Huijian Ge1, Hengwei Jin1, Youxiang Li1, Xianli Lv2.
Abstract
A 56-year-old woman was admitted to stent-assisted coiling for a 2-mm A1 aneurysm of the left anterior cerebral artery and a left 3-mm internal carotid artery aneurysm. While coiling the A1 aneurysm, the first 2 mm × 20 mm coil migrated through the 4.5 mm × 37 mm Enterprise stent struts, lodging at the distal anterior cerebral artery. A 4 mm × 15 mm Solitaire AB stent was used successfully in this case to remove the displaced coil. The A1 aneurysm was re-treated with a 2 mm × 40 mm coil after placement of the Enterprise stent, and the ophthalmic ICA aneurysm was also coiled through the stent struts. The patient was neurologically intact after treatment.Entities:
Keywords: Coil migration; Solitaire AB stent; embolisation; intracranial aneurysm
Mesh:
Year: 2016 PMID: 27558991 PMCID: PMC5131764 DOI: 10.1177/1971400916666556