Early coverage of drug-eluting stents analysed by optical coherence tomography: evidence of the impact of stent apposition and strut characteristics on the neointimal healing process.

Renick Lee1, Nicolas Foin, Jaryl Ng, John Allen, Nicole Soh, Ivy Ang, Winston Shim, Ryo Torii, Philip Wong.   

Abstract

AIMS: Previous studies have associated issues such as incomplete stent apposition with delayed healing and adverse events (stent thrombosis). The aim of this study was to evaluate the impact of strut apposition and stent type on the progression of stent strut coverage. METHODS AND
RESULTS: We evaluated in vivo in porcine models the follow-up response and coverage characteristics of well-apposed and malapposed segments of drug-eluting stents (DES) (CYPHER, PROMUS Element and Orsiro) and the Absorb bioresorbable vascular scaffold (BVS) by optical coherence tomography (OCT) sequentially, at baseline, and at one week and four weeks of follow-up. Supporting results were provided by histological analysis performed at four-week follow-up and computer simulation describing the shear characteristics around apposed and non-apposed struts. A total of 325 cross-sections containing 3,166 struts were analysed. The extent of malapposition decreased over time as a result of neointimal healing (from 7.1% at baseline to 0% at four weeks; p=0.03). At one week, 13.6% of struts in well-apposed segments were still uncovered versus 19.2% of struts in malapposed cross-sections and 77.8% of NASB struts (p<0.01). At four-week follow-up, 3.1% of struts were uncovered in well-apposed cross-sections vs. 1.6% in malapposed cross-sections and 35.7% of NASB struts (p<0.01). A comparison of the apposed segments revealed that the thin-strut Orsiro had only 1.3% of uncovered struts at one week while PROMUS Element, CYPHER and BVS had 6.6%, 48.4% and 16.2% of struts still uncovered, respectively.
CONCLUSIONS: This study shows that early coverage is influenced by stent apposition as well as platform strut characteristics (stent type). At four weeks, NASB struts remained a focus of delayed endothelialisation.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27497360     DOI: 10.4244/EIJV12I5A100

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  2 in total

1.  The Proximal Optimisation Technique for Intervention of Coronary Bifurcations.

Authors:  Angela Hoye
Journal:  Interv Cardiol       Date:  2017-09

2.  Very early neointimal coverage of new biodegradable polymer drug-eluting stent compared with durable polymer everolimus-eluting stent evaluated by optical frequency domain imaging.

Authors:  Norihiro Kobayashi; Yoshiaki Ito; Masahiro Yamawaki; Motoharu Araki; Tsuyoshi Sakai; Yasunari Sakamoto; Shinsuke Mori; Masakazu Tsutsumi; Masahiro Nauchi; Yohsuke Honda; Takahiro Tokuda; Kenji Makino; Shigemitsu Shirai; Keisuke Hirano
Journal:  Int J Cardiovasc Imaging       Date:  2017-11-20       Impact factor: 2.357

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.