| Literature DB >> 27491718 |
Mijin Kim1, Min Ji Jeon2, Won Gu Kim1, Jong Jin Lee3, Jin Sook Ryu3, Eun Jung Cho4, Dae Hyun Ko4, Woochang Lee4, Sail Chun4, Won Ki Min4, Tae Yong Kim1, Young Kee Shong1, Won Bae Kim1.
Abstract
BACKGROUND: Second-generation thyroglobulin immunometric assays (Tg-IMAs) have been developed with improved sensitivity. Our aim was to compare the diagnostic value of Tg-IMA measurements using a Kryptor (BRAHMS AG) kit (Tg-K) and an ACCESS (Beckman Coulter) kit (Tg-A) with that of the first-generation Tg measurement using a Tg-plus (BRAHMS AG) kit (Tg+).Entities:
Keywords: Anti-thyroglobulin; Differentiated thyroid cancer; Immunoassay; Thyroglobulin
Year: 2016 PMID: 27491718 PMCID: PMC5053060 DOI: 10.3803/EnM.2016.31.3.462
Source DB: PubMed Journal: Endocrinol Metab (Seoul) ISSN: 2093-596X
Comparison of Anti-Thyroglobulin Antibody Status of Study Subjects according to Assay Kit
| Variable | TgAb+a | Total | |
|---|---|---|---|
| Positive (>60 U/mL) | Negative (≤60 U/mL) | ||
| TgAb-Kb | |||
| Positive (>33 U/mL) | 4 | 4 | 8 |
| Negative (≤33 U/mL) | 0 | 74 | 74 |
| TgAb-Ac | |||
| Positive (>4 U/mL) | 4 | 1 | 5 |
| Negative (≤4 U/mL) | 0 | 76 | 76 |
TgAb, anti-thyroglobulin antibody.
aTgAb+, TgAb concentration measured with the BRAHMS anti-Tg RIA kit (BRAHMS AG); bTgAb-K, TgAb concentration measured with the BRAHMS anti-Tg Kryptor kit (BRAHMS AG); cTgAb-A, TgAb concentration measured with the Beckman Access Antibody II Calibrators (Beckman Coulter).
Fig. 1Concordance between (A) suppressed Tg+ and suppressed Tg-K and (B) suppressed Tg+ and suppressed Tg-A in anti-thyroglobulin antibody negative patients. Tg, thyroglobulin; Tg+, Tg level measured with the BRAHMS Tg-plus RIA kit (BRAHMS AG); Tg-K, Tg level measured with the BRAMHS hTg sensitive Kryptor kit (BRAHMS AG); Tg-A, Tg level measured with the Beckman Coulter ACCESS immunoassay kit (Beckman Coulter).
Fig. 2Concordance between stimulated Tg+ and (A) suppressed Tg+, (B) suppressed Tg-K, and (C) suppressed Tg-A as measured with each of the three different immunoassay kits in anti-thyroglobulin antibody negative patients. Tg, thyroglobulin; Tg+, Tg level measured with the BRAHMS Tg-plus RIA kit (BRAHMS AG); Tg-K, Tg level measured with the BRAMHS hTg sensitive Kryptor kit (BRAHMS AG); Tg-A, Tg level measured with the Beckman Coulter ACCESS immunoassay kit (Beckman Coulter).
Comparison of Stimulated Tg+ with Suppressed Tg+, Suppressed Tg-K, and Suppressed Tg-A in TgAb Negative Patients
| Variable | Stimulated Tg+a | Total | |
|---|---|---|---|
| ≤1 ng/mL | >1 ng/mL | ||
| Suppressed Tg+a | |||
| ≤0.3 ng/mL | 54 (87) | 8 (13) | 62 |
| >0.3 ng/mL | 0 | 16 (100) | 16 |
| Suppressed Tg-Kb | |||
| ≤0.2 ng/mL | 46 (92) | 4 (8) | 50 |
| >0.2 ng/mL | 5 (21) | 19 (79) | 24 |
| Suppressed Tg-Ac | |||
| ≤0.2 ng/mL | 42 (98) | 1 (2) | 43 |
| >0.2 ng/mL | 9 (27) | 24 (73) | 33 |
Values are expressed as number (%).
Tg, thyroglobulin; TgAb, anti-thyroglobulin antibody.
aTg+, Tg level measured with the BRAHMS Tg-plus RIA kit (BRAHMS AG); bTg-K, Tg level measured with the BRAMHS hTg sensitive Kryptor kit (BRAHMS AG); cTg-A, Tg level measured with the Beckman Coulter ACCESS immunoassay kit (Beckman Coulter).
Fig. 3Sensitivity, specificity, accuracy, and positive predictive value (PPV) and negative predictive value (NPV) of (A) suppressed Tg+, (B) suppressed Tg-K, and (C) suppressed Tg-A in predicting positivity of stimulated Tg+. Tg, thyroglobulin; Tg+, Tg level measured with the BRAHMS Tg-plus RIA kit (BRAHMS AG); Tg-K, Tg level measured with the BRAMHS hTg sensitive Kryptor kit (BRAHMS AG); Tg-A, Tg level measured with the Beckman Coulter ACCESS immunoassay kit (Beckman Coulter).