An assessment of the mechanical properties of leaflets from four second-generation porcine bioprostheses with biaxial testing techniques.

A S Mayne1, G W Christie, B H Smaill, P J Hunter, B G Barratt-Boyes.   

Abstract

The aim of this study was to determine whether second-generation porcine bioprostheses, glutaraldehyde fixed at pressures said to be less than 4 mm Hg, exhibit more natural leaflet material properties than earlier valves fixed at 80 to 100 mm Hg. Biaxial mechanical testing techniques were used to compare Carpentier-Edwards SAV, St. Jude Medical BioImplant, Hancock II, and Medtronic Intact bioprostheses (12 leaflets from four valves in each case) with fresh porcine aortic valves and high pressure-fixed Carpentier-Edwards 6625 bioprostheses (14 leaflets from five valves in each case). The circumferential extensibility of leaflets from Medtronic Intact bioprostheses and from fresh porcine aortic valves were not significantly different (p greater than 0.05), whereas leaflets from the other second-generation valves tested and from Carpentier-Edwards 6625 valves were highly inextensible in the circumferential direction. The radial material properties of leaflets from all bioprostheses differed from those of fresh porcine aortic valves, which were very extensible with a high pretransitional compliance. The radial extensibility and compliance of Hancock II, St. Jude Medical BioImplant, and Carpentier-Edwards 6625 leaflets were not significantly different (p greater than 0.05). In the radial direction, Carpentier-Edwards SAV and Medtronic Intact valve leaflets were substantially more extensible than Carpentier-Edwards 6625 leaflets (p less than 0.01), whereas Medtronic Intact leaflets were more compliant than all other bioprostheses. These data demonstrate (1) that second-generation porcine bioprosthetic valves do not necessarily exhibit more natural leaflet material properties than earlier high pressure-fixed xenografts and (2) that Medtronic Intact valve leaflets have material properties most closely approximating the fresh porcine aortic valve.

Entities:  

Mesh:

Year:  1989        PMID: 2755150

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  8 in total

1.  Structural mechanism for alteration of collagen gel mechanics by glutaraldehyde crosslinking.

Authors:  Preethi L Chandran; David C Paik; Jeffrey W Holmes
Journal:  Connect Tissue Res       Date:  2012-03-21       Impact factor: 3.417

Review 2.  Heart valve function: a biomechanical perspective.

Authors:  Michael S Sacks; Ajit P Yoganathan
Journal:  Philos Trans R Soc Lond B Biol Sci       Date:  2007-08-29       Impact factor: 6.237

3.  Age-related changes in material behavior of porcine mitral and aortic valves and correlation to matrix composition.

Authors:  Elizabeth H Stephens; Nicky de Jonge; Meaghan P McNeill; Christopher A Durst; K Jane Grande-Allen
Journal:  Tissue Eng Part A       Date:  2010-03       Impact factor: 3.845

Review 4.  Heart Valve Biomechanics and Underlying Mechanobiology.

Authors:  Salma Ayoub; Giovanni Ferrari; Robert C Gorman; Joseph H Gorman; Frederick J Schoen; Michael S Sacks
Journal:  Compr Physiol       Date:  2016-09-15       Impact factor: 9.090

5.  Biaxial tension of fibrous tissue: using finite element methods to address experimental challenges arising from boundary conditions and anisotropy.

Authors:  Nathan T Jacobs; Daniel H Cortes; Edward J Vresilovic; Dawn M Elliott
Journal:  J Biomech Eng       Date:  2013-02       Impact factor: 2.097

6.  Functional collagen fiber architecture of the pulmonary heart valve cusp.

Authors:  Erinn M Joyce; Jun Liao; Frederick J Schoen; John E Mayer; Michael S Sacks
Journal:  Ann Thorac Surg       Date:  2009-04       Impact factor: 4.330

Review 7.  On the biomechanics of heart valve function.

Authors:  Michael S Sacks; W David Merryman; David E Schmidt
Journal:  J Biomech       Date:  2009-06-21       Impact factor: 2.712

8.  The Mosaic Mitral Valve Bioprosthesis: A Long-Term Clinical and Hemodynamic Follow-Up.

Authors:  Michele Celiento; Stefania Blasi; Andrea De Martino; Stefano Pratali; Aldo D Milano; Uberto Bortolotti
Journal:  Tex Heart Inst J       Date:  2016-02-01
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.