OBJECTIVES: to obtain information from scientific literature concerning infusion pumps used in administering erythrocyte (red blood cells) and to evaluate the implications in the practical use of this equipment by nurses when conducting transfusions. METHOD: an integrative revision of the following scientific databases: Pubmed/Medline, Scopus, the Virtual Library for Health, SciELO, Web of Science and Cochrane. The following descriptors were used: "infusion pumps", "blood transfusion", "transfused erythrocyte" and "hemolyis". There were no restrictions on the scope of the initial data and it was finalized in December 2014. 17 articles were identified in accordance with the inclusion and exclusion criteria. RESULTS: all of the publications included in the studies were experimental in vitro and covered the use of infusion pumps in transfusion therapy. A summary of the data was presented in a synoptic chart and an analysis of it generated the following categories: cellular damage and the infusion mechanism. CONCLUSION: infusion pumps can be harmful to erythrocytes based on the infusion mechanism that is used, as the linear peristaltic pump is more likely to cause hemolysis. Cellular damage is related to the plasmatic liberation of markers that largely dominate free hemoglobin and potassium. We reiterate the need for further research and technological investments to guide the development of protocols that promote safe practices and that can contribute to future clinical studies. OBJETIVOS: identificar na literatura a produção científica acerca dos efeitos de bomba de infusão na administração de hemácias e avaliar as implicações do uso desses equipamentos na prática transfusional de enfermagem. MÉTODO: revisão integrativa de literatura nas bases de dados Pubmed/Medline, Scopus, Biblioteca Virtual em Saúde, SciELO, Web of Science e Cochrane, utilizando os descritores "bombas de infusão", "transfusão de sangue", "transfusão de eritrócitos" e "hemólise". A data inicial não foi delimitada e a final foi dezembro de 2014. Foram identificados 17 artigos de acordo com os critérios de inclusão e exclusão. RESULTADOS: Todas as publicações incluídas eram estudos experimentais in vitro e abordavam o uso de bombas de infusão na terapêutica transfusional. A síntese dos dados foi apresentada em quadro sinóptico e a análise gerou duas categorias relevantes: lesão celular e mecanismo de infusão. CONCLUSÃO: bombas de infusão podem ser deletérias às hemácias de acordo o mecanismo de infusão, sendo que as peristálticas lineares foram mais predisponentes à hemólise. A lesão celular foi relacionada à liberação plasmática de marcadores, predominando a hemoglobina livre e potássio. Reitera-se a necessidade de pesquisas adicionais e investimentos tecnológicos para nortear o desenvolvimento de protocolos que promovam prática segura e que subsidiem futuros estudos clínicos. OBJETIVOS: identificar en la literatura la producción científica acerca de los efectos de la bomba de transfusión en la administración de glóbulos rojos y evaluar las implicaciones del uso de esos equipos en la práctica transfusional de enfermería. MÉTODO: revisión integradora de literatura en las bases de datos Pubmed/Medline, Scopus, Biblioteca Virtual en Salud, SciELO, Web of Science y Cochrane, utilizando los descriptores "bombas de infusión", "transfusión de sangre", "transfusión de eritrocitos" y "hemolisis". La fecha inicial no fue delimitada y la final fue en diciembre de 2014. Fueron identificados 17 artículos de acuerdo con los criterios de inclusión y exclusión. RESULTADOS: todas las publicaciones incluidas eran estudios experimentales in vitro y abordaban el uso de bombas de infusión en la terapéutica transfusional. La síntesis de los datos se presenta en un cuadro sinóptico y el análisis generó dos categorías relevantes: lesión celular y mecanismo de transfusión. CONCLUSIÓN: las bombas de infusión pueden ser perjudiciales para los glóbulos rojos de acuerdo el mecanismo de transfusión, siendo que las peristálticas lineares fueron más predisponentes para la hemolisis. La lesión celular fue relacionada a la liberación plasmática de marcadores, predominando la hemoglobina libre y el potasio. Se reitera la necesidad de realizar investigaciones adicionales e inversiones tecnológicas que puedan orientar el desarrollo de protocolos que promuevan la práctica segura y que ayuden a futuros estudios clínicos.
OBJECTIVES: to obtain information from scientific literature concerning infusion pumps used in administering erythrocyte (red blood cells) and to evaluate the implications in the practical use of this equipment by nurses when conducting transfusions. METHOD: an integrative revision of the following scientific databases: Pubmed/Medline, Scopus, the Virtual Library for Health, SciELO, Web of Science and Cochrane. The following descriptors were used: "infusion pumps", "blood transfusion", "transfused erythrocyte" and "hemolyis". There were no restrictions on the scope of the initial data and it was finalized in December 2014. 17 articles were identified in accordance with the inclusion and exclusion criteria. RESULTS: all of the publications included in the studies were experimental in vitro and covered the use of infusion pumps in transfusion therapy. A summary of the data was presented in a synoptic chart and an analysis of it generated the following categories: cellular damage and the infusion mechanism. CONCLUSION: infusion pumps can be harmful to erythrocytes based on the infusion mechanism that is used, as the linear peristaltic pump is more likely to cause hemolysis. Cellular damage is related to the plasmatic liberation of markers that largely dominate free hemoglobin and potassium. We reiterate the need for further research and technological investments to guide the development of protocols that promote safe practices and that can contribute to future clinical studies. OBJETIVOS: identificar na literatura a produção científica acerca dos efeitos de bomba de infusão na administração de hemácias e avaliar as implicações do uso desses equipamentos na prática transfusional de enfermagem. MÉTODO: revisão integrativa de literatura nas bases de dados Pubmed/Medline, Scopus, Biblioteca Virtual em Saúde, SciELO, Web of Science e Cochrane, utilizando os descritores "bombas de infusão", "transfusão de sangue", "transfusão de eritrócitos" e "hemólise". A data inicial não foi delimitada e a final foi dezembro de 2014. Foram identificados 17 artigos de acordo com os critérios de inclusão e exclusão. RESULTADOS: Todas as publicações incluídas eram estudos experimentais in vitro e abordavam o uso de bombas de infusão na terapêutica transfusional. A síntese dos dados foi apresentada em quadro sinóptico e a análise gerou duas categorias relevantes: lesão celular e mecanismo de infusão. CONCLUSÃO: bombas de infusão podem ser deletérias às hemácias de acordo o mecanismo de infusão, sendo que as peristálticas lineares foram mais predisponentes à hemólise. A lesão celular foi relacionada à liberação plasmática de marcadores, predominando a hemoglobina livre e potássio. Reitera-se a necessidade de pesquisas adicionais e investimentos tecnológicos para nortear o desenvolvimento de protocolos que promovam prática segura e que subsidiem futuros estudos clínicos. OBJETIVOS: identificar en la literatura la producción científica acerca de los efectos de la bomba de transfusión en la administración de glóbulos rojos y evaluar las implicaciones del uso de esos equipos en la práctica transfusional de enfermería. MÉTODO: revisión integradora de literatura en las bases de datos Pubmed/Medline, Scopus, Biblioteca Virtual en Salud, SciELO, Web of Science y Cochrane, utilizando los descriptores "bombas de infusión", "transfusión de sangre", "transfusión de eritrocitos" y "hemolisis". La fecha inicial no fue delimitada y la final fue en diciembre de 2014. Fueron identificados 17 artículos de acuerdo con los criterios de inclusión y exclusión. RESULTADOS: todas las publicaciones incluidas eran estudios experimentales in vitro y abordaban el uso de bombas de infusión en la terapéutica transfusional. La síntesis de los datos se presenta en un cuadro sinóptico y el análisis generó dos categorías relevantes: lesión celular y mecanismo de transfusión. CONCLUSIÓN: las bombas de infusión pueden ser perjudiciales para los glóbulos rojos de acuerdo el mecanismo de transfusión, siendo que las peristálticas lineares fueron más predisponentes para la hemolisis. La lesión celular fue relacionada a la liberación plasmática de marcadores, predominando la hemoglobina libre y el potasio. Se reitera la necesidad de realizar investigaciones adicionales e inversiones tecnológicas que puedan orientar el desarrollo de protocolos que promuevan la práctica segura y que ayuden a futuros estudios clínicos.
Blood transfusion is a therapeutic technology that is commonly used in clinical
practices in many different health establishments. Approximately 85 million blood
transfusions are done annually in the world, with 15 million of them being carried just
in the United States
. In Brazil in 2014 3,127,957 transfusions were carried out, in which
concentrated erythrocytes (CH) were the blood components most used, covering 57.98% of
all transfusions conducted in outpatient units and hospitals in the country
.An indication for a CH transfusion is both clinical and laboratorial which is based on
hemoglobin and hematocrit levels as well as any signs and symptoms presented by the
patient
,
.The implementation of transfusion therapy requires the use of an integrated
multidisciplinary team where the following occurs: obtaining blood donors, collection,
processing, quality control, distribution, the therapy is prescribed for someone, the
transfusion takes place and monitoring the clinical responses
-
. The role of nurses is fundamental in this process according to the Nursing
Federal Council Resolution (COFEN) nº 306/2006 which provides directives on the
procedures to be used by nurses in hemotherapy. Nurses are professionally qualified to
plan, execute, coordinate, supervise and watch patients during their transfusion
therapy
.In relation to the administration of blood components, nurses are required to use
infusion devices that are on the market that present differences concerning: control
methods for quality, price, presentation and recommendation of use. The national
practice for transfusions involves the use of blood transfusion products that work
through the use of gravity with a manual flux control system
,
.The infusion pumps (BI) are devices that regulate the flux of liquid administered under
positive pressure to the patient. They are used in intravenous therapy in different
areas of health care
-
. Currently BIs are gradually being introduced into the market to be used in
blood transfusions.With reference to flux control, the BIs can be classified as: volumetric infusion pumps,
Flow meters and syringe pumps. Volumetric equipment is a type of device that controls
the liquid to be infused in volume by unit of time through programming the flow done by
the operator. This then controls the syringe pumps. Flow meters or drips work through
permitting flows and are also controlled by an operator. However the control of the
infused volume is carried out through the counting of drips per unit of time using an
electronic sensor. Syringes pumps are instruments in which the volume administered to
the patient is stored through the use of one or more syringes which is pushed by a
movable piston controlled by equipment. The operator can select the flow rate and can
indicate volumes through units of time on the equipment. Barring the BIs syringe pumps,
the other devices essentially work through the use of peristaltic mechanisms and
cassettes
-
.The peristaltic mechanism permits the infusion of liquid through forcefully pushing a
part of it from the equipment through which the liquid passes. This can be done through
two methods: the peristaltic linear method or the rotatory method. The two methods are
different in that one works through the use of wave movement and the other works based
on compression on the linear plaques or gyrating rollers respectively. This results in
pushing the liquid from the bottle that has the solution to be infused in the
circulatory vein network in the patient
-
.The cassette mechanism involves infusion through the use of pistons. Such devices have
cassettes in side of them that are generally inserted into the center of the equipment.
The pistons, upon being actioned, move in and out of the cylinders that are contained in
the cassettes. The internal movement pushes the liquid in the direction of the patient,
whilst the external movement drains the liquid from the bottle in order to refill the
cassette. Additionally there can be diaphragm mounted on the movable pistons inserted in
the cassettes. The engine transmits a movement to the pistons that move in and out of
the bottles, compressing the siliconized, diaphragmatic membrane permitting it to be
filled or to release the liquid. When the pistons go into the cylinders, the liquid is
pushed in the direction of the patient and when the pistons move out, the liquid is
sucked up from the container that has it, allowing the cassettes bottles to be
refilled
-
.Although there are innumerable advantages in relation to the safety of the patient
through the use of BI in intravenous therapy such as the alarms that it has, the control
of the infused volume and providing the adequate time for the liquid to be administered,
there is still some uncertainty on their use in transfusion therapy due to the effects
of the infusion mechanism on the erythrocytes which may result in hemolysis
-
.Patients that are transfused with erythrocyte hemolysate, aside from receiving low
levels of functional hemoglobin, can be subject to deleterious effects to the organisms
through the presence of freed biomarkers when hemolysis occurs
-
. With consideration for the above, the guiding question for this study was: If
one transfuses erythrocytes using BI, will this result in cellular damage and
hemolysis?
Objectives
To obtain information from scientific literature concerning the effects of infusion
pumps used in administering erythrocyte (red blood cells) and to evaluate the
implications in the practical use of this equipment by nurses when conducting
transfusions.
Material and method
We undertook an integrative revision of the relevant literature. We collected data
from secondary sources based on a list that we had sourced out. We then analyzed the
data obtained in a systemized manner.The study had six steps: 1) An identification of the theme and a selection of the
theory or questions for the research, 2) establishing inclusion and exclusion
criteria for the study, as well as conducting searches in the literature, 3) defining
the information to be extracted from the research that was selected, 4)
categorization and evaluation of the included studies, 5) interpreting the results
and 6) summarizing the knowledge obtained
-
.This research method brings scientific rigor to clinical practices. This permits the
inclusion of different types of studies which allows for the following: the
maximization of research, critical evaluation and the summarization of evidence that
is obtained based on the theme
-
.In order to select the articles, we used the following portals and databases:
U.S. National Library of Medicine (PUBMED), Virtual Library for
Health (BVS), SciELO, The Cochrane Library (Cochrane),
SCOPUS and ISI Web of Science. When using the SciELO database,
the descriptors were researched using the basic (not advanced) research field.In the Science of Health (DECS) and Medical Subject Headings Section
(MESH) we used the following descriptors in response to the research question:
"infusion pumps", "blood transfusion",
"erythrocyte transfusion" and "hemolysis" . In
our searches we used the words AND and OR.Our research covered all of the articles published up until the 31 December 2014. The
inclusion criteria for the publications were the following: 1) articles published in
Portuguese, English and Spanish; 2) complete articles that covered the use of BIs in
transfusion therapy; 3) any of relevant study. The exclusion criteria was: 1)
Opinions from specialists, chapters from books, summaries of journals, patents and
editorials, 2) articles that described cellular damage that occurs when using
extracorporeal circulation devices and those which provide circulatory assistance
with oxygen for extracorporeal membranes (ECMO).The strategy used to find the articles was modified based on each the database due to
the different ways to access each of them. As a guide, we used the research question
and the inclusion criteria that had been previously defined. Figure 1 shows the strategy used for the searches done in the
Pubmed database. This strategy was used in analyzing the other databases.
Figure 1
Search Strategy in the Pubmed database- Sao Paulo, 2014
The Procedures for collecting the data
Initially we looked at the title and the summary of the studies to check that they
met the inclusion criteria. The next step was to analyze the complete article that
was selected. We needed to see whether the study was pertinent to our study. The
publications that did not have the complete text were solicited through the
comut/SCAD. Data collection forms were designed by the
researchers and were adapted to the objectives of the research, having the following
items: identification of original article, methodological characteristics, an
evaluation of the methodological rigor used, interventions and the main results
found.The results were presented in a descriptive way permitting anyone to evaluate the
applicability of our integrative revision. It would also help them in making
decisions in clinical practices related to transfusions and it would show any
knowledge gaps. This could help in developing and enhancing future research.
Procedures for analyzing the data
From the six databases included in the study, we managed to obtain a total of 566
articles. Of these, 511 fell into our inclusion criteria and our proposed objectives,
ending in 55 articles. From this total, 38 were taken out as they were repeats from
the databases, which left 17 articles (figure 2).All of the studies that were included were experimental designs in vitro. No
classification of the levels of the evidence was carried out as they were considered
to be pre-clinical studies.Further on in the research, the studies were analyzed and grouped based on their
similar content with two categories to be analyzed: the mechanism of the BI and the
cellular damage.The presentation of the results from the data that was obtained, was presented in a
summary chart which had the following information: information on the authors, the
objective of the study, type of study, results, conclusion and the implications in
the practices of transfusions done by nurses. This last element had relevant points
that were summarized from each article that we analyzed that covered the use of
infusion pumps in transfusions done by nurses. It was noted that in some places the
aforementioned devices were used in blood transfusions.
Results
Out of the 17 articles that we found, the majority were in English being: 15 (82%), 5.9%
were in Spanish and 5.9% were in Portuguese. With reference to the where the studies had
been conducted, 10 (58.9%) of them had been produced in the United States, one (5.9%) in
England and the remaining six (35.2%) were produced in the following countries: Brazil,
Spain, Australia, Denmark, Holland and Switzerland. Amongst the selected studies, the
oldest was produced in 1981 and the most recent in 2011.In relation to the design of the 17 studies selected, all of them were experimental
having simulations of the procedures in laboratories. With reference to the listed
categories corresponding to the infusion mechanism, we analyzed the peculiarities in
relation to blood transfusions and the chances of cellular damage due to the mechanical
force of the equipment. In the cellular damage category, we grouped together related
variables to the blood components and cellular damaged markers.Figure 3 is a presentation of a summary of the
studies identified and included in the present integrative revision and our principal
results.
Figure 3
Summary of the studies that were analyzed with reference to the authors,
the objective of the study, the results, type of study, results and conclusion
and implications for practices in nursing.
Infusion pump mechanisms
Amongst the 17 studies that were selected, we identified 40 types of infusion systems
used in transfusion therapy with 39 (97.5%) being BI and 01 (2.5%) being a pressure
system. Some work opted for analyzing just the infusion mechanism in isolation or the
different manufacturers while others evaluated the differences between the
mechanisms.The studies covered volumetric infusion pumps, Flow meters and syringe pumps that, in
one study, the type of devices were not described. Of the 39 that were described,
volumetric infusion pumps were described the most (32 or 82.0%). The next were 5
syringe pumps (12.8%) and lastly 2 flow meters (5.2%).Amongst the total number of pumps described, the following infusion mechanism were
found: five (12.8%) syringe pumps, nine (23.1%) cassette types, 22 (56.4%) linear
peristaltic pumps, two (5.1%) rotary pumps and one (2.6%) using the shuttle
mechanism.With reference to the infusion mechanism, the linear peristaltic pump was the type
that provoked the most amount of hemolysis. It was stated that hemolytic events could
be predicted according to 10 (76.9%) publications of the 13 that were studied
covering linear peristaltic pumps. Others studies state that the BI that is
considered the safest for therapeutic transfusions is the volumetric infusion pumps
using the cassette mechanism. This was mentioned in four studies (or 66.7% out of
six) with results that showed how safe it was.Simulation studies were done with reference to infusion flux based on clinical
practices for transfusions done in pediatrics/neonatal care and for adults. The
speeds of infusion varied from five mil liters per hour (mL/h) to 999mL/h.In relation to infusion pressure, two of the studies (11.8%) covered an inflatable
pressurized device with a pressure gauge. The other was able to obtain the maximum
pressure for linear peristaltic pumps
,
.
Cellular damage
Out of the 17 studies, the blood components which were prevalent were the CH in 10 of
them (58.8%). One (5.9%) only used the ST and six (35.3%) evaluated both blood
products in the experiments.In relation to the preservative solution, eight (47.1%) used the citrate phosphate
dextros and adenine solutions (CPDA-1). Four (23.5%) used additives based on
mannitol. Four (23.5%) did not specify the preservative solution used and one (5.9%)
analyzed samples of both solutions. None of the publications focused on the
propensity for hemolytic effects to occur based on the preservative solution. Only
two studies suggested (11.8%) that solutions with additives reduced the final
hematocrit in the bag. This in turn reduces the viscosity of the blood
components.In relation to storage time, the time period for storing hemolysis varied from 24 to
44 days. In eight studies (47.1%) there was a description of a lot of hemolysis when
the erythrocytes were near to expiration
,
,
-
,
.In the life span of the hemolysis there is a liberation of hemoglobin in the plasma.
There is also an increase in potassium and lactic dehydrogenase (LDH) amongst other
biomarkers. In the selected article, the integrity of the cells were analyzed through
variables of outcome such as free hemoglobin, hematocrit, potassium, LDH, percentage
of hemolysis, a wide distribution of red blood cells, average corpuscular volume, the
number of erythrocytes, alanine aminotransferase and aspartate aminotransferase. Free
hemoglobin was the biomarker most present in all of the publications. Potassium was
the second hemolysis marker that was most prevalent being described in 11 (64.7%) of
the studies included. Of these, nine (53.0%) related to the increase in potassium at
the longest storage period for the hemotherapeutic product.With reference to establishing hemolysis as the results of the action of the infusion
system in the cells, 10 (58,9%) articles covered this area. Four publications (23.5%)
noted significant statistical alterations in the outcome variables. However they
opted on determining the presence of hemolysis being clinically significant for
values where the level of hemolysis was above 0.8%
,
,
,
.None of the publications mentioned the clinical consequences of cellular damage and
the liberation of biomarkers to patients.
Discussion
The studies selected in the literature show that alterations in the integrity of the
erythrocytes can occur when CH and ST are transfused by BI. This is that case for the
action with the infusion mechanism and the variables related to the equipment such as
flux.All of the evidence found in the study came from simulations done in laboratories.
Humans were not used in the simulations. These studies were called pre-clinical studies.
Publications that rate the level of scientific research on a scale of I to VIII with
level I covering systematic meta-analysis revisions which are considered the best
evidence, refers to pre-clinical studies with animals. In vitro studies are considered
to be VIII on the scale
. However they are considered fundamental evidence for investigating theories
which can be subsequently evaluated and implemented in future clinical studies.Even with a low level of evidence in the in vitro studies, it is still possible to
evaluate the methodological rigor of the publications from the way how the studies have
been designed, with 16 (94.1%) covering controlled evaluations and 01 (5.9%) covering
controlled randomized studies.Cellular damage of the red blood cells during the period of the extravascular hemolysis
can have damaging clinical consequences for the patient. This is because low levels of
functioning hemoglobin are produced and renal problems can occur (such as
hemoglobinemia, hemoglobinuria and acute renal problems). There can also be alterations
in substances that point to hemolysis such as LDH, haptoglobin and potassium
-
,
.Free potassium in the plasma can bring about adverse events for the patient such as
arrhythmia and even sudden death. Other correlated studies increased the level of
potassium with an increase in storage time and the preservative solution in the
collection bag33-35). There is evidence of the occurrence of hyperkalemia and
even heart attacks when there are transfusions with CH after long storage periods
-
. The concentrations of potassium in the stored blood increase about 1
milliequivalent (mEq) per day
. However none of the publications that were selected touched on the clinical
consequences of cellular damage to patients because they mainly dealt with in vitro
studies. In spite of this, they placed a lot of emphasis on the biomarkers as a
consequence of cellular damage and hemolysis.Nowadays, national agencies, Europeans and North Americans establish a maximum level of
hemolysis at 0.8% until the last storage day. This is obligatory in the control of the
quality of blood banks
. In some included studies the authors designated hemolysis through alterations
in the markers after the experiments. Others adopt a reference value for the level of
hemolysis at 0.8% for clinical relevant hemolysis. The value designated to control the
quality of the CH in blood banks in Brazil is defined by the Resolution from the
Governing Collegiate (RDC) Number 34, 11 June 2014. It stated that the acceptable
hemolysis is a maximum of 0.8% at the last time period for storage which is about the
35th day in conservative solutions that have CPDA-1
.Another point that was explored as a factor in damaging red blood cells, was the storage
time for ST and CH. Some research noted that red blood cells at the last time period for
expiration are very fragile and are susceptible to hemolytic effects. The solutions
commonly used CPDA-1 that conserves the CH for 35 days and the additives solutions that
preserve erythrocytes for 42 days. Publications suggest that the preservative solutions
saline-adenine-glucosemannitol (SAG-M) reduces the final hematocrit for blood
components, however further research should focus on the differences between
preservative solutions, hematocrit and blood components such as red blood cells
leukoreduced, washed and radiated
,
.The linear peristaltic mechanisms were the most susceptible at producing hemolysis,
according to the articles. 13 studies were evaluated in this revision. 3 (23.1%) of them
considered it to be a safe mechanism for transfusions. The flow meters BIs mentioned in
two publications are not currently recommended for intravenous therapy because they
require electronic sensors that count the drips to measure and administer the volume of
liquid. However it does not consider the viscosity, density, superficial tension and
solution temperature which are important determinants for measuring the drips to be
administered
..The volumetric mechanism with the cassette is efficient and excellent for intravenous
therapy. This is because it has little interference with the mechanical force of the BI
on the fluid to be administered. In the articles included in the present revision, it
was identified as a safe mechanism for blood transfusions(6,17,21-22.26).An English studied noted that hemolysis is caused by multi factors associated with the
increase in hematocrit, storage time for the blood components and the pressure placed on
the red blood cells
. The manufacturers ought to pay attention to the international standard in the
International Organization for Standardization (ISO) 1135-4
, that determines the maximum pressure level for infusion at 40 kilopascal
(kPa)
,
,
. No publication stated the value of the infusion pressure as a possible factor
for cellular damage. They only stated the pressor variation in the equipment in high
infusion fluxes
. Aspects related to accessories to infusion pumps such as catheters, were
described and evaluated only in conjunction with the infusion system. This gave us
inconclusive and conflicting results.The linear peristaltic pumps are the most common pumps in the health care system in the
country. They have advantages in comparison to the cassette system, such as similar
alarms and their accessories are less onerous
. More studies in the linear peristaltic pumps should be done to established
standards for its use. The studies should also cover: infusion pressure and occlusion,
defining the worst case scenario, fluxes that mimic the practice of transfusions,
temperature control during the procedure and a wide analysis of biomarkers for
evaluation of hemolysis which has consensus in the literature.We reiterate the importance of institutional protocols for blood banks and assistant
units that ensure safety in the transfusion process to prevent untoward events. For
example there can be double checks of the blood component data and the individual who
will receive the blood. Health and safety analysis of the equipment can be conducted for
transfusions in institutions. There can also be: more improvements made in the
technology used and more visual inspections of the collection bags because this visual
check may be useful in detecting hemolysis (note this has not be proven)
.Multi professional teams need to do the following to implement the use of this
transfusion therapy: prescribe the blood product, plan the installation of the
transfusion therapy, choose the catheter and the accessories for adequate infusion,
obtain an access route, technical installation, monitoring patient responses, infusion
control (time, volume, adverse reactions), prevention of complications and constantly
monitoring the infusion. In the team, the nurse evaluates and implements the intravenous
therapy and selects the adequate materials for the patient and the treatment
. In the publications that we studied there was a greater likelihood for
hemolysis with catheters at low gauges principally where there are high flows. This is
due to the force of the blood with few lumes. Nevertheless, all of the researchers
evaluated the infusion system as a whole being part of BIs with the catheter and not
only the catheter in isolation
,
,
.The technology is a part of everyday nursing. Progress in the use of BI in intravenous
therapy has provided greater safety and efficiency in the process through the use of
resources that facilitate and improve nursing in relation to: volume control, infusion
time, memory of stunted infusions and the establishment of infusion pressure
.The theme of safety for the patient is an important tool in the management of
institutional processes and it has been receiving special attention in the world at
large. In order to promote safety it is essential that all those involved understand
clinical practices and techniques that are used to ensure low risks in transfusions.
This being the case, an evaluation of the target public's characteristics is fundamental
covering: age group, clinical state and blood component indicators. Other aspects
include: variables related to devices and equipment, accuracy, influence of the
hydrostatic pressure, solution type, the quality of the flux, system safety and evaluate
costing issues.The following should be implemented and adhered to, in order to ensure transfusion
safety: the development of protocols covering the parameters for uniformed evaluations
and improvements in future equipment through better technology. For example, incentives
should be given for the use of devices which are intelligent BIs that can alert the
professional where there is the possibility of errors or when there are alterations in
the safety standards. These devices are computerized and are thus connected to the
patient's medical records when admitted to hospital. They can be adjusted to clinical
needs and produce more efficiency in nursing. This is because notes can be directly
placed on the electronic patient record as well as noting: the dose, time, type and
volume of the medicine of infused solution
. This is technology that has been well developed and used in north American and
European countries in spite of being expensive. Future studies covering cost and
benefits for the implementation of intravenous therapy in relation to transfusion ought
to focus on improving the process and work of nurses and the safety for patients at the
edges of the hospital beds.
Conclusion
The studies selected in the literature show that alterations in the integrity of the
erythrocytes can occur when CH and ST are transfused by BI. Amongst the different types
and BI mechanisms, the safest for transfusions are volumetric infusion pumps with the
cassette mechanism. The linear peristaltic pumps are more likely to produce hemolytic
effects. With reference to variable infusion speeds, based on the analyzed equipment, we
found a divergence in the results.We noted that the storage time for the blood components to be transfused can influence
the increase in cellular fragility. In other words, the nearer the expiration deadline
for the blood product was, the greater the chances of there being hemolysis. Red blood
cell damage was mentioned in the publications when biomarkers were released with the
most common being free hemoglobin due to potassium. The level of hemolysis was a
determinant in some publications, which opted to designate the hemolysis from the
reference value of 0.8%. None of the studies focused on the clinical consequences of
extravascular hemolysis for the patient.There was no conclusive evidence on the influence of the needle gauge with the catheter
connected in the final infusion line, on the integrity of the erythrocytes for ST and
CH. We opted to evaluate the whole infusion system for some of the publications.Investment in research and technology in relation to transfusion safety for the
erythrocytes when using automated erythrocytes is extremely relevant. We need new
research showing experiments that analyze the multi factors involved in hemolysis, which
will in turn aid future studies that aim to promote safe practices for the protection of
patients.
Limitations of the Study
We opted for not limiting the scope of the data and our searches, but some of the
data was far from being relevant to the study. Although the theme was related to
technology and innovation in transfusion therapy, there was a scarcity in
publications on this topic covering the use of BIs in blood transfusions.Additional we could not evaluate the quality of the methods used in the in vitro
experimental studies as they did not fit into the system for classifying data from
epidemiological studies. They were considered to be pre-clinical. We therefore opted
to describe the rigor in the methods used.
Authors: J Tomás Rojas; F Burdallo Heras; M R Sánchez García; E Perea-Milla López; M C Medina Vilchez; J Díaz Rodríguez Journal: Rev Enferm Date: 2001-11
Authors: Meng-Chun Hsu; Mehran Mansouri; Nuzhet N N Ahamed; Stephen M Larson; Indranil M Joshi; Adeel Ahmed; David A Borkholder; Vinay V Abhyankar Journal: Sci Rep Date: 2022-06-24 Impact factor: 4.996