Literature DB >> 27496586

Comparative study of the failure rates among 3 implantable defibrillator leads.

Sophie C H van Malderen1, Tamas Szili-Torok1, Sing C Yap1, Sanne E Hoeks2, Felix Zijlstra1, Dominic A M J Theuns3.   

Abstract

BACKGROUND: After the introduction of the Biotronik Linox S/SD high-voltage lead, several cases of early failure have been observed.
OBJECTIVE: The purpose of this article was to assess the performance of the Linox S/SD lead in comparison to 2 other contemporary leads.
METHODS: We used the prospective Erasmus MC ICD registry to identify all implanted Linox S/SD (n = 408), Durata (St. Jude Medical, model 7122) (n = 340), and Endotak Reliance (Boston Scientific, models 0155, 0138, and 0158) (n = 343) leads. Lead failure was defined by low- or high-voltage impedance, failure to capture, sense or defibrillate, or the presence of nonphysiological signals not due to external interference.
RESULTS: During a median follow-up of 5.1 years, 24 Linox (5.9%), 5 Endotak (1.5%), and 5 Durata (1.5%) leads failed. At 5-year follow-up, the cumulative failure rate of Linox leads (6.4%) was higher than that of Endotak (0.4%; P < .0001) and Durata (2.0%; P = .003) leads. The incidence rate was higher in Linox leads (1.3 per 100 patient-years) than in Endotak and Durata leads (0.2 and 0.3 per 100 patient-years, respectively; P < .001). A log-log analysis of the cumulative hazard for Linox leads functioning at 3-year follow-up revealed a stable failure rate of 3% per year. The majority of failures consisted of noise (62.5%) and abnormal impedance (33.3%).
CONCLUSION: This study demonstrates a higher failure rate of Linox S/SD high-voltage leads compared to contemporary leads. Although the mechanism of lead failure is unclear, the majority presents with abnormal electrical parameters. Comprehensive monitoring of Linox S/SD high-voltage leads includes remote monitoring to facilitate early detection of lead failure.
Copyright © 2016. Published by Elsevier Inc.

Entities:  

Keywords:  Durata; Endotak; Implantable cardioverter-defibrillator; Lead failure; Linox

Mesh:

Year:  2016        PMID: 27496586     DOI: 10.1016/j.hrthm.2016.08.001

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  2 in total

1.  First time description of early lead failure of the Linox Smart lead compared to other contemporary high-voltage leads.

Authors:  Vanessa Weberndörfer; Tobias Nyffenegger; Ian Russi; Miriam Brinkert; Benjamin Berte; Stefan Toggweiler; Richard Kobza
Journal:  J Interv Card Electrophysiol       Date:  2018-05-01       Impact factor: 1.900

2.  Conductor cable externalization in an atrial hemodynamic sensor lead in a patient presenting with inappropriate shocks.

Authors:  Sing-Chien Yap; Mark G Hoogendijk; Suzanne D A Valk; Sweder W Van de Poll; Peter van der Kemp; Tamas Szili-Torok
Journal:  HeartRhythm Case Rep       Date:  2019-09-10
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.