Literature DB >> 27541000

Long-term Results of the Cartilage Shoe Technique to Anchor a Titanium Total Ossicular Replacement Prosthesis on the Stapes Footplate After Type III Tympanoplasty.

Antoniu-Oreste Gostian1, Jean-Marc Kouamé1, Martin Bremke1, Magdalene Ortmann2, Karl-Bernd Hüttenbrink1, Dirk Beutner1.   

Abstract

Importance: Multiple techniques for a more secure and stable anchoring of the foot of a total ossicular replacement prosthesis (TORP) on the footplate have been proposed. To address this issue, a technique was developed that fits a cartilage shoe into the oval niche to guide the base of the prosthesis into a preformed central perforation. Objective: To evaluate the long-term results of the cartilage shoe technique used to anchor a titanium TORP on the stapes footplate in patients after type III tympanoplasty. Design, Setting, and Participants: Medical record review of total ossiculoplasties at a single center. The study included all patients who had undergone a tympanoplasty using a TORP between January 1, 2004, and December 31, 2008, at the Department of Otorhinolaryngology-Head and Neck Surgery, University of Cologne, Cologne, Germany. Main Outcomes and Measures: Hearing thresholds were determined by a 4-frequency (500-4000 Hz) pure-tone average air-bone gap (PTA-ABG). Intraoperative findings from revision surgery and second-look operations are reported.
Results: Forty-two ears were eligible for follow-up examination, averaging 6.8 years (range, 4.8-9.1 years) after surgery, that comprised otoscopy and audiometry. The mean age of 22 women and 20 men was 42.8 years (age range, 6-78 years). The overall PTA-ABG decreased from a mean (SD) of 33.0 (8.4) decibels (dB) before surgery to a mean (SD) of 22.0 (10.1) dB after surgery (P ≤ .001, η = 0.402). Before surgery, 64% (27 of 42) of the patients had a PTA-ABG exceeding 30 dB, which was 30 dB or less in 76% (32 of 42) of the patients after surgery. After canal wall down (n = 18) and tympanoplasties with intact canal wall (n = 24), the PTA-ABG was reduced from a mean (SD) of 33.0 (8.9) dB to 24.6 (11.2) dB (P = .01, η = 0.271) and from a mean (SD) of 32.0 (7.3) dB to 19.6 (9.2) dB (P ≤ .001, η = 0.511), respectively. For transmeatal revision ossiculoplasty (n = 33), the PTA-ABG improved from a mean (SD) of 32.0 (8.6) dB to 21.0 (10.2) dB (P ≤ .001, η = 0.389), similar to primary ossiculoplasty (n = 9), with a mean (SD) of 33.0 (5.5) dB PTA-ABG before surgery and a mean (SD) of 21.0 (9.8) dB PTA-ABG after surgery (P = .005, η = 0.478). Intraoperative findings from revision surgery (n = 11) during the study period and second-look procedures (n = 7) showed that the cartilage shoe prevented a lift off the footplate in all but 2 patients, even in the presence of recurrent cholesteatoma. Conclusions and Relevance: The use of the cartilage shoe that anchors a titanium TORP on the stapes footplate can be advocated for good and reliable long-term results after total ossiculoplasty.

Entities:  

Mesh:

Year:  2016        PMID: 27541000     DOI: 10.1001/jamaoto.2016.2118

Source DB:  PubMed          Journal:  JAMA Otolaryngol Head Neck Surg        ISSN: 2168-6181            Impact factor:   6.223


  3 in total

1.  Audiological Results of Total Ossicular Replacement Prosthesis with Cartilage Shoe Technique.

Authors:  Hacı Kaplankıran; Mehmet Emrah Ceylan; Gökçe Aksoy Yıldırım; Gözde Ceylan; Abdullah Dalğıç; Levent Olgun
Journal:  Turk Arch Otorhinolaryngol       Date:  2018-06-01

2.  Effect of a boomerang-shaped Conchal cartilage graft for Type three Tympanoplasty on Hearing.

Authors:  Snigdha Elaprolu; Arun Alexander; Sivaraman Ganesan; Coimbatore Balakrishnan Manu
Journal:  Int Arch Otorhinolaryngol       Date:  2020-06-30

3.  Study on the Prosthesis Structural Design and Vibration Characteristics Based on the Conduction Effect of Human Middle Ear.

Authors:  Wu Ren; Huijuan Yan; Yi Yu; Jinghong Ren; Jinlong Chang; Yidong Wang; Yibo Han
Journal:  Appl Bionics Biomech       Date:  2020-05-20       Impact factor: 1.781

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.