Abstract
A method for calibrating iodine-125 seeds in terms of exposure has been established. The standard free-air ionization chamber, used for measuring soft x rays, was chosen for the measurements. Arrays of four to six seeds were used to enhance the ionization-current-to-background-current ratio. Seeds from an array were measured individually in a re-entrant chamber. The quotient of the exposure rate for the array by the sum of the ionization currents in the re-entrant chamber is the calibration factor for the re-entrant chamber. Calibration factors were established for three types of iodine-125 seeds. The overall uncertainty for the seed exposure calibrations is less than 6%.Entities:
Keywords: calibration; exposure rate; free-air chamber; iodine-125 seed; re-entrant chamber; standards
Year: 2001 PMID: 27500052 PMCID: PMC4862818 DOI: 10.6028/jres.106.042
Source DB: PubMed Journal: J Res Natl Inst Stand Technol ISSN: 1044-677X
Figure 1Cross section of gold-marker and silver-wire type iodine-125 seeds.
FAC dimensions, and corrections for measurement of iodine-125 radiation.
| Diaphragm area | 0.7867 cm2 |
| Collector plate length | 7.0003 cm |
| Air attenuation length | 12.7 cm |
| Corrections: | |
| Air attenuation | 1.018 |
| Photon scattering | 0.995 |
| Electron loss | 1.000 |
| Recombination | 1.000 |
Iodine-125 seeds used for exposure measurements.
| Type | Number of seeds | Apparent activity | Assay date |
|---|---|---|---|
| First series | |||
| Gold-marker | 4 | 60 | 1980/12/01 |
| No marker | 4 | 142 | 1982/09/06 |
| Silver-wire | 5 | 114 | 1981/06/01 |
| Second series | |||
| Gold-marker | 6 | 84 | 1983/02/21 |
| No marker | 4 | 103 | 1983/03/07 |
| Silver-wire | 6 | 95.4 | 1983/02/21 |
The apparent activity, as determined by the 3M Co., is a measure of the effective radiation output from the seeds and is independent of seed wall thickness or source self-absorption. The actual activity in the seed is greater than the apparent activity. The apparent activities are taken from the labels on the 3M lead acontainers. The 3M Co. has used the term “mCi Comp.” for apparent activity.
First series exposure rate measurements for gold-marker type seeds (µR m2/s)
| Array | FAC dist. (m) | Number of measurements | Array means | Mean for arrays |
|---|---|---|---|---|
| 1 | 0.50 | 4 | 2.281 | |
| 2 | 0.50 | 1 | 2.350 | 2.320 |
| 3 | 0.50 | 7 | 2.328 | |
| 3 | 0.25 | 5 | 2.350 | |
| 4 | 0.25 | 1 | 2.359 | 2.353 |
| 5 | 0.25 | 3 | 2.351 |
Each measurement consists of an independent series of current measurements taken at different times.
First series gold-marker type iodine-125 seed measurements in the re-entrant ionization chamber. Data corrected to STP and reference time 1981 June 1.5.
| Source number | Mean current for random seed position (pA) | Number of seed positions | Standard deviation of the mean (%) |
|---|---|---|---|
| 1 | 18.50 | 6 | 1.0 |
| 2 | 18.91 | 5 | 0.3 |
| 3 | 18.76 | 5 | 1.2 |
| 4 | 19.52 | 4 | 1.9 |
Nuclear-decay and energy-absorption data used to calculate Γ0 for iodine-125.
| Radionuclide | Radiation type | Energy | Number per transition | Energy-absorption coefficient | |
|---|---|---|---|---|---|
| Tellurium-125 | x ray K | 27.2 | 0.398 | 0.0212 | 0.230 |
| Tellurium-125 | x ray K | 27.5 | 0.742 | 0.0204 | 0.416 |
| Tellurium-125 | x ray K | 31 | 0.258 | 0.0140 | 0.112 |
| Iodine-125 | 35.5 | 0.0667 | 0.0094 | 0.022 |
First series exposure-rate measurements of Ag-wire type seeds measured at 0.25 m from the FAC (µR m2/s)
| Array number | Number of measurements | Mean for 1981 June 1.5 | Mean for all arrays | Standard deviation of the mean (%) |
|---|---|---|---|---|
| 1 | 12 | 4.639 | ||
| 2 | 5 | 4.593 | ||
| 3 | 2 | 4.699 | 4.633 | 0.8 |
| 4 | 2 | 4.560 | ||
| 5 | 2 | 4.762 | ||
| 6 | 14 | 4.543 |
First series Ag-wire type iodine-125 seed measurements in the re-entrant ionization chamber. Data corrected to STP and reference time 1981 July 23.5.
| Source number | Mean current (pA) | Standard deviation of the mean (%) |
|---|---|---|
| 11 | 115.8 | 0.1 |
| 12 | 112.4 | 0.5 |
| 13 | 110.4 | 0.4 |
| 14 | 110.3 | 0.9 |
| 15 | 111.3 | 0.7 |
First series exposure data for No-marker type seeds measured at 0.25 m from the FAC (µR m2/s).
| Array number | Means for 1982 Sept. 12.5 | Mean exposure rate ( | Standard deviation of the mean (%) |
|---|---|---|---|
| 1 | 5.040 | ||
| 2 | 5.059 | ||
| 3 | 5.060 | 5.068 | 0.2 |
| 4 | 5.070 | ||
| 5 | 5.092 | ||
| 6 | 5.086 |
First series No-marker type seed measurements in the re-entrant chamber. Data are corrected to STP and reference time 1982 Sept 12.5.
| Source number | Mean current for seven measurements (pA) | Standard deviation of the mean (%) | |
|---|---|---|---|
| 1 | 338.8 | 0.2 | |
| 2 | 316.5 | 0.3 | |
| 3 | 338.4 | 0.4 | |
| 4 | 369.9 | 0.3 | |
Second series gold-marker type iodine-125 source exposure measurements at 0.25 m from the FAC (μR m2/s).
| Type of measurement | Number of measurements | Mean exposure data 1983 Feb 1.5 | Standard deviation of the mean (%) |
|---|---|---|---|
| Single source, sum. | 4 | 4.047 | 0.9 |
| Arrays | 5 | 4.054 | 0.7 |
Re-entrant chamber current measurements for second series gold-marker iodine-125 seeds.
| Source number | Number of measurements | Mean current 1983 Feb 1.5 (pA) | Standard deviation of the mean (%) |
|---|---|---|---|
| 1 | 17 | 170.8 | 0.7 |
| 2 | 15 | 167.4 | 0.5 |
| 3 | 16 | 181.9 | 0.4 |
| 4 | 16 | 179.5 | 0.3 |
| 5 | 16 | 175.7 | 0.4 |
| 6 | 16 | 166.4 | 0.5 |
| Sum 10.42 nA | |||
Second series exposure measurements for silver-wire type iodine-125 sources at 0.25 m from the FAC (μR m2/s).
| Type of measurement | Number of measurements | Mean exposure data 1983 Feb 1.5 | Standard deviation of the mean (%) |
|---|---|---|---|
| Single source, sum. | 5 | 5.613 | 1.3 |
| arrays | 8 | 5.688 | 1.0 |
Re-entrant chamber current measurements for second series silver-wire sources.
| Source number | Mean current 1983 Feb 1.5 (pA) | Number of measurements | Standard deviation of the mean (%) |
|---|---|---|---|
| 1 | 208.6 | 6 | 0.2 |
| 2 | 219.5 | 6 | 0.5 |
| 3 | 195.0 | 6 | 0.2 |
| 4 | 228.9 | 7 | 0.7 |
| 5 | 230.4 | 11 | 0.8 |
| 6 | 198.7 | 6 | 0.9 |
Second series exposure measurements for No-marker type iodine-125 sources at 0.25 m from the FAC (μR m2/s).
| Type of measurement | Number of measurments | Mean exposure data 1983 Feb 1.5 | Standard deviation of the mean (%) |
|---|---|---|---|
| Single source, sum. | 3 | 5.490 | 0.3 |
| Arrays | 9 | 5.595 | 0.3 |
Second series re-entrant chamber current measurements for the No-marker type sources.
| Source number | Mean current | Standard deviation of the mean (%) |
|---|---|---|
| 1 | 367.4 | 0.2 |
| 2 | 388.4 | 0.2 |
| 3 | 361.8 | 0.2 |
| 4 | 381.1 | 0.3 |
Currents for nine random source positions in re-entrant tube.
Re-entrant chamber calibration factors (kR m2/C).
| Source type | First series | Second series | Mean calibration factor |
|---|---|---|---|
| Gold-marker | 3.842 | 3.872 | 3.857 |
| Silver-wire | 4.467 | 4.400 | 4.433 |
| No-marker | 3.716 | 3.701 | 3.708 |
Measurements of gold-marker iodine-125 source spectrum (1983 Feb 21.44) and energy-absorption data.
| Radiation type | ||||
|---|---|---|---|---|
| Te K | 27.4 | 2.402×106 | 0.0210 | 1.382×106 |
| Te K | 31 | 5.867×105 | 0.0140 | 2.546×105 |
| | 35.5 | 1.340×105 | 0.00935 | 4.448×104 |
Measurements of silver-wire iodine-125 source spectrum (1983 Feb 20.45) and energy-absorption data.
| Radiation type | ||||
|---|---|---|---|---|
| Te K | 27.4 | 2.836×106 | 0.0210 | 1632×106 |
| Te K | 31 | 7.028×105 | 0.0140 | 3.050×105 |
| | 35.5 | 1.962×105 | 0.00935 | 6.512×104 |
| Ag K | 22.1 | 6.59×105 | 0.0400 | 5.826×105 |
| Ag K | 25.2 | 1.61×105 | 0.0268 | 1.087×105 |
Comparison of exposure calibration data, Ẋ(FAC) and exposure data computed from measurements of spectrum, Ẋ(A).
| Source type | Exposure calibration (nR m2/s) | Exposure from spectrum (nR m2/s) | Ratio |
|---|---|---|---|
| Gold-marker | 3.271 | 3.117 | 1.049 |
| Silver-wire | 5.138 | 4.936 | 1.041 |
Figure 2Standard free-air ionization chamber measurements of exposure rate from an array of gold-marker type iodine-125 seeds. The array was rotated through 180° starting and ending with the ends of the seeds facing the FAC diaphragm. The data are normalized to measurements with the plane of the array perpendicular to the FAC axis.
Figure 3Standard free-air ionization chamber measurements of exposure rate from an array of silver-wire type iodine-125 seeds. The array was rotated through 180° starting and ending with the ends of the seeds facing the FAC diaphragm. The data are normalized to measurements with the plane of the array perpendicular to the FAC axis.
Constants for the relation Ẋ sin θ, and the normalized mean exposure rates for two seed types. The mean exposure rate is derived from data which were normalized to the exposure rate from the array in a direction perpendicular to the plane of the array.
| Source array |
| ||
|---|---|---|---|
| Silver-wire | 0.20 | 0.85 | 0.87 |
| Gold-marker | 0.09 | 0.92 | 0.81 |
Summary of standard deviations of the mean, σm, for exposure and re-entrant chamber current summations.
| First series | Second series | |||||||
|---|---|---|---|---|---|---|---|---|
|
| ||||||||
| Source type | No. of arrays | n | No. of arrays | n | ||||
| Gold-marker | 5 | 0.6 | 4 | 2.5 | 9 | 0.5 | 6 | 1.8 |
| Silver-wire | 6 | 0.8 | 5 | 1.3 | 13 | 0.8 | 6 | 1.5 |
| No-marker | 6 | 0.2 | 4 | 0.6 | 12 | 0.3 | 4 | 0.2 |
Combined and overall uncertainties for iodine-125 seed calibrations. The first and second series uncertainties are averaged to form the combined uncertainties.
| Seed type | Combined (%) | Overall (%) |
|---|---|---|
| Gold-marker | 2.6 | 5 |
| Silver-wire | 2.1 | 4 |
| No-marker | 1.5 | 3 |