Literature DB >> 27548343

Noise Texture Deviation: A Measure for Quantifying Artifacts in Computed Tomography Images With Iterative Reconstructions.

Fabian Morsbach1, Lotus Desbiolles, Rainer Raupach, Sebastian Leschka, Bernhard Schmidt, Hatem Alkadhi.   

Abstract

OBJECTIVES: The aims of this study were to introduce the measure noise texture deviation as quantitative parameter for evaluating iterative reconstruction (IR)-specific artifacts in computed tomography (CT) images and to test whether IR-specific artifacts, quantified through this measure, are reduced in advanced modeled IR (ADMIRE) as compared with sinogram-affirmed IR (SAFIRE) images of the liver ex vivo and in patients with hypodense liver lesions.
MATERIALS AND METHODS: In the ex vivo study part, an abdominal phantom was used. In the institutional review board-approved in vivo study part, 40 consecutive patients (mean age, 63 years) with hypodense liver lesions undergoing abdominal CT in the portal-venous phase were included. Images were reconstructed with filtered back projection, with the second-generation IR algorithm SAFIRE and with the third-generation IR algorithm ADMIRE. Noise power spectra and noise texture deviation were calculated in the phantom; image noise was measured in the phantom and in patients. Two blinded readers evaluated all image data regarding IR-specific artifacts (plastic-like, blotchy appearance); patient data were evaluated regarding conspicuity and confidence for detecting hypodense liver lesions.
RESULTS: Image noise was significantly reduced at increasing IR levels (P < 0.001) with both algorithms, with no significant differences between corresponding strength levels of SAFIRE and ADMIRE (all, P > 0.05). Noise power spectra were similar at corresponding strength levels of SAFIRE and ADMIRE (all, P > 0.05). Noise texture deviation in ADMIRE was reduced compared with corresponding strength levels of SAFIRE (all, P < 0.001) and strongly correlated with subjective IR-specific artifacts (r = 0.88, P < 0.001). Iterative reconstruction-specific artifacts were significantly reduced in ADMIRE compared with that in SAFIRE images at strength levels 3 or greater, both ex vivo and in vivo (all, P < 0.001). There were no significant differences in the readers' ratings of lesion conspicuity and lesion confidence in detecting hypodense liver lesions between SAFIRE and ADMIRE (P > 0.05). Only lesion conspicuity was superior with SAFIRE and ADMIRE compared with filtered back projection (all, P < 0.001).
CONCLUSIONS: Noise texture deviation is a quantitative measure reflecting IR-specific artifacts and is reduced in CT images with ADMIRE compared with SAFIRE.

Entities:  

Mesh:

Year:  2017        PMID: 27548343     DOI: 10.1097/RLI.0000000000000312

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  8 in total

1.  Diagnostic accuracy of contemporary multidetector computed tomography (MDCT) for the detection of lumbar disc herniation.

Authors:  S Notohamiprodjo; R Stahl; M Braunagel; P M Kazmierczak; K M Thierfelder; K M Treitl; S Wirth; M Notohamiprodjo
Journal:  Eur Radiol       Date:  2016-12-17       Impact factor: 5.315

2.  Color-coded virtual non-calcium dual-energy CT for the depiction of bone marrow edema in patients with acute knee trauma: a multireader diagnostic accuracy study.

Authors:  Christian Booz; Jochen Nöske; Lukas Lenga; Simon S Martin; Ibrahim Yel; Katrin Eichler; Tatjana Gruber-Rouh; Nicole Huizinga; Moritz H Albrecht; Thomas J Vogl; Julian L Wichmann
Journal:  Eur Radiol       Date:  2019-07-26       Impact factor: 5.315

3.  Whole-body ultra-low dose CT using spectral shaping for detection of osteolytic lesion in multiple myeloma.

Authors:  Saravanabavaan Suntharalingam; Christian Mikat; Axel Wetter; Nika Guberina; Ahmed Salem; Philipp Heil; Michael Forsting; Kai Nassenstein
Journal:  Eur Radiol       Date:  2018-01-10       Impact factor: 5.315

4.  Application of 80-kVp scan and raw data-based iterative reconstruction for reduced iodine load abdominal-pelvic CT in patients at risk of contrast-induced nephropathy referred for oncological assessment: effects on radiation dose, image quality and renal function.

Authors:  Yasunori Nagayama; Shota Tanoue; Akinori Tsuji; Joji Urata; Mitsuhiro Furusawa; Seitaro Oda; Takeshi Nakaura; Daisuke Utsunomiya; Eri Yoshida; Morikatsu Yoshida; Masafumi Kidoh; Machiko Tateishi; Yasuyuki Yamashita
Journal:  Br J Radiol       Date:  2018-03-02       Impact factor: 3.039

5.  Task-Based Model Observer Assessment of A Partial Model-Based Iterative Reconstruction Algorithm in Thoracic Oncologic Multidetector CT.

Authors:  David C Rotzinger; Damien Racine; Catherine Beigelman-Aubry; Khalid M Alfudhili; Nathalie Keller; Pascal Monnin; Francis R Verdun; Fabio Becce
Journal:  Sci Rep       Date:  2018-12-07       Impact factor: 4.379

6.  Quantum Iterative Reconstruction for Low-Dose Ultra-High-Resolution Photon-Counting Detector CT of the Lung.

Authors:  Thomas Sartoretti; Damien Racine; Victor Mergen; Lisa Jungblut; Pascal Monnin; Thomas G Flohr; Katharina Martini; Thomas Frauenfelder; Hatem Alkadhi; André Euler
Journal:  Diagnostics (Basel)       Date:  2022-02-18

7.  Finding the optimal tube current and iterative reconstruction strength in liver imaging; two needles in one haystack.

Authors:  Bibi Martens; Joris G A Bosschee; Sander M J Van Kuijk; Cécile R L P N Jeukens; Maikel T H Brauer; Joachim E Wildberger; Casper Mihl
Journal:  PLoS One       Date:  2022-04-07       Impact factor: 3.240

8.  Influence of Acquisition Time on MR Image Quality Estimated with Nonparametric Measures Based on Texture Features.

Authors:  Rafał Obuchowicz; Adam Piórkowski; Andrzej Urbanik; Michał Strzelecki
Journal:  Biomed Res Int       Date:  2019-11-20       Impact factor: 3.411

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.