| Literature DB >> 27514007 |
G S Payne1, L M Harris2, G S Cairns2, C Messiou2, N M deSouza2, A Macdonald2, F Saran2, M O Leach2.
Abstract
(1) H MRS measurements of lactate are often confounded by overlapping lipid signals. Double-quantum (DQ) filtering eliminates lipid signals and permits single-shot measurements, which avoid subtraction artefacts in moving tissues. This study evaluated a single-voxel-localized DQ filtering method qualitatively and quantitatively for measuring lactate concentrations in the presence of lipid, using high-grade brain tumours in which the results could be compared with standard acquisition as a reference. Paired standard acquisition and DQ-filtered (1) H MR spectra were acquired at 3T from patients receiving treatment for glioblastoma, using fLASER (localization by adiabatic selective refocusing using frequency offset corrected inversion pulses) single-voxel localization. Data were acquired from 2 × 2 × 2 cm(3) voxels, with a repetition time of 1 s and 128 averages (standard acquisition) or 256 averages (DQ-filtered acquisition), requiring 2.15 and 4.3 min respectively. Of 37 evaluated data pairs, 20 cases (54%) had measureable lactate (fitted Cramér-Rao lower bounds ≤ 20%) in either the DQ-filtered or the standard acquisition spectra. The measured DQ-filtered lactate signal was consistently downfield of lipid (1.33 ± 0.03 ppm vs 1.22 ± 0.08 ppm; p = 0.002), showing that it was not caused by lipid breakthrough, and that it matched the lactate signal seen in standard measurements (1.36 ± 0.02 ppm). In the absence of lipid, similar lactate concentrations were measured by the two methods (mean ratio DQ filtered/standard acquisition = 1.10 ± 0.21). In 7/20 cases with measurable lactate, signal was not measureable in the standard acquisition owing to lipid overlap but was quantified in the DQ-filtered acquisition. Conversely, lactate was undetected in seven DQ-filtered acquisitions but visible using the standard acquisition. In conclusion, the DQ filtering method has proven robust in eliminating lipid and permits uncontaminated measurement of lactate. This is important validation prior to use in tissues outside the brain, which contain large amounts of lipid and which are often susceptible to motion.Entities:
Keywords: MRS; brain; double-quantum filter; glioblastoma; lactate; single voxel
Mesh:
Substances:
Year: 2016 PMID: 27514007 PMCID: PMC5042032 DOI: 10.1002/nbm.3587
Source DB: PubMed Journal: NMR Biomed ISSN: 0952-3480 Impact factor: 4.044
Figure 1Examples of standard acquisition and DQ‐filtered single‐voxel fLASER spectra in glioblastoma. Voxel location is shown on the left, with standard acquisition and DQ‐filtered acquisition on the right, plotted in each case on the same vertical scale. The smooth curve is the Amares fit to the data. All spectra were acquired at 3T from a cubic voxel with 2 cm sides, 2048 complex sampling points over a 4 kHz acquisition bandwidth, and T R/T E = 1000/144 ms. 128 averages were acquired for the unfiltered spectrum, and 256 averages for the DQ‐filtered spectrum. A, Example with no apparent lipid overlying the lactate. B, Example showing significant overlap between lactate and lipid, with consequent partial signal cancellation. C, Example showing a dominant lipid peak in the standard acquisition but lactate detected in the DQ‐filtered spectrum. D, Example of a dominant lipid resonance in which no lactate was present (below detection threshold in the DQ‐filtered spectrum). Only in A can a reasonable estimate of lactate concentration be obtained from the standard acquisition spectrum.
Comparison of lactate measurements using standard acquisition or DQ‐filtered acquisition in patients in which lactate was quantifiable using both methods. The lactate frequency is referenced to water at 4.68 ppm. The concentration calculations have been corrected for T 1 and T 2 effects as described in the text
|
|
|
|
| ||
|---|---|---|---|---|---|
|
|
|
|
| ||
| 1 | 6.69 | 1.344 | 1.05 | 1.342 | 1.38 |
| 2 | 4.45 | 1.345 | 1.16 | 1.283 | 0.88 |
| 3 | 3.00 | 1.310 | 1.86 | 1.356 | 2.12 |
| 4 | 4.85 | 1.395 | 0.78 | 1.348 | 2.75 |
| 5 | 5.04 | 1.385 | 1.26 | 1.349 | 1.43 |
| 6 | 3.94 | 1.363 | 1.18 | 1.354 | 1.16 |
| 7 | 5.82 | 1.387 | 1.37 | 1.375 | 1.77 |
Data Set 4 contained high lipid (Figure 1B), and therefore yielded very low lactate in the standard acquisition.
Figure 2Comparison of lactate concentrations measured using standard acquisition and DQ‐filtered acquisition. The line of identity is indicated. The two measurements yield similar results, except for the outlier with the DQ‐filtered concentration of 2.75 mM. This corresponds to the case shown in Figure 1B, in which the lactate is strongly overlapped with lipid.