Literature DB >> 27544479

Automated Measurement of Visual Acuity in Pediatric Ophthalmic Patients Using Principles of Game Design and Tablet Computers.

Tariq M Aslam1, Humza J Tahir2, Neil R A Parry3, Ian J Murray4, Kun Kwak5, Richard Heyes5, Mahani M Salleh5, Gabriela Czanner6, Jane Ashworth7.   

Abstract

PURPOSE: To report on the utility of a computer tablet-based method for automated testing of visual acuity in children based on the principles of game design. We describe the testing procedure and present repeatability as well as agreement of the score with accepted visual acuity measures.
DESIGN: Reliability and validity study.
METHODS: Setting: Manchester Royal Eye Hospital Pediatric Ophthalmology Outpatients Department. PATIENT POPULATION: Total of 112 sequentially recruited patients. INTERVENTION: For each patient 1 eye was tested with the Mobile Assessment of Vision by intERactIve Computer for Children (MAVERIC-C) system, consisting of a software application running on a computer tablet, housed in a bespoke viewing chamber. The application elicited touch screen responses using a game design to encourage compliance and automatically acquire visual acuity scores of participating patients. Acuity was then assessed by an examiner with a standard chart-based near ETDRS acuity test before the MAVERIC-C assessment was repeated. MAIN OUTCOME MEASURE: Reliability of MAVERIC-C near visual acuity score and agreement of MAVERIC-C score with near ETDRS chart for visual acuity.
RESULTS: Altogether, 106 children (95%) completed the MAVERIC-C system without assistance. The vision scores demonstrated satisfactory reliability, with test-retest VA scores having a mean difference of 0.001 (SD ±0.136) and limits of agreement of 2 SD (LOA) of ±0.267. Comparison with the near EDTRS chart showed agreement with a mean difference of -0.0879 (±0.106) with LOA of ±0.208.
CONCLUSIONS: This study demonstrates promising utility for software using a game design to enable automated testing of acuity in children with ophthalmic disease in an objective and accurate manner.
Copyright © 2016 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27544479     DOI: 10.1016/j.ajo.2016.08.013

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  5 in total

1.  A new visual acuity test on touchpad for vision screening in children.

Authors:  Elsa Di Foggia; Noémie Stoll; Hélène Meunier; Adam Rimelé; Pascal Ance; Pierre-Henri Moreau; Claude Speeg-Schatz; Arnaud Sauer
Journal:  Int J Ophthalmol       Date:  2020-09-18       Impact factor: 1.779

2.  A computerized resolution visual acuity test in preschool and school age children.

Authors:  Ying-Yan Qin; Zhen-Zhen Liu; Li-Yuan Zhu; Xuan Bao; Fu-Rong Luo; Yi-Zhi Liu; Young Tsau; Ming-Xing Wu
Journal:  Int J Ophthalmol       Date:  2020-02-18       Impact factor: 1.779

3.  Impact of Children's Postural Variation on Viewing Distance and Estimated Visual Acuity.

Authors:  Lisa M Hamm; Kishan Mistry; Joanna M Black; Cameron C Grant; Steven C Dakin
Journal:  Transl Vis Sci Technol       Date:  2019-01-30       Impact factor: 3.283

4.  Development and validation of a new method for visual acuity assesment on tablet in pediatric population: eMOVA test.

Authors:  Noémie Stoll; Elsa Di Foggia; Claude Speeg-Schatz; Hélène Meunier; Adam Rimele; Pascal Ancé; Pierre-Henri Moreau; Arnaud Sauer
Journal:  BMC Ophthalmol       Date:  2022-04-19       Impact factor: 2.086

5.  Evaluation of a New Method to Track Changes in Vision at Home for Children Undergoing Amblyopia Treatment.

Authors:  Anna O'Connor; Martha Waters; Laura England; Ashli Milling; Hazel Kay
Journal:  Br Ir Orthopt J       Date:  2021-04-09
  5 in total