direction of the applied force had a significant impact on the primary stability of miniscrews at various insertion angles. The highest primary stability values were achieved at 45° when the miniscrews were loaded by shear force and at 90° when pullout force was applied.
Copyright © 2016 American Association of Orthodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27585771     DOI: 10.1016/j.ajodo.2016.02.020

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  4 in total

1.  Effect of steam and dry heat sterilization on the insertion and fracture torque of orthodontic miniscrews.

Authors:  Shiva Alavi; Farnaz Asadi; Seyed Amir Hossein Raji; Soroor Samie
Journal:  Dent Res J (Isfahan)       Date:  2020-05-23

Review 2.  Revisiting the Complications of Orthodontic Miniscrew.

Authors:  Van Mai Truong; Soyeon Kim; Jaeheon Kim; Joo Won Lee; Young-Seok Park
Journal:  Biomed Res Int       Date:  2022-08-01       Impact factor: 3.246

3.  Anatomical consideration for optimal position of orthodontic miniscrews in the maxilla: a CBCT appraisal.

Authors:  Maha S Al Amri; Hanadi M Sabban; Doaa H Alsaggaf; Fahad F Alsulaimani; Ghassan A Al-Turki; Mohammad S Al-Zahrani; Khalid H Zawawi
Journal:  Ann Saudi Med       Date:  2020-08-06       Impact factor: 1.526

4.  Effect of miniscrew insertion angle in the maxillary buccal plate on its clinical survival: a randomized clinical trial.

Authors:  Amin Golshah; Kimia Gorji; Nafiseh Nikkerdar
Journal:  Prog Orthod       Date:  2021-08-02       Impact factor: 2.750

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.