Literature DB >> 27498918

Single rescuer ventilation using a bag-valve mask with internal handle: a randomized crossover trial.

Baruch Zobrist1, Monica Casmaer1, Michael D April2.   

Abstract

OBJECTIVE: To compare tidal volume received during single rescuer ventilation with a modified bag-valve mask (BVM) with integrated internal handle vs standard BVM among healthy volunteers using a manikin model.
METHODS: This study was a randomized crossover trial of adult healthcare provider volunteers performing ventilation on a manikin. We randomized participants to perform single rescuer ventilation first using either a modified BVM with integrated internal handle or a standard unmodified BVM. Participants were responsible for mask placement and delivery of 10 breaths per minute for 3 minutes as guided by a metronome. After a 3-minute rest period, they performed ventilation using the alternative device. The primary outcome measure was mean received tidal volume as measured by the manikin (IngMar RespiTrainer model). Secondary outcomes included participant reported device preference.
RESULTS: Of 70 recruited participants, all completed the study. Mean received tidal volume was higher using the modified BVM with integrated internal handle vs standard BVM by 90 mL (95% confidence interval, 60-120 mL; P< .0001). There were no significant differences in mean received tidal volume based on the order of study arm allocation. Eighty percent of participants reported preference for the modified BVM over the standard BVM (95% confidence interval, 70.0%-80.0%).
CONCLUSIONS: The modified BVM with integrated internal handle results in greater mean received tidal volume compared with standard BVM during single rescuer ventilation in a manikin model. This modified BVM design may prove a useful airway adjunct for ventilation. Published by Elsevier Inc.

Entities:  

Mesh:

Year:  2016        PMID: 27498918     DOI: 10.1016/j.ajem.2016.07.030

Source DB:  PubMed          Journal:  Am J Emerg Med        ISSN: 0735-6757            Impact factor:   2.469


  3 in total

1.  Mask Ventilation Grip: A Life-Saving Innovation.

Authors:  Jacquelin Peck; Omar Viswanath; Gerald Rosen
Journal:  Ochsner J       Date:  2018

2.  Use of a novel pedal-operated compressor is non-inferior to the use of a standard hand-compressed bag-valve mask.

Authors:  Sakina H Sojar; Zachary J Neronha; Brian Vuong; Julia R Puzone; Paul C Decerbo; Janette Baird; Robyn Wing
Journal:  J Am Coll Emerg Physicians Open       Date:  2022-02-09

3.  Efficacy of Amflow®, a Real-Time-Portable Feedback Device for Delivering Appropriate Ventilation in Critically Ill Patients: A Randomised, Controlled, Cross-Over Simulation Study.

Authors:  Jong Won Kim; Sang O Park; Kyeong Ryong Lee; Dae Young Hong; Kwang Je Baek
Journal:  Emerg Med Int       Date:  2020-04-24       Impact factor: 1.112