CLINICAL EVALUATION OF THE RAPID ACCESS VITREAL INJECTION GUIDE: A Handheld Instrument for Assisting Intravitreal Injections.

Dennis P Han1, Kaitlin C McKenney, Judy E Kim, David V Weinberg, David C Musch, Ravi S J Singh.   

Abstract

PURPOSE: The Rapid Access Vitreal Injection (RAVI) guide combines the function of an eyelid speculum and measuring caliper into a single instrument for assisting intravitreal injections. This study clinically evaluated the RAVI guide with respect to patient acceptance, complication rates, and operative goals.
METHODS: A prospective study was performed on 54 patients undergoing intravitreal injections using the RAVI guide (n = 32) or the speculum/caliper (n = 22). Device-related pain was assessed using the Wong-Baker scoring system, scaled from 0 (no pain) to 10 (agonizing pain).
RESULTS: Mean device-related pain score did not differ significantly between the 2 groups, with scores of 0.6 and 0.7 for the RAVI guide and speculum groups, respectively. The rate of significant pain (score of ≥2) was twice as high in the speculum group (7 of 22, 32%) compared with the RAVI guide group (5 of 32, 16%), but this difference was not statistically significant (P = 0.19, Fisher's exact test). Operative goals of avoiding needle touch to lashes/lids and guiding needle insertion to the intended site were achieved in all patients.
CONCLUSION: The RAVI guide appeared equivalent to the eyelid speculum in achieving operative goals, with similarly low pain scores. It has the potential for facilitating efficient, accurate, and safe intravitreal injections.

Entities:  

Mesh:

Year:  2017        PMID: 27491046     DOI: 10.1097/IAE.0000000000001229

Source DB:  PubMed          Journal:  Retina        ISSN: 0275-004X            Impact factor:   4.256


  2 in total

1.  Comparison of Two Different Intravitreal Injection Techniques.

Authors:  Harvey S Uy; Jose Carlo M Artiaga
Journal:  Clin Ophthalmol       Date:  2021-06-08

2.  Intravitreal Injection with a Conjunctival Injection Device: A Single-Center Experience.

Authors:  Yu Qiang Soh; Nathalie Pei Yu Chiam; Andrew Shih Hsiang Tsai; Gemmy Chui Ming Cheung; Tien Yin Wong; Ian Yew San Yeo; Edmund Yick Mun Wong; Anna Cheng Sim Tan
Journal:  Transl Vis Sci Technol       Date:  2020-07-17       Impact factor: 3.283

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.