Production of Spherical Ablations Using Nonthermal Irreversible Electroporation: A Laboratory Investigation Using a Single Electrode and Grounding Pad.

Michael B Sano1, Richard E Fan2, Gloria L Hwang3, Geoffrey A Sonn2, Lei Xing4.   

Abstract

PURPOSE: To mathematically model and test ex vivo a modified technique of irreversible electroporation (IRE) to produce large spherical ablations by using a single probe.
MATERIALS AND METHODS: Computed simulations were performed by using varying voltages, electrode exposure lengths, and tissue types. A vegetable (potato) tissue model was then used to compare ablations created by conventional and high-frequency IRE protocols by using 2 probe configurations: a single probe with two collinear electrodes (2EP) or a single electrode configured with a grounding pad (P+GP). The new P+GP electrode configuration was evaluated in ex vivo liver tissue.
RESULTS: The P+GP configuration produced more spherical ablation volumes than the 2EP configuration in computed simulations and tissue models. In prostate tissue, computed simulations predicted ablation volumes at 3,000 V of 1.6 cm(3) for the P+GP configurations, compared with 0.94 cm(3) for the 2EP configuration; in liver tissue, the predicted ablation volumes were 4.7 times larger than those in the prostate. Vegetable model studies verify that the P+GP configuration produces larger and more spherical ablations than those produced by the 2EP. High-frequency IRE treatment of ex vivo liver with the P+GP configuration created a 2.84 × 2.21-cm ablation zone.
CONCLUSIONS: Computer modeling showed that P+GP configuration for IRE procedures yields ablations that are larger than the 2EP configuration, creating substantial ablation zones with a single electrode placement. When tested in tissue models and an ex vivo liver model, the P+GP configuration created ablation zones that appear to be of clinically relevant size and shape.
Copyright © 2016 SIR. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27478129     DOI: 10.1016/j.jvir.2016.05.032

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  1 in total

1.  Asymmetric Waveforms Decrease Lethal Thresholds in High Frequency Irreversible Electroporation Therapies.

Authors:  Michael B Sano; Richard E Fan; Lei Xing
Journal:  Sci Rep       Date:  2017-01-20       Impact factor: 4.379

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.