Abstract
This study found large discrepancies between in-bed weights recorded in the medical record and carefully obtained standing weights with a calibrated, electronic bedside scale. This discrepancy appears to be related to inadequate bed calibration before patient admission and having excessive linen, clothing, and/or equipment on the bed during weighing by caregivers.Entities: