| Literature DB >> 27562965 |
Jon B Stephenson1, Melanie L Flater2, Lisa T Bain2.
Abstract
The Georgia Bureau of Investigation utilized a silylation method of analysis for low molecular weight carboxylic acids in the past. Due to the negative impact such derivatizations can have on gas chromatography-mass spectrometry (GC-MS) systems an alternative means of analysis was investigated. The described method is a whole blood solid phase extraction of valproic acid, salicylic acid and ibuprofen utilizing butylation for sensitivity and improved chromatography by GC-MS. The method produced a limit of detection and limit of quantitation at 1 mg/L for valproic acid, 2 mg/L for salicylic acid and 0.25 mg/L for ibuprofen. The variability based upon the middle of the calibration curve estimated to be 7% for valproic acid, 8% for salicylic acid and 11% for ibuprofen established upon a 95% confidence interval, with the highest percent coefficient of variation being 5.3% for ibuprofen.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27562965 DOI: 10.1093/jat/bkw079
Source DB: PubMed Journal: J Anal Toxicol ISSN: 0146-4760 Impact factor: 3.367