Scanning Electron Microscopy Evaluation of an EX-PRESS Mini Glaucoma Shunt After Explantation.

Daniele Tognetto1, Paolo Cecchini, Rossella D'Aloisio, Odilla Vattovani, Gianluca Turco.   

Abstract

We report a case of an explanted stainless steel miniature glaucoma drainage device (EX-PRESS) implanted under a scleral flap for pseudoexfoliation open-angle glaucoma surgical treatment. The glaucoma shunt was implanted in a 75-year-old white man with medically refractive glaucoma. Cataract surgery was performed simultaneously. After 2 years, the shunt extruded through the scleral flap and the conjunctiva and it was, therefore, explanted. Scanning electron microscopy images of the EX-PRESS mini glaucoma shunt were acquired to verify the patency of the device lumen and the presence of fibrosis or cellular adhesion on the device. Energy dispersive spectroscopy for chemical surface characterization of the EX-PRESS shunt was performed. Scanning electron microscopy-acquired images showed minimal extracellular material proliferation on the lumen device. The energy dispersive spectroscopy analysis revealed a high peak of carbon suggesting the organic nature of the residuals found on the shunt lumen. The surface showed few superficial pits, likely due to an initial corrosion process.

Entities:  

Mesh:

Year:  2017        PMID: 27552510     DOI: 10.1097/IJG.0000000000000521

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  2 in total

1.  Factors influencing the filtration-bleb volume after Ex-PRESS® surgery.

Authors:  Naoki Tojo; Atsushi Hayashi; Mitsuya Otsuka
Journal:  Clin Ophthalmol       Date:  2018-09-05

2.  Xen45 gel stent ab interno trimming for ostium occlusion: case report.

Authors:  Filippo Tatti; Pietro Gentile; Lorenzo Mangoni; Giuseppe Demarinis; Pietro Napoli; Maurizio Fossarello
Journal:  BMC Ophthalmol       Date:  2021-12-27       Impact factor: 2.209

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.