Jong Eun Park1, Ji Youn Kim2, Sun Ae Yun2, Myoung Keun Lee1, Hee Jae Huh3, Jong Won Kim1, Chang Seok Ki4.
Abstract
Standardized cytomegalovirus (CMV) DNA quantification is important for managing CMV disease. We evaluated the performance of the Real-Q CMV Quantification Kit (Real-Q assay; BioSewoom, Korea) using whole blood (WB), with nucleic acid extraction using MagNA Pure 96 (Roche Diagnostics, Germany). Real-time PCR was performed on two platforms: the 7500 Fast real-time PCR (7500 Fast; Applied Biosystems, USA) and CFX96 real-time PCR detection (CFX96; Bio-Rad, USA) systems. The WHO international standard, diluted with CMV-negative WB, was used to validate the analytical performance. We used 90 WB clinical samples for comparison with the artus CMV RG PCR kit (artus assay; Qiagen, Germany). Limits of detections (LODs) in 7500 Fast and CFX96 were 367 and 479 IU/mL, respectively. The assay was linear from the LOD to 10⁶ IU/mL (R² ≥0.9886). The conversion factors from copies to IU in 7500 Fast and CFX96 were 0.95 and 1.06, respectively. Compared with the artus assay, for values <1,000 copies/mL, 100% of the samples had a variation <0.7 log₁₀ copies/mL; >1,000 copies/mL, 73.3% and 80.6% of samples in 7500 Fast and CFX96, respectively, had <0.5 log₁₀ copies/mL. The Real-Q assay is useful for quantifying CMV in WB with the two real-time PCR platforms.Entities:
Keywords: Cytomegalovirus; Quantification; Real-time PCR; Validation; WHO International Standard; Whole blood
Mesh:
Substances:
Year: 2016 PMID: 27578516 PMCID: PMC5011116 DOI: 10.3343/alm.2016.36.6.603
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Limits of detection, limits of quantification, and conversion factors for the MagNA Pure 96 system coupled to two distinct amplification platforms
| Amplification platform | Limit of detection (IU/mL) | Limit of quantification (IU/mL) | Conversion factor (IU/copies) |
|---|---|---|---|
| 7500 Fast real-time PCR system | 367 | 367 | 0.95 |
| CFX96 real-time PCR detection system | 479 | 479 | 1.06 |
Fig. 1Linearity of the Real-Q cytomegalovirus (CMV) Quantification kit on different real-time PCR platforms: (A) the 7500 Fast real-time PCR system and (B) the CFX96 real-time PCR detection system.
Fig. 2Bland-Altman plot comparing the artus cytomegalovirus (CMV) RG PCR kit and the Real-Q CMV Quantification kit on different real-time PCR platforms: (A) the 7500 Fast Real-time PCR system, (B) the CFX96 real-time PCR detection system. Solid lines are the mean differences between the values; dashed lines are the mean difference plus or minus 1.96 SD (95% confidential interval of mean difference).