Massic Activity Ratios of the NBS/NIST Tritiated-Water Standards Issued Between 1954 and 1999.

L L Lucas1.   

Abstract

As part of the preparation and calibration of three new National Institute of Standards and Technology (NIST) tritiated-water radioactivity Standard Reference Materials (SRMs), the massic-activity (activity of the sample divided by the mass of the sample) ratios of all of the available NBS/NIST tritiated-water SRMs issued between 1954 and 1999 were measured using liquid-scintillation (LS) counting. Four of the tritiated-water standards (SRMs 4361, 4926B, 4927C, and 4927D) were not available for measurement. All of the other tritiated-water standards (SRMs 4361B, 4361C, 4926, 4926C, 4926D, 4926E, 4927, 4927B, 4927E, and 4927F) were available, having been stored in flame-sealed glass ampoules. Where possible, massic-activity ratios measured by liquid-scintillation counting are compared with massic-activity ratios calculated from gravimetric dilution factors. The agreement is well within the stated uncertainties. For two of the tritiated-water standards that were not available for measurement (SRMs 4361 and 4926B), massic-activity ratios calculated from gravimetric dilution factors are available.

Entities:  

Keywords:  NBS; NIST; hydrogen-3; massic activity; standards; tritiated water; tritium

Year:  2000        PMID: 27551620      PMCID: PMC4877154          DOI: 10.6028/jres.105.042

Source DB:  PubMed          Journal:  J Res Natl Inst Stand Technol        ISSN: 1044-677X


1. Introduction

The National Institute of Standards and Technology (NIST), previously the National Bureau of Standards (NBS), has issued solution standards of tritiated water since 1954. A chronology is given in Table 1. The original solutions, NBS Standard No. 4926 and NBS Standard No. 4927, were calibrated using calorimetry [1]. In 1961, the massic activity (activity of the sample divided by the mass of the sample) of NBS Standard No. 4927 was measured in the (then) new NBS length-compensated internal gas proportional counters [2] and a revised certificate was issued. The massic activities of NBS Standard No. 4927 and of the new Standard Reference Material (SRM) 4927B were measured in 1978 using the same counters [3]. The massic activities of SRM 4927E and of NBS Standard No. 4927 were measured in 1998 and 1999, respectively, in the same counters [4].
Table 1

Chronology of the NBS/NIST tritiated-water standards issued between 1954 and 1999

Year IssuedHigh-level standardMedium-level standardLow-level standardComments
1954NBS 4927NBS 4926Calibrated by calorimetry. NBS 4926 is a quantitative dilution of NBS 4927.
1961NBS 4927NBS 4926Recalibrated by gas counting. Revised certificate issued.
1965New standards are now called NBS Standard Reference Materials (SRMs).
1977SRM 4926BSRM 4926B is a quantitative dilution of NBS 4927.
1979SRM 4927BSRM 4926CCalibrated by gas counting. SRM 4926C is a quantitative dilution of SRM 4927B.
1981SRM 4361SRM 4361 is a quantitative dilution of NBS 4926.
1985SRM 4927Ca
1987SRM 4361BSRM 4361B is a quantitative dilution of SRM 4926C.
1989SRM 4927DSRM 4926DSRM 4926D is not a quantitative dilution of SRM 4927D.a
1991SRM 4927Ea
1999SRM 4927FSRM 4926ESRM 4361CCalibrated by gas counting. SRM 4926E is a quantitative dilution of SRM 4927F. SRM 4361C is a quantitative dilution of SRM 4926E.

The massic activities on the Certificates for SRMs 4927C, 4927D, 4926D, and 4927E were obtained using liquid-scintillation counting to determine the massic-activity ratios to SRM 4927B.

As part of the preparation and calibration of three new NIST tritiated-water radioactivity SRMs, the massic-activity ratios of all of the available NBS/NIST tritiated-water SRMs issued between 1954 and 1999, and of samples of the Level 1 and Level 3 tritiated-water standards produced by the National Physical Laboratory (NPL) in 1996, were measured using liquid-scintillation (LS) counting. The results for the NPL samples are reported elsewhere [4]. Four of the tritiated-water standards (SRMs 4361, 4926B, 4927C, and 4927D) were not available for measurement. All of the other tritiated-water standards (SRMs 4361B, 4361C, 4926, 4926C, 4926D, 4926E, 4927, 4927B, 4927E, and 4927F) were available, having been stored in flame-sealed glass ampoules. Where possible, massic-activity ratios measured by liquid-scintillation counting are compared with massic-activity ratios calculated from gravimetric dilution factors. The agreement is well within the stated uncertainties. For two of the tritiated-water standards that were not available for measurement (SRMS 4361 and 4926B), massic-activity ratios calculated from gravimetric dilution factors are available.

2. Liquid-Scintillation Measurements

As part of the preparation and calibration of a new series of tritiated-water SRMs (4361C, 4926E, and 4927F), five batches of liquid-scintillation samples were prepared during 1998 and 1999. The first batch compared the massic activities of the new medium-level (SRM 4926E) and the new and old low-level (4361C and 4361B) tritiated-water SRMs. The second, third, fourth, and fifth batches compared the massic activities of all of the available high-level (4927, 4927B, 4927E, and 4927F) and medium-level (4926, 4926C, 4926D, and 4926E) tritiated-water SRMs, and of several samples of the Level 1 and Level 3 tritiated-water standards produced by NPL in 1996. For most of the SRMs, several ampoules of the SRM were sampled. Several independent dilutions of each high-level SRM or standard were made so that all of the LS vials had approximately the same count rate. This minimized any uncertainty due to dead time variation. For each batch, four LS vials were prepared from each sample (ampoule or dilution) of each SRM or NPL standard. Each LS vial consisted of 10 mL of Beckman ReadySafe cocktail in a Beckman 18 mL polyethylene vial1. Each of the four LS vials was quenched with a different (but reproducible) amount of distilled water. The range of quench corresponded to a change in average pulse height of about 20 %. Approximately the same mass of tritiated water was weighed into each of the LS vials. The LS vials were then counted in one of two commercial LS counters, a Beckman LS7800 or a Packard 2500TR. Both counters use an external gamma-ray source to determine a quench indicating parameter (QIP). The counting window was set for tritium (0 keV to 18.6 keV). One counting cycle consisted of sequentially counting each LS vial in a given batch for 15 min. One run consisted of this counting cycle repeated 10 times. The data for each run were analyzed separately. The LS vials were then counted for 10 cycles in the other LS counter. This was repeated twice, so that the total count time for each LS vial was 600 minutes (2 counters times 2 runs per counter times 10 cycles per run times 15 min. per cycle) over a period of about 4 weeks. For each run, a second-order polynomial (y = massic count rate, x = QIP) was fitted to the data (typically 40 points) for each set of four LS vials using a least-squares technique. Three values of the QIP were selected (low, middle, and high) that could be used for all of the data in a given run. The massic count rate for each sample, at each of the three QIP values, was then calculated using the polynomial. For any two samples, three ratios of the massic count rate (one at each of the selected QIP values) were obtained. The ratio of massic activities of any two samples in a given run was taken to be the average of these three massic count rate ratios. The standard deviation of the mean of the three ratios was used as an indication of the uncertainty associated with the ratio. The final ratio of the massic activities of any two samples was taken to be the average of the ratios from all of the runs in which the two samples were intercompared. Because the liquid-scintillation samples had been so carefully matched with regard to other parameters, the relative standard deviation of the mean was used as the best estimate of the relative standard uncertainty associated with the final ratio.

3. Discussion of Results

Table 2 gives the massic-activity ratios for the high-level and medium-level SRMs. Table 3 gives the massic-activity ratios for the medium-level and low-level SRMs. For some of the SRMs, massic-activity ratios calculated from gravimetric dilution factors are available. These are given in Table 4 and, where possible, are compared with the massic-activity ratios measured by liquid-scintillation counting. It can be seen that the agreement is very good. This suggests that the massic activity ratios for SRM 4927 / SRM 4926B and SRM 4926 / SRM 4361, calculated from the gravimetric dilution factors, are probably also reliable to within the estimated uncertainty.
Table 2

Massic-activity ratios measured by liquid-scintillation counting for NBS/NIST tritiated-water standards issued between 1954 and 1999: high-level and medium-level standards. Each ratio is the massic activity of the SRM listed at the top of the table divided by the massic activity of the SRM listed on the left side of the table. The relative standard uncertainties of the ratios are shown as percentages. The number of sets of liquid-scintillation measurements are shown in parentheses

Ratio(Top/Side)SRM4927FSRM4927ESRM4927DSRM4927CSRM4927BSRM4927
SRM4927F1.00000
SRM4927E1.715240.052 % (7)1.00000
SRM4927D1.00000
SRM4927C1.00000
SRM4927B2.845830.064 % (7)1.658920.036 % (11)1.00000
SRM49275.753850.056 % (4)3.356030.046 % (8)2.023720.016 % (12)1.00000
SRM4926E125.959a0.041 % (7)73.45850.013 % (7)44.28980.071 % (11)21.90700.005 % (8)
SRM4926D369.5460.086 % (7)215.4500.034 % (7)129.9390.107 % (11)64.29390.020 % (8)
SRM4926C576.7580.066 % (4)336.0820.049 % (4)202.760a0.047 % (8)100.1840.055 % (8)
SRM4926Ba
SRM4926576.3980.049 % (4)336.2610.058 % (8)202.778a0.029 % (12)100.201a0.023 % (12)

The massic-activity ratio calculated from the gravimetric dilution factor is available; see Table 4.

Table 3

Massic-activity ratios measured by liquid-scintillation counting for NBS/NIST tritiated-water standards issued between 1954 and 1999: medium-level and low-level standards. Each ratio is the massic activity of the SRM listed at the top of the table divided by the massic activity of the SRM listed on the left side of the table. The relative standard uncertainties of the ratios are shown as percentages. The number of sets of liquid-scintillation measurements are shown in parentheses

Ratio(Top/Side)SRM4926ESRM4926DSRM4926CSRM4926BSRM4926SRM4361CSRM4361B
SRM4926E1.00000
SRM4926D2.933820.038 % (11)1.00000
SRM4926C4.573160.050 % (8)1.558230.074 % (8)1.00000
SRM4926B1.00000
SRM49264.573570.022 % (8)1.558370.003 % (8)1.000190.042 % (8)1.00000
SRM4361C2504.65a0.17 % (4)1.00000
SRM4361B8404.44b0.58 % (4)a3.35548b0.53 % (4)1.00000
SRM4361a

The massic-activity ratio calculated from the gravimetric dilution factor is available; see Table 4.

The uncertainty associated with this measurement is much larger than for most of the liquid-scintillation measurements because of the very low massic activity of SRM 4361B at the time of measurement.

Table 4

Massic-activity ratios measured by liquid-scintillation counting compared to massic-activity ratios calculated from gravimetric dilution factors. The relative standard uncertainties of the ratios are shown as percentages. The number of sets of liquid-scintillation measurements are shown in parentheses

Massic-activity RatioFrom the dilution factorFrom liquid-scintillationDifference
SRM 4927F / SRM 4926E125.880≈0.05 %125.9590.041 % (7)+0.06 %
SRM 4927B / SRM 4926C202.905≈0.05 %202.7600.047 % (8)−0.07 %
SRM 4927 / SRM 4926B100.236≈0.06 %
SRM 4927 / SRM 4926100.148≈0.05 %100.2010.023 % (12)+0.05 %
SRM 4926E / SRM 4361C2507.56≈0.12 %2504.650.17 % (4)−0.12 %
SRM 4926C / SRM 4361B1837.54≈0.15 %1837.78a0.58 % (4)+0.01 %
SRM 4926 / SRM 43612596.08≈0.20 %

Calculated as (SRM 4926E / SRM 4361B) / (SRM 4926E / SRM 4926C), using the massic-activity ratios given in Table 3.

Using the NIST length-compensated internal gas proportional counters, the massic activity of SRM 4927E was determined to be 370 060 Bq·g−1 as of 1200 EST, 3 September 1998, with a relative expanded uncertainty (coverage factor k = 2) of 0.70 % [4]. Using the massic activity value for SRM 4927E and the massic-activity ratios in the tables, one can calculate the massic activity of any of the other SRMs (except SRMs 4927C and 4927D). Where a different reference time is desired for the massic activity, the recommended half-life is (4500 ± 8) d, where 8 d corresponds to one standard uncertainty [5].
  2 in total

1.  Calibration of the National Institute of Standards and Technology tritiated-water standards

Authors: 
Journal:  Appl Radiat Isot       Date:  2000-03       Impact factor: 1.513

2.  Comprehensive Review and Critical Evaluation of the Half-Life of Tritium.

Authors:  L L Lucas; M P Unterweger
Journal:  J Res Natl Inst Stand Technol       Date:  2000-08-01