Evaluation of the V8 E-Class, a Novel Automated Capillary Isoelectric Focusing Instrument for Hemoglobinopathy Screening.

Toon Schiemsky1, Caroline Brusselmans2, Elke Nackers2, Pieter Vermeersch3, Koen Poesen4, Koen J O Desmet2, Nancy Boeckx5, Davy M J Kieffer6.   

Abstract

OBJECTIVES: We evaluated the performance of a novel capillary isoelectric focusing (CIEF) application for hemoglobinopathy screening on the recently introduced V8 E-Class platform.
METHODS: Analytical performance of the V8 E-Class was evaluated and included assessment of hemoglobin A2 (HbA2) imprecision; linearity for HbA2, fetal hemoglobin (HbF), and sickle hemoglobin (HbS); and carryover for HbS. Furthermore, a method comparison with the Minicap Flex Piercing (Sebia, Lisses, France), the Variant Classic (Bio-Rad Laboratories, Hercules, CA), and the G8 (Tosoh Europe, Amsterdam, the Netherlands) was done to assess analytical and clinical concordance.
RESULTS: Total HbA2 imprecision was 3.26% and 3.14% for normal and elevated HbA2 controls and 5.16% and 3.58% for a normal and a heterozygous HbS patient sample, respectively. HbA2, HbF, and HbS showed acceptable linearity, and no carryover was observed. The method comparison showed good analytical concordance (r > 0.95) except for a homozygous HbS subset (r = 0.532-0.704). A comparable phenomenon was seen for the clinical concordance with good agreement in samples without variants (weighted κ > 0.80) but poorer agreement in HbS samples (κ < 0.30).
CONCLUSIONS: Good analytical performance was demonstrated for this novel CIEF application for hemoglobinopathy screening. Method comparison showed generally good correlation but highlights the need for standardization. Finally, software optimization could further add to its use for routine hemoglobinopathy screening. © American Society for Clinical Pathology, 2016. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.

Entities:  

Keywords:  Capillary isoelectric focusing; Capillary zone electrophoresis; Cation-exchange HPLC; Helena Biosciences V8 E-Class; Hemoglobinopathy screening

Mesh:

Substances:

Year:  2016        PMID: 27543981     DOI: 10.1093/ajcp/aqw124

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  1 in total

1.  Comparison between capillary zone electrophoresis and capillary isoelectric focusing for thalassemia screening in southern China.