Man-Xia Lin1, Ming Kuang2, Ming Xu1, Bo-Wen Zhuang1, Wen-Shuo Tian1, Jie-Yi Ye1, Xiao-Hua Xie1, Xiao-Yan Xie3.
Abstract
The objective of this study was to describe the performance of ultrasound (US) and contrast-enhanced ultrasound (CEUS) within 2 h after irreversible electroporation (IRE) ablation of porcine liver. Six IRE ablations were performed on porcine liver in vivo; ultrasound assessments were performed within 2 h after IRE ablation. On US images, the ablation zone appeared as a hypo-echoic area within 10 min after the ablation, and then the echo of the ablation zone gradually increased. On CEUS images, the ablation zone appeared as a non-enhanced area within 10 min after ablation and then was gradually centripetally filled by microbubbles. A hyper-echoic rim on US images and a hyper-enhanced rim on CEUS images appeared in the periphery of the ablation zone 60 min after the ablation. Characteristic and dynamic ultrasound images of the IRE ablation zone were obtained within 2 h after IRE ablation of in vivo porcine liver.Keywords: Contrast-enhanced ultrasound; Imaging; Irreversible electroporation ablation; Ultrasound
Mesh:
Substances:
Year: 2016 PMID: 27546157 DOI: 10.1016/j.ultrasmedbio.2016.07.007
Source DB: PubMed Journal: Ultrasound Med Biol ISSN: 0301-5629 Impact factor: 2.998