| Literature DB >> 27500857 |
Nguyen Quoc Hung1, Huynh Dinh Chuong2, Le Quang Vuong3, Tran Thien Thanh4, Chau Van Tao1.
Abstract
In this study, in situ gamma spectra using NaI(Tl) detector have been compared with the laboratory measurements by using HPGe detector on geological samples. The results for measuring naturally occurring terrestrial gamma radiation of 40K and the decay series of 232Th and, 238U respectively of both detectors show a maximum deviation about 5%. The mass activities series from both detectors were checked for coherence using proficiency test procedure from the International Atomic Energy Agency. The reliability and precision pass for final scores for all the analytical determinations of are received "acceptable" for all radionuclides.Keywords: Acceptable; IAEA; Mass activity; Proficiency test
Mesh:
Substances:
Year: 2016 PMID: 27500857 DOI: 10.1016/j.jenvrad.2016.07.035
Source DB: PubMed Journal: J Environ Radioact ISSN: 0265-931X Impact factor: 2.674