Evaluation of a novel device, high-intensity focused ultrasound with a contact cooling for subcutaneous fat reduction.

Heun Joo Lee1, Mi-Hye Lee1, Su Gun Lee2, Un-Cheol Yeo3, Sung Eun Chang4.   

Abstract

BACKGROUND: Non-invasive devices for fat reduction involving high-intensity focused ultrasound (HIFU) are attracting attention. HIFU can deliver energy to the desired depth and can ablate subcutaneous adipose tissue (SAT), but purpura and pain may still limit its use.
OBJECTIVES: The aim of this study was to investigate the effects of a novel HIFU device for fat destruction with a contact cooling system compared to HIFU without contact cooling.
METHODS: A group of three pigs were administered a series of four HIFU treatments with or without contact cooling over a period of 12 weeks. Energy fluence parameters ranged from 60 to 300 J/cm2 . Immediately after the treatment and at 1, 4, and 12 weeks, the tissue was studied by hematoxylin and eosin (H&E), Masson-trichrome, toluidine blue, CD68 staining, and transmission electron microscopy. Three human volunteers also received treatment with this HIFU device with cooling and were evaluated subjectively and objectively by computed tomography (CT).
RESULTS: HIFU treatment with a contact cooling decreased the skin surface temperature and prevented epidermal damage. Ecchymosis was observed on the non-cooled area immediately after HIFU treatment, but not on the cooled area. Histological analyses on both areas (cooled and non-cooled) revealed disrupted adipocytes in the treatment area immediately, at 1 and 4 weeks following treatment. Lipophagic histiocytic fat necrosis was evident at 4 weeks. Finally, at 12 weeks all inflammation subsided, and the lobules were markedly atrophied with reduced SAT thickness. The human volunteers experienced reduction of a few centimeter-range reduction in waist circumference after 4 weeks and pain was tolerable without bruising.
CONCLUSIONS: HIFU treatment with a cooling system efficiently destroyed adipocytes. This novel HIFU device with an added contact cooling system may provide an effective, safe and less painful treatment as a non-invasive device for fat reduction. Lasers Surg. Med. 48:878-886, 2016.
© 2016 Wiley Periodicals, Inc. © 2016 Wiley Periodicals, Inc.

Entities:  

Keywords:  contact cooling; fat reduction; high-intensity focused ultrasound (HIFU)

Mesh:

Year:  2016        PMID: 27551954     DOI: 10.1002/lsm.22576

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  3 in total

1.  Experimental evaluation of high intensity focused ultrasound for fat reduction of ex vivo porcine adipose tissue.

Authors:  Antria Filippou; Christakis Damianou
Journal:  J Ultrasound       Date:  2022-02-01

2.  Role of mechanical and thermal damage in pericapsular inflammatory response to injectable silicone in a rabbit model.

Authors:  Joon Seok; Soo Hyun Woo; Tae Rin Kwon; Jong Hwan Kim; Guk Jin Jeong; Kapsok Li; Woo Seob Kim; Beom Joon Kim
Journal:  PLoS One       Date:  2019-05-14       Impact factor: 3.240

3.  Design and evaluation of an open-source, conformable skin-cooling system for body magnetic resonance guided focused ultrasound treatments.

Authors:  Robb Merrill; Henrik Odéen; Christopher Dillon; Rachelle Bitton; Pejman Ghanouni; Allison Payne
Journal:  Int J Hyperthermia       Date:  2021       Impact factor: 3.914

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.