| Literature DB >> 27478855 |
Robert J Kubiak1, Nancy Lee1, Yuan Zhu1, William R Franch2, Sophia V Levitskaya2, Surekha R Krishnan1, Varghese Abraham1, Peter F Akufongwe1, Christopher J Larkin2, Wendy I White1.
Abstract
Consistent performance of anti-drug antibody (ADA) assays through all stages of clinical development is critical for the assessment of immunogenicity and interpretation of PK, PD, safety, and efficacy. The electrochemiluminescent assays commonly employed for ADA measurement use drug conjugated with ruthenium and biotin to bind ADA in samples. Here we report an association between high nonspecific ADA responses in certain drug-naïve individuals and the storage buffer of the conjugated reagents used in a monoclonal antibody ADA assay. Ruthenylated reagents stored in phosphate-buffered saline (PBS) buffer had increased levels of aggregate and produced variable and high baseline responses in some subjects. Reagents stored in a histidine-sucrose buffer (HSB) had lower aggregate levels and produced low sample responses. In contrast to PBS, conjugated reagents formulated in HSB remained low in aggregate content and in sample response variability after 5 freeze/thaw cycles. A reagent monitoring control (RMC) serum was prepared for the real-time evaluation of conjugated reagent quality. Using appropriate buffers for storage of conjugated reagents together with RMCs capable of monitoring of reagent aggregation status can help ensure consistent, long-term performance of ADA methods.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27478855 PMCID: PMC4958433 DOI: 10.1155/2016/1485615
Source DB: PubMed Journal: J Immunol Res ISSN: 2314-7156 Impact factor: 4.818
Figure 1HPSEC traces of ruthenylated antibody conjugate stored at 1 mg/mL in PBS or in HSB. Each preparation was subjected to multiple freeze-thaw cycles by cycling between −80°C (approximately 16 hours) and 5°C (approximately 8 hours) prior to analysis by HPSEC: the light blue color: 0 F/T cycles; the red color: 1 F/T cycle; the green color: 3 F/T cycles; and the purple color: 5 F/T cycles. The retention times, structural designations, and peak areas are shown in Table 1.
Summary of ruthenylated antibody analysis by HPSEC. Ruthenylated antibody was stored at 1 mg/mL in PBS) or in HSB. Each preparation was subjected to multiple freeze-thaw (F/T) cycles by cycling between −80 ± 10°C (approximately 16 hours) and 5 ± 3°C (approximately 8 hours) prior to analysis by HPSEC.
| Retention time [min] | Molecular weight [kDa] | Structure designation | % peak area | |||||||
|---|---|---|---|---|---|---|---|---|---|---|
| 0 F/T cycles | 1 F/T cycle | 3 F/T cycles | 5 F/T cycles | |||||||
| PBS | HSB | PBS | HSB | PBS | HSB | PBS | HSB | |||
| 11.45 | ~1900 | Unordered multimer | 0.09 | 0.0 | 0.22 | 0.0 | 0.44 | 0.0 | 0.31 | 0.0 |
| 12.45 | ~360 | Dimers, trimers | 2.6 | 0.2 | 2.9 | 0.2 | 3.6 | 0.2 | 2.8 | 0.2 |
| 14.89 | ~150 | Monomer | 97.3 | 99.7 | 96.9 | 99.7 | 95.9 | 99.6 | 96.9 | 99.7 |
Figure 2Screening responses of individual disease A and disease B serum samples tested using ruthenylated antibody reagents stored in PBS and in HSB. The solid red line on each graph indicates screening cut point values for disease A (1.37) and disease B (1.14) populations obtained during method validation.
Figure 3ECL signal in relative light units (RLU) of the system suitability controls: negative control (NQC: pooled serum) and low and high positive controls (LQC: 15 ng/mL, HQC: 1000 ng/mL of affinity-purified anti-idiotype goat polyclonal antibody in NQC) analyzed with reagents stored in PBS and HSB subjected to multiple freeze-thaw (F/T) cycles. Error bars show mean with standard deviation.
Figure 4Preparation of the reagent monitoring control. (a) 160 commercial serum samples from drug-naïve subjects diagnosed with disease B were screened using conjugated reagents stored in PBS and subjected to one freeze/thaw cycle. Solid green circles indicate samples with S : B responses at or above the screening cut point. (b) Samples selected in step (a) were tested in the screening assay using conjugated reagents stored in PBS and in HSB. Solid green circles indicate samples that changed classification from negative when tested using reagents in HSB to positive using reagents in PBS. (c) Samples identified in step (b) were pooled together and measured three times using conjugated reagents stored in HSB and using reagents in PBS. Green circles in this panel represent the three separate measurements of the final RMC pool. Red line indicates assay cut point for disease B.
Figure 5Screening and confirmatory responses of individual drug-naïve disease A and disease B samples tested using conjugated reagents stored in HSB or in PBS. The vertical red and horizontal green lines on each graph show screening (SCP) and confirmatory cut points (CCP), respectively. Open circles (○) indicate negative responses and closed circles (●) correspond to confirmed positives.