Literature DB >> 27480288

Maneuverability and safety of a magnetic-controlled capsule endoscopy system to examine the human colon under real-time monitoring by colonoscopy: a pilot study (with video).

Hongxiang Gu1, Huiling Zheng1, Xiaobing Cui1, Ying Huang1, Bo Jiang1.   

Abstract

BACKGROUND AND AIMS: Colon capsule endoscopy has become an alternative method to evaluate the colonic mucosa without pain, sedation, and gas insufflation in recent years. The magnetic-controlled capsule endoscopy (MCCE) system is an innovative ingestible colon capsule. We performed a pilot study to assess its maneuverability and safety among subjects who volunteered to undergo colorectal cancer screening.
METHODS: Fifty-seven volunteers underwent both MCCE and colonoscopy procedures. The position of the MCCE was monitored after the capsule was swallowed. When the MCCE reached the cecum, it was controlled by a magnetic manipulator to observe the colonic mucosa under real-time monitoring by colonoscopy. The primary measurements included maneuverability, the level of cleanliness, lesions, and safety.
RESULTS: Five volunteers (8.78%) were excluded because of bowel preparation protocol deviations or failure to reach the cecum before the battery was exhausted. There was no capsule retention. Maneuverability of the MCCE to match the guidance of the magnetic manipulator was graded as good in 49 subjects (94.23%) and moderate in 3 (5.77%). It took 3.63 ± 1.14 hours for the MCCE to reach the cecum. In 52 subjects (100%) the MCCE reached the transverse colon positively, and in 41 subjects (78.84%) the MCCE reached the rectosigmoid colon within a limited time of 25 minutes. The bowel preparation for MCCE was rated as good or excellent in 84.61% of the volunteers. Six positive findings were identified by the MCCE in the colon, which were also confirmed by colonoscopy.
CONCLUSIONS: The MCCE showed promising maneuverability under real-time monitoring by colonoscopy. (Clinical trial registration number: NCT02536144.).
Copyright © 2017 American Society for Gastrointestinal Endoscopy. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27480288     DOI: 10.1016/j.gie.2016.07.053

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  7 in total

Review 1.  Colon capsule endoscopy: toward the future.

Authors:  Naoki Muguruma; Kumiko Tanaka; Satoshi Teramae; Tetsuji Takayama
Journal:  Clin J Gastroenterol       Date:  2017-01-13

Review 2.  A global perspective on capsule endoscopy.

Authors:  Tanja Nowak
Journal:  Ann Transl Med       Date:  2017-11

Review 3.  Review: capsule colonoscopy-a concise clinical overview of current status.

Authors:  Diana E Yung; Emanuele Rondonotti; Anastasios Koulaouzidis
Journal:  Ann Transl Med       Date:  2016-10

4.  First prospective European study for the feasibility and safety of magnetically controlled capsule endoscopy in gastric mucosal abnormalities.

Authors:  Milán Szalai; Krisztina Helle; Barbara Dorottya Lovász; Ádám Finta; András Rosztóczy; László Oczella; László Madácsy
Journal:  World J Gastroenterol       Date:  2022-05-28       Impact factor: 5.374

5.  Standing-type magnetically guided capsule endoscopy versus gastroscopy for gastric examination: multicenter blinded comparative trial.

Authors:  Hua-Sheng Lai; Xin-Ke Wang; Jian-Qun Cai; Xin-Mei Zhao; Ze-Long Han; Jie Zhang; Zhen-Yu Chen; Zhi-Zhao Lin; Ping-Hong Zhou; Bing Hu; Ai-Min Li; Si-de Liu
Journal:  Dig Endosc       Date:  2019-10-10       Impact factor: 7.559

6.  Automatic lumen detection and magnetic alignment control for magnetic-assisted capsule colonoscope system optimization.

Authors:  Sheng-Yang Yen; Hao-En Huang; Gi-Shih Lien; Chih-Wen Liu; Chia-Feng Chu; Wei-Ming Huang; Fat-Moon Suk
Journal:  Sci Rep       Date:  2021-03-19       Impact factor: 4.379

7.  Application of robotic technologies in lower gastrointestinal tract endoscopy: A systematic review.

Authors:  Harpreet Kaur Sekhon Inderjit Singh; Emily Rose Armstrong; Sujay Shah; Reza Mirnezami
Journal:  World J Gastrointest Endosc       Date:  2021-12-16
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.