Yoon-Soo Jang1, Guochao Wang1,2, Sangwon Hyun1, Hyun Jay Kang1, Byung Jae Chun1, Young-Jin Kim1, Seung-Woo Kim1. 1. Department of Mechanical Engineering, Korea Advanced Institute of Science and Technology (KAIST), Science Town, Daejeon, 305-701, South Korea. 2. Department of Instrument Science and Technology, College of Mechatronic Engineering and Automation, National University of Defence Technology (NUDT), Hunan, Changsha, 410073, China.
Abstract
A prototype laser distance interferometer is demonstrated by incorporating the frequency comb of a femtosecond laser for mass-production of optoelectronic devices such as flat panel displays and solar cell devices. This comb-referenced interferometer uses four different wavelengths simultaneously to enable absolute distance measurement with the capability of comprehensive evaluation of the measurement stability and uncertainty. The measurement result reveals that the stability reaches 3.4 nm for a 3.8 m distance at 1.0 s averaging, which further reduces to 0.57 nm at 100 s averaging with a fractional stability of 1.5 × 10(-10). The uncertainty is estimated to be in a 10(-8) level when distance is measured in air due to the inevitable ambiguity in estimating the refractive index, but it can be enhanced to a 10(-10) level in vacuum.
A prototype laser distance interferometer is demonstrated by incorporating the frequency comb of a femtosecond laser for mass-production of optoelectronic devices such as flat panel displays and solar cell devices. This comb-referenced interferometer uses four different wavelengths simultaneously to enable absolute distance measurement with the capability of comprehensive evaluation of the measurement stability and uncertainty. The measurement result reveals that the stability reaches 3.4 nm for a 3.8 m distance at 1.0 s averaging, which further reduces to 0.57 nm at 100 s averaging with a fractional stability of 1.5 × 10(-10). The uncertainty is estimated to be in a 10(-8) level when distance is measured in air due to the inevitable ambiguity in estimating the refractive index, but it can be enhanced to a 10(-10) level in vacuum.
Laser distance interferometers have been playing key roles for the advance of precision machines for large-volume manufacture of nanotechnology-based products such as semiconductor integrated circuits, flat panel displays and solar photovoltaic panels123. Such interferometers rely on continuous-wave lasers, measuring distance or length by accumulating the interferometric phase with sub-nanometer resolutions uninterruptedly over ranges up to several meters4. The laser wavelength used for distance measurement is calibrated to standard lasers stabilized to atomic absorption lines as recommended by the Bureau International des Poids et Mesures (BIPM) for realization of the current definition of the meter567. Recently the frequency comb of mode-locked femtosecond lasers (hereafter simply referred to as ‘comb’) began to be adopted as a new wavelength ruler, permitting calibration of the laser wavelength with direct traceability to the radio-frequency standard89 or an optical clock10. In addition, the comb can be used for simultaneous generation of multiple arbitrary wavelengths; by phase-locking tunable laser diodes individually to selected optical modes within the comb11 or by filtering out several optical modes separately with power amplification by injection-locking12. This approach of comb-referenced wavelength generation offers excellent frequency stability and uncertainty, allowing for implementation of advanced techniques such as absolute distance measurement by multi-wavelength interferometry131415161718 and real-time compensation of the refractive index of air by two-colour interferometry1920.In fact, during the past decade, many attempts have been made to use the comb as the light source for long-distance measurements with various progressive principles; inter-mode synthetic wavelength interferometry212223, many-wavelength dispersive interferometry24252627, multi-wavelength interferometry131415161718, dual-comb frequency-down-conversion interferometry28293031, and optical cross-correlation time-of-flight measurement323334. These newly proposed methods proved effective in extending the non-ambiguity distance range by taking advantage of the comb’s unique spectral or temporal characteristics. Nonetheless, for practical use in precision positioning and dimensional measurement, they need to be appraised systematically in terms of the stability and uncertainty in consideration of all possible error sources.Our work described here is concerned with the method of comb-referenced multi-wavelength interferometry. A prototype interferometer system is used to demonstrate its competence as an absolute-type positioning transducer. As illustrated in Fig. 1, the interferometer employs four different wavelengths concurrently phase-locked to the comb of an Er-doped fiber laser. The absolute position is determined through synchronized phase detection along with parallel data-processing of the multiple interference phases so as to enable fast, precise and stable measurements continuously over a few meters of axis-travel. In fact, the interferometer system was previously proposed in ref. 18 by the authors together with primary measurement results. In this study, with further experimental data obtained via more elaborate environment control, comprehensive analysis is made to validate the proposed interferometer system as a practical means of realizing the definition of the meter with an utmost level of uncertainty and stability.
Figure 1
Comb-referenced multi-wavelength interferometry for realization of distance measurement with improved uncertainty evaluation.
Results
System configuration
Figure 2 illustrates the comb-referenced multi-wavelength interferometer system constructed in this study. The light source comprises four distributed feedback (DFB) diode lasers, each being phase-locked to a distinct wavelength within the comb of an Er-doped fiber femtosecond laser of a 1550 nm center wavelength. The comb used as the wavelength ruler is stabilized to the Rb atomic clock by phase-locked loop (PLL) control of the repetition rate (f) precisely to 100 MHz and also the carrier-envelop-offset frequency (f) to 30 MHz. The f is extracted using a fiber-type f-2f interferometer3536. The four DFB laser output beams, each having an optical power of 10 mW, are combined into a single-mode fiber so that they propagate along the same optical path together all the way within the interferometer. For heterodyne interferometric phase detection, the combined DFB lasers are divided into two fibers, one of which is connected to a pair of acousto-optic modulator (AOM) to provide a frequency-shift of 40 kHz. The configured interferometer optics is basically of Michelson type, employing four non-polarizing beam splitters so as to divide and also recombine the reference and measurement beams without the polarization leakage causing non-linearity errors. Once the original and frequency-shifted beams have been recombined after round-trip to the reference and measurement arms, the DFB lasers are separated using fiber-Bragg-grating (FBG) arrays so that the interferometric phase of each laser wavelength is detected using a multi-channel phase meter of 0.1° resolution18.
The light source, in addition to the four DFB lasers, incorporates a continuous-wave (CW) laser of a 1530 nm wavelength. This near-infrared CW laser is intensity-modulated with a 40 MHz rf frequency to produce a long synthetic wavelength of 7.50 m. This extra source is used to perform coarse phase-measuring interferometry to provide an initial estimate of the target distance needed for the process of multi-wavelength interferometry. In addition, for comparison purpose, a commercial HeNe laser interferometer of incremental distance measurement is installed via a dichroic mirror along the same optical path of the multi-wavelength interferometer. The whole interferometer system is shielded to prevent drastic temperature variation, while being set up on a granite frame seated on a passive anti-vibration foundation. The target mirror is a corner-cube type mounted on an aerostatic-bearing stage moving along the granite guideway. By monitoring the ambient parameters of the air temperature, relative humidity, air pressure and CO2 concentration, the refractive index of air is estimated for real-time wavelength compensation of the DFB lasers using the Ciddor’s formula37.
Wavelength generation
For proper implementation of multi-wavelength interferometry, the four wavelengths employed as the light source have to be carefully selected. For convenience, the wavelengths in terms of their vacuum values are denoted as λ with i being 1, 2, 3, 4 and λ1 < λ2 < λ3 < λ4. As depicted in Fig. 3a, the selection rule adopted in this study is to maximize the non-ambiguity range (NAR) that is half of the synthetic wavelength Λ1234 defined by the equation of 1/Λ1234 = |1/Λ12 − 1/Λ34| = |(1/λ1 − 1/λ2) − (1/λ3 − 1/λ4)|. In order to make Λ1234 large, the separation between λ1 and λ2, and also between λ3 and λ4 are narrowed. Besides, λ1 and λ4 are placed far apart so that an intermediate synthetic wavelength Λ14 is created in between Λ12 and λ1 or λ4. The comb spectrum used as the reference in this study for wavelength stabilization has a 100 MHz mode-to-mode spacing, but the permissible mini-mum frequency separation is practically limited by the 100 GHz channel spacing of the FBGAs used for wavelength separation and recombination. Accordingly, λ1 and λ2 are located at 1530.279693 nm and 1531.040888 nm, generating a synthetic wavelength Λ12 of ~3.4 mm. Similarly, λ3 and λ4 are placed in the higher wavelength end of the comb spectrum, at 1554.179409 nm and 1554.937151 nm, respectively. In consequence, as illustrated in Fig. 3a, the resulting NAR extension can explained in three steps: First, the far-separated λ1 and λ4 makes a synthetic wavelength Λ14 = λ1λ4/(λ1 − λ4) being worked out to be 90 μm. Second, the closely located pair of λ1 and λ2 creates a longer synthetic wavelength Λ12 of 3.4 mm. The other pair of λ3 and λ4 yields a similar synthetic wavelength Λ34 of ~3.4 mm. Third, combing all four wavelengths leads to the largest synthetic wavelength Λ1234 of 90 mm, offering an overall NAR of 45 mm.
Figure 3
Four-wavelength-combined light source.
(a) Measured optical spectrum and estimation of non-ambiguity range (NAR) of the combined light source. (b) Frequency stability measurement result.
Figure 3a shows the optical spectrum of the light source comprising the four wavelengths actually produced with a signal-to-noise ratio of 50 dB in this study. As already explained, the comb was stabilized to the Rb clock by PLL control of the repetition rate (f) and the carrier-envelop-offset frequency (f). The optical frequency generated for each wavelength is represented by f = c/λ with c being the speed of light in vacuum. After the comb-referenced stabilization, the frequency can be expressed as f = N × f + f + f in which N is a large integer and f denotes the beat frequency with the optical mode designated within the comb11. The expression leads to the first-order frequency variation Δf in a fractional form of Δf/f = [(Δf/f)2 + (Δf/f)2 + (Δf/f)2 + (Δf/f)2]1/2 with f being the Rb atomic clock frequency. Figure 3b shows the temporal stability of each frequency term actually measured in terms of the Allan deviation with increasing the averaging time from 1 to 500 s. Note that the Δf was measured by monitoring the beat signal of the Rb clock in use with another Rb clock of the same kind. Other Δf, Δf and Δf were monitored using a frequency counter referenced to the Rb clock. The stability test result indicates that the Δf/f term is the most dominant instability factor, being 2.57 × 10−11 at 1 s averaging. All other terms, Δf/f, Δf/f and Δf/f, are found smaller than Δf/f, making no significant contributions to the overall stability of Δf/f. Since Δf/f = Δλ/λ, the fractional stability obtained in frequency is directly transferred to the fractional stability in wavelength. Note that the frequency instability causes random errors, influencing the measurement repeatability. Complete estimation of the frequency uncertainty demands that not only the random frequency stability but also the systematic frequency offset of the Rb clock be taken into consideration together as will be discussed in detail later.
Absolute distance determination
For each wavelength λ, the refractive index of air n is estimated using the Ciddor’s equation in this study. Hence, the target distance L can be expressed as L = (λ/2n)(m + e) with m and e being an integer and an excess fraction, i.e. 0 ≤ e < 1, respectively. The excess fraction e is determined one by one directly from the interferometric phase measured for each λ. Next, the integer mi is decided so as to satisfy the inequality constraint of |L − L| < α(λ/2) for all λ simultaneously with α being a positive constant selected as small as 1/100. This process is implemented in real time by numerical iteration using LabVIEW software (National Instruments) by adopting the conventional algorithm of the excess fraction method38, or smart algorithms devised to minimize the computation time in dealing with a large group of mi candidates3940. Convergence to a unique true solution of m requires that the uncertainty of L should be less than a quarter of λ, i.e. ΔL < λ/4. Otherwise, the numerical computation would fail to converge at all or reach a false solution that deviates from the true solution with a large error. If abnormal convergence happens, the calculated integer value of mi shows a drastic change from its previous value. This situation can be ruled out using a simple criterion. i.e. |Δm| between two consecutive calculations should be either 0 or 1 for every λ.Beyond the non-ambiguity range (NAR4) of 45 mm, the measurement range is extended by operating the cw laser added in the light source. The cw laser is intensity-modulated with a 40 MHz frequency so that radio-frequency synthetic wavelength interferometry (RFSWI) is implemented with a prolonged non-ambiguity range (NARRFSWI) of 3.75 m as illustrated in Fig. 4. The RFSWI permits coarse measurement of L, which is subsequently taken as the initial guess of L for the numerical computation of multi-wavelength interferometry to accurately determine m for each λ. The NAR extension demands that the RFSWI measurement uncertainty be less than a half of NAR4, i.e. <22.5 mm over the entire range of NARRFSWI. This uncertainty requirement was confirmed using the HeNe laser interferometer installed in parallel with the multi-wavelength light source as explained in Fig. 2, revealing that the RFSWI offers an accuracy of 7.7 mm in peak-to-valley as presented in Fig. 4b.
Figure 4
Non-ambiguity range (NAR) extension with radio-frequency synthetic wavelength interferometry (RFSWI).
(a) Synthetic wavelength chain with uncertainty requirement. (b) Comparative measurement result between the RFSWI and cw HeNe laser interferometer.
Measurement results
Figure 5 presents a test result in which the comb-referenced multi-wavelength interferometer built in this study was operated from 1.0 to 3.8 m in steps of 150 mm. At each step, the stage was brought to a complete stop on the granite guideway; the motion control servo and also the aerostatic bearing were turned off so that distance readings were taken without disturbance from servo-induced motion jitters or turbulence-induced bearing vibration. Figure 5a shows a typical time-dependent variation of the measured distance of L1 from λ1 over a 50 s period monitored at an updated 100 Hz rate. The stage holding the target mirror was positioned at a 3.8 m distance. The standard deviation of the repeated measurements of L1 was 20 nm, while the moving average of 100 consecutive measurements yields a peak-to-valley variation of 42 nm (standard deviation: 7.8 nm). The L1 measurement data is Fourier-transformed as shown in Fig. 5b; the high frequency fluctuation seen in the 20–40 Hz range is estimated to arise from random mechanical vibration existing on the granite guideway, while the low frequency disturbance below 1 Hz is attributable to the slow environmental drift of the refractive index of air and also the thermal expansion of the granite guideway due to the ambient temperature change.
Figure 5
Performance test result.
(a) Time-dependent variation of the measured distance L1 over 50 s. (b) Fourier-transformed data of L1. (c) Allan deviation of L1. (d) Linearity comparison with an incremental HeNe laser interferometer. Inter-comparisons between individual distances of four wavelengths are also shown.
Further, the effect of random noise on the measurement stability is analysed by calculating the Allan deviation of the measurement data as a function of the averaging time as plotted in Fig. 5c; the measurement stability, or repeatability, improves better than 1 nm as the averaging time increases from 0.01 to 1.0 s. With further increasing the averaging time, the measurement stability begins to worsen in proportion because the slow varying environmental drift becomes accumulated while fast varying noise is cancelled out. It is also noted in Fig. 5c that when the distance update rate is slowed down from 100 Hz to 1 Hz, the measurement stability worsens, by an order of magnitude, due to ineffective suppression of high frequency random noise. Lastly, Fig. 5d shows a comparison of the distance L1 with the accumulated distance of the HeNe laser interferometer over the absolute distance range of 1.0–3.8 m. Note that the HeNe laser interferometer reading has a zero-datum offset of ~1.00 m from the absolute distance reading. The comparison was made at every 150 mm step of the stage movement by averaging 300 measurements taken at a 100 Hz update rate. The result reveals a peak-to-valley discrepancy of 35.3 nm, which corresponds to a ±4.6 × 10−9 linearity error between the two interferometers over an entire range of 3.0 m. Figure 5(d) also shows comparisons between absolute distances of L1, L2, L3 and L4 obtained the from multi-wavelength interferometer. With distance differences of L1 − L2, L1 − L3 and L1 − L4 lying within a 20 nm peak-to-valley bound, the maximum linearity error is worked out to be ±2.6 × 10−9 between the measured distances from individual wavelengths. The inter-wavelength differences are reckoned to arise from the imperfect compensation of the refractive index of the ambient air due to the 10−8 level uncertainty of the Ciddor’s equation formula.Figure 6 shows another test in which the measured distance was monitored over a long period of 12 hours at a 1 Hz update rate. The stage holding the target mirror was positioned stationary at the farthest end of a 3.8 m distance throughout the test measurement. When the absolute distances L1, L2, L3 and L4 are plotted together, they all appear to follow a common long-term variation of gradual decrease by ~5 μm. The distance decrease was caused by a temperature change of 0.3 °C that was monitored during the entire test period as in Fig. 6a. Despite the temperature variation, the individual distances of L1, L2, L3 and L4 shows no significant difference, so they are not well distinguishable from each other as plotted in Fig. 6b. The inter-wavelength distance differences of L3 − L2 and L4 − L1 remain nearly constant, without a notable long-term drift, within 12.8 nm and 7.1 nm in terms of 1,000-point averaging, respectively. When the differences are evaluated in terms of the Allan deviation as shown in Fig. 6c, the stability is 3.4 nm and 4.0 nm at 1 s averaging with a fractional stability of 8.9 × 10−10 and 1.1 × 10−9, respectively. The small discrepancy between L3 − L2 and L4 − L1 is attributable to the phase detection error that is random and independent of wavelengths as discussed in the next section of uncertainty evaluation. The stability reaches 0.57 nm and 0.78 nm at 100 s averaging, respectively, which corresponds to 1.5 × 10−10 and 2.1 × 10−10 of fractional stability. This result implies that even though the Ciddor’s formula is limited to a 10−8 level of uncertainty in estimating air refractive indices, the stability of distance measurement can be achieved better by about two orders of magnitude by properly adjusting the averaging time. However, if the averaging time is taken unnecessarily long, the measurement stability weakens due to the long-term drift of the environment monitoring sensors and also the phase-detection electronics.
Figure 6
Long-term test result.
(a) Time-dependent variation of the distances measured from the four wavelengths along with the stage temperature change over 12 hours. (b) Distance differences between four wavelengths. (c) Allan deviation of distance differences between wavelengths.
Uncertainty evaluation
Table 1 lists the uncertainty evaluation made for the multi-wavelength interferometry described so far. As derived in [ref. 16], the uncertainty u(L) of the distance measured by optical wave-length interferometry is contributed by three major terms; u(L) = [{u(e) × (λ/2)}2 + {u(n) × L}2 + {u(f) × L}2]1/2. The first term represents the electronic error with u(e) being the uncertainty of phase detection. Using a well-calibrated function generator, the multi-channel phase meter used in our interferometer system was assessed to have a maximum random error of 0.3° over the entire 360° range, i.e. u(e) = ~0.00083, bringing about a 0.64 nm distance error. This phase-induced error is not affected by increasing L, thus the least significant except when measuring short distances with sub-nm resolutions. The next important contribution is concerned with the frequency uncertainty u(f) of the light source. The random error caused by the frequency instability was estimated to be ~2.57 × 10−11 at 1 s averaging as discussed in the previous section. In addition, the systematic frequency offset of the Rb clock was provided by the manufacturer to be 5.00 × 10−11 as listed in Table 1. Thus, by combining the random error with the systematic offset, the frequency uncertainty u(f) is calculated as 5.63 × 10−11. This implies that the comb-reference wavelength generation performed in this work makes the frequency uncertainty causes no larger errors than the previous phase error until the measured distance L exceeds hundreds of meters.
Table 1
Uncertainty evaluation of distance measurement.
Uncertainty sources
Uncertainty
Uncertainty for phase detection
8.34 × 10−4 × λ/2 (0.64 nm)
Phase offset (systematic)
8.33 × 10−4 × λ/2
Repeatability (random)
2.94 × 10−5 × λ/2
Uncertainty for refractive index
1.62 × 10−8 × L
Temperature of air (ΔT = 5 mK)
4.56 × 10−9 × L
Pressure of air (ΔP = 2.5 Pa)
6.52 × 10−9 × L
Humidity of air (ΔH = 1%)
8.79 × 10−9 × L
CO2 concentration (ΔPCO2 = 41 ppm)
4.67 × 10−9 × L
Ciddor’s formula
1.00 × 10−8 × L
Uncertainty for frequency
5.63 × 10−11 × L
Frequency offset of fRb (systematic)
5.00 × 10−11 × L
Stability of fRb (random)
2.57 × 10−11 × L
Stability of fr(random)
3.08 × 10−12 × L
Stability of fceo(random)
1.16 × 10−14 × L
Stability of fb(random)
8.49 × 10−14 × L
Combined standard uncertainty (k = 1)
[(0.64 nm)2 + (1.62 × 10−8 × L)2]1/2
It is now obvious that the most dominant contribution comes from the uncertainty u(n) related to the refractive index of air. This environment-dependent uncertainty has two distinct error sources; one is the inaccuracy arising in monitoring the ambient conditions and the other is the uncertainty of the dispersion formulas available for the air. The Ciddor’s equation or updated Edlen’s equation is deliberated to offer a 10−8 level uncertainty due to empirical-analytical limitations. In this work, adopting the state-of-the-art precision sensors permitted observing the ambient temperature within an uncertainty of 5 mK, the air pressure within 2.5 Pa, the air humidity within 1% and the carbon dioxide con-centration within 41 ppm. This precision enabled compensation of the refractive index of air to the extent of making the most of the 10−8 uncertainty, leaving the air dispersion formulas as the ultimate limit. In consequence, the combined uncertainty (k = 1) of the measured distance L is worked out to be u(L) = [(0.64 nm)2 + (1.62 × 10−8 × L)2]1/2, which estimates a total error of 62 nm for our experiments performed over a target distance of 3.8 m.Another important issue related to the uncertainty evaluation is the process reliability of multi-wavelength interferometry for absolute distance measurement. As discussed in the previous section, convergence to the true solution of L is guaranteed only when the uncertainty is less than a quarter of λ, i.e. u(L) < λ/4. This condition implies that the maximum measurable distance L is constrained as u(L) scales with L as depicted in Fig. 7. Based on the uncertainty data given in Table 1, L in air corresponds to 23.6 m for a 68% level of confidence (k = 1), being limited mainly by the uncertainty of estimating the refractive index of air. On the other hand, L in vacuum extends to 6.8 km as it is affected only by the uncertainty of the light source frequency. This reasoning supports the potential of the comb-referenced multi-wavelength interferometry in outer space use for various applications such as multiple satellites operation in formation41. Further improvements in the uncertainty of the source frequency may be made by several orders of magnitudes by stabilizing the reference comb to well-certified optical clocks42.
Figure 7
Uncertainty as a function of the distance to be measured in air.
The λ/4-uncertainty limit line for multi-wavelength interferometry is shown to estimate the maximum measureable distance in air and vacuum.
Discussions
The multi-wavelength interferometer proposed and tested in this study is found capable of performing distance measurements with several fundamental advantages over conventional laser interferometers. First, the wavelengths generated with reference to the comb being stabilized to the Rb clock are highly stable and also accurate within an uncertainty level of 10−11, which is not readily achievable by traditional lasers used in distance interferometry. Second, the comb-referenced wavelength generation allows for reliable absolute distance measurements with a minimum number of wavelengths optimally selected for an extended non-ambiguity range well suited for precision machine-axis control and length metrology. Third, a comprehensive uncertainty evaluation can be made on the overall measurement performance with systematic identification of optical, electrical and environmental error sources.The experimental work conducted in this study by measuring distances up to 3.8 m demonstrates a 0.57 nm repeatability and a 10 nm linearity, equivalent to 1.5 × 10−10 and 2.6 × 10−9 in fractional terms, respectively. The combined measurement uncertainty is estimated to be 1.62 × 10−8, which is mainly contributed by the ambiguity in estimating the refractive index of the ambient air. In vacuum environment, the measurement uncertainty is expected to improve to a 10−10 level which is accurate, stable enough for mass-production of optoelectronic devices with direct traceability to the atomic clock. It is also anticipated that the proposed method could be exploited for outer space scientific applications such as gravitation wave detection in the near future with further improvements on frequency stabilization by adopting optical clocks and also interferometric phase detection by incorporating high performance electronics.
Additional Information
How to cite this article: Jang, Y.-S. et al. Comb-referenced laser distance interferometer for industrial nanotechnology. Sci. Rep. 6, 31770; doi: 10.1038/srep31770 (2016).