Kazuko Uchikawa Graziano1, Marta Elisa Auler Pereira2, Elaine Koda3.
Abstract
OBJECTIVE: to elaborate and apply a method to assess the efficacy of automated flexible endoscope reprocessors at a time when there is not an official method or trained laboratories to comply with the requirements described in specific standards for this type of health product in Brazil.Entities:
Mesh:
Year: 2016 PMID: 27508915 PMCID: PMC4990042 DOI: 10.1590/1518-8345.0595.2745
Source DB: PubMed Journal: Rev Lat Am Enfermagem ISSN: 0104-1169
Figure 1Test specimens fitted into the connectors of the tested automated endoscope reprocessor.
Results of the validation analysis of disinfection efficacy. São Paulo-SP, Brazil, 2014.
| Tested microorganism | Microorganism count in test specimens used as positive controls | Viable microorganism count in test specimens after reprocessing cycle 51 |
|
| 2 x106 CFU*/ 37.5 cm test specimen | Absence/ 37.5 cm test specimen |
|
| 5x106 CFU/ 37.5 cm test specimen | Absence/ 37.5 cm test specimen |
|
| 3x104 CFU/ 37.5 cm test specimen | Absence/ 37.5 cm test specimen |
|
| 5x106 CFU/ 37.5 cm test specimen | Absence/ 37.5 cm test specimen |
|
| 1x106 CFU/ 37.5 cm test specimen | Absence/ 37.5 cm test specimen |
*CFU = colony forming unit
Figure 2List of test microorganisms in RDC no. 35/2010(7) and ISO 15883-4 selected for the present methodological proposal.
Figure 3Decreasing order of resistance of microbial groups to chemical germicides according to the CDC(21)