Literature DB >> 27542782

The Multi-center Evaluation of the Accuracy of the Contrast MEdium INduced Pd/Pa RaTiO in Predicting FFR (MEMENTO-FFR) Study.

Antonio Maria Leone1, Roberto Martin-Reyes, Sergio B Baptista, Nicolas Amabile, Luis Raposo, Juan Antonio Franco Pelaez, Carlo Trani, Pio Cialdella, Eloisa Basile, Giuseppe Zimbardo, Francesco Burzotta, Italo Porto, Cristina Aurigemma, Antonio G Rebuzzi, Mariana Faustino, Giampaolo Niccoli, Pedro F Abreu, Michel S Slama, Vincent Spagnoli, Miren Telleria Arrieta, Ignacio J Amat Santos, Jose M de la Torre Hernandez, Ramon Lopez Palop, Filippo Crea.   

Abstract

AIMS: Adenosine administration is needed for the achievement of maximal hyperaemia fractional flow reserve (FFR) assessment. The objective was to test the accuracy of Pd/Pa ratio registered during submaximal hyperaemia induced by non-ionic contrast medium (contrast FFR [cFFR]) in predicting FFR and comparing it to the performance of resting Pd/Pa in a collaborative registry of 926 patients enrolled in 10 hospitals from four European countries (Italy, Spain, France and Portugal). METHODS AND
RESULTS: Resting Pd/Pa, cFFR and FFR were measured in 1,026 coronary stenoses functionally evaluated using commercially available pressure wires. cFFR was obtained after intracoronary injection of contrast medium, while FFR was measured after administration of adenosine. Resting Pd/Pa and cFFR were significantly higher than FFR (0.93±0.05 vs. 0.87±0.08 vs. 0.84±0.08, p<0.001). A strong correlation and a close agreement at Bland-Altman analysis between cFFR and FFR were observed (r=0.90, p<0.001 and 95% CI of disagreement: from -0.042 to 0.11). ROC curve analysis showed an excellent accuracy (89%) of the cFFR cut-off of ≤0.85 in predicting an FFR value ≤0.80 (AUC 0.95 [95% CI: 0.94-0.96]), significantly better than that observed using resting Pd/Pa (AUC: 0.90, 95% CI: 0.88-0.91; p<0.001). A cFFR/FFR hybrid approach showed a significantly lower number of lesions requiring adenosine than a resting Pd/Pa/FFR hybrid approach (22% vs. 44%, p<0.0001).
CONCLUSIONS: cFFR is accurate in predicting the functional significance of coronary stenosis. This could allow limiting the use of adenosine to obtain FFR to a minority of stenoses with considerable savings of time and costs.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27542782     DOI: 10.4244/EIJV12I6A115

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  6 in total

1.  Simplifying the assessment of coronary artery stenosis by enhancing instantaneous wave free ratio.

Authors:  Vincent Spagnoli; Fabien Picard; Victor-Xavier Tadros; Daniel Cournoyer; Serge Doucet; Jean François Tanguay; Gilbert Gosselin; Pierre de Guise; Richard Gallo
Journal:  Cardiovasc Diagn Ther       Date:  2018-04

Review 2.  The Role of Coronary Physiology in Contemporary Percutaneous Coronary Interventions.

Authors:  Federico Marin; Roberto Scarsini; Dimitrios Terentes-Printzios; Rafail A Kotronias; Flavio Ribichini; Adrian P Banning; Giovanni Luigi De Maria
Journal:  Curr Cardiol Rev       Date:  2022

Review 3.  Invasive physiological indices to determine the functional significance of coronary stenosis.

Authors:  Firas R Al-Obaidi; William F Fearon; Andy S C Yong
Journal:  Int J Cardiol Heart Vasc       Date:  2018-02-23

4.  Utility of Saline-Induced Resting Full-Cycle Ratio Compared with Resting Full-Cycle Ratio and Fractional Flow Reserve.

Authors:  Takao Sato; Sonoka Goto; Yusuke Ohta; Yuji Taya; Sho Yuasa; Minoru Takahashi; Masaaki Okabe; Yoshifusa Aizawa
Journal:  J Interv Cardiol       Date:  2020-04-06       Impact factor: 2.279

5.  Contrast medium Pd/Pa ratio in comparison to fractional flow reserve, quantitative flow ratio and instantaneous wave-free ratio for evaluation of intermediate coronary lesions.

Authors:  Paweł Kleczyński; Artur Dziewierz; Łukasz Rzeszutko; Dariusz Dudek; Jacek Legutko
Journal:  Postepy Kardiol Interwencyjnej       Date:  2020-12-29       Impact factor: 1.426

6.  Evaluation of Coronary Artery Stenosis by Quantitative Flow Ratio During Invasive Coronary Angiography: The WIFI II Study (Wire-Free Functional Imaging II).

Authors:  Jelmer Westra; Shengxian Tu; Simon Winther; Louise Nissen; Mai-Britt Vestergaard; Birgitte Krogsgaard Andersen; Emil Nielsen Holck; Camilla Fox Maule; Jane Kirk Johansen; Lene Nyhus Andreasen; Jo Krogsgaard Simonsen; Yimin Zhang; Steen Dalby Kristensen; Michael Maeng; Anne Kaltoft; Christian Juhl Terkelsen; Lars Romer Krusell; Lars Jakobsen; Johan H C Reiber; Jens Flensted Lassen; Morten Bøttcher; Hans Erik Bøtker; Evald Høj Christiansen; Niels Ramsing Holm
Journal:  Circ Cardiovasc Imaging       Date:  2018-03       Impact factor: 7.792