Literature DB >> 27577512

Feasibility of spectral imaging with low-concentration contrast medium in abdominal CT angiography of obese patients.

Yijun Liu1, Ailian Liu1, Lei Liu1, Shifeng Tian1, Jinghong Liu1, Renwang Pu1, Xin Fang1, Xiaofeng Liu1, Gang Yuan1.   

Abstract

OBJECTIVE: To explore the application value of dual-energy spectral imaging scanning with low radiation dose and low-concentration contrast medium (270 mg I/mL) in abdominal CT angiography (CTA) of obese patients.
METHODS: A total of 127 obese patients (BMI≥25 kg/m(2) and waist circumference ≥900 mm) referred for abdominal CTA were prospectively enrolled in the study. The patients were divided into two groups; in group A (n=69), a spectral imaging scan mode and 270 mg I/mL iodine concentration contrast medium was used, and in group B (n=58), a conventional imaging scan mode using 120 kVp and 350 mg I/mL iodine concentration contrast medium was used. The image quality of the right renal artery in the two groups was evaluated by two observers using a 5-point scale, and the scores were compared using the Mann-Whitney U-test; the inter-observer agreement for the scores was analysed using the Kappa test. The CT values of the abdominal aorta, the superior mesenteric artery, the common hepatic artery and the splenic artery, and the CT value and standard deviation (SD) of the erector spinae at the level of the right renal hilum in groups A and B were measured by two observers; the inter-observer agreement of the measurement data was analysed using the inter-class correlation coefficient test. The following parameters were compared between the two groups using an independent sample t-test: the CT values of the abdominal aorta and its main branches; the image contrast-to-noise ratio (CNR) and figure of merit (FOM) of the abdominal aorta; the CT dose index (CTDIvol ); the dose length product (DLP); and the total iodine intake of the patients. P<.05 suggested a statistically significant difference.
RESULTS: The image scores of the right renal artery in groups A (4.59±0.60) and B (4.53±0.63) were the same (P=.57), with good inter-observer agreement. The CT values of the abdominal aorta, the superior mesenteric artery, the common hepatic artery and the splenic artery were >300 HU in both the groups; there was no statistically significant difference between the two groups (all P>.05), and inter-observer agreement was also good. Group A had significantly higher CNR and FOM values in the abdominal aorta than group B (all P<.001). Compared with group B, the CTDIvol and DLP values in group A were decreased by 46% and 35%, respectively. The total iodine intake for patients in group A was 27 g, 23% lower than the 35 g intake for patients in group B.
CONCLUSION: Compared with conventional CT, spectral CT imaging significantly reduces both radiation dose and contrast dose while maintaining image quality in abdominal CTA for patients with central obesity and high BMI.
© 2016 John Wiley & Sons Ltd.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27577512     DOI: 10.1111/ijcp.12856

Source DB:  PubMed          Journal:  Int J Clin Pract        ISSN: 1368-5031            Impact factor:   2.503


  3 in total

1.  Aortic CT angiography using the double region of interest timing bolus technique: feasibility of 80 kVp scanning in lean patients.

Authors:  Nobuo Tomizawa; Shingo Ito; Tastuya Nakao; Hiroaki Arakawa; Kodai Yamamoto; Shinichi Inoh; Takeshi Nojo; Sunao Nakamura
Journal:  Int J Cardiovasc Imaging       Date:  2019-07-02       Impact factor: 2.357

2.  Automatic spectral imaging protocol and iterative reconstruction for radiation dose reduction in typical hepatic hemangioma computed tomography with reduced iodine load: a preliminary study.

Authors:  Wei Li; Aiyin Li; Bin Wang; Xiuyuan Niu; Xin Cao; Xinyi Wang; Hao Shi
Journal:  Br J Radiol       Date:  2018-05-17       Impact factor: 3.039

3.  Mixed reality models based on low-dose computed tomography technology in nephron-sparing surgery are better than models based on normal-dose computed tomography.

Authors:  Guan Li; Zhiqiang Cao; Jinbao Wang; Xin Zhang; Longjiang Zhang; Jie Dong; Guangming Lu
Journal:  Quant Imaging Med Surg       Date:  2021-06
  3 in total