| Literature DB >> 27549545 |
Jessica G Y Luc1, Miriam Shanks2, Benjamin D Tyrrell3, Robert C Welsh2, Craig R Butler2, Steven R Meyer4.
Abstract
Transcatheter aortic valve replacement (TAVR) by valve-in-valve (VIV) implantation is an alternative treatment for high-risk patients with a degenerating aortic bioprosthesis. We present a case of transapical TAVR VIV with a 29-mm Edwards SAPIEN XT (ESV) (Edwards Lifesciences, Irvine, CA) into a 29-mm Medtronic Freestyle stentless bioprosthesis (Medtronic Inc, Minneapolis, MN) in which unanticipated dilatation of the Freestyle bioprosthesis resulted in intraprocedural embolization of the TAVR valve, necessitating urgent conversion to a conventional surgical aortic valve replacement (AVR). Our experience suggests that TAVR VIV with the 29-mm ESV in the setting of a degenerated 29-mm Freestyle stentless bioprosthesis must be undertaken with caution.Entities:
Mesh:
Year: 2016 PMID: 27549545 DOI: 10.1016/j.athoracsur.2016.01.045
Source DB: PubMed Journal: Ann Thorac Surg ISSN: 0003-4975 Impact factor: 4.330