Christopher G Chapman1, Uzma D Siddiqui1, Maricarmen Manzano2, Vani J Konda1, Cynthia Murillo3, Emily M Landon3, Irving Waxman1. 1. Center for Endoscopic Research and Therapeutics, The University of Chicago Medicine and Biological Sciences, Chicago, Illinois, USA. 2. National Cancer Institute of Mexico, Mexico City, Mexico. 3. Department of Medicine, Section of Infectious Diseases & Global Health, The University of Chicago Medicine and Biological Sciences, Chicago, Illinois, USA.
Abstract
BACKGROUND AND AIMS: The complex design of the elevator mechanism in duodenoscopes has been recognized as a challenge for disinfection and recently implicated as a potential source of persistent bacterial contamination. Curvilinear array (CLA) echoendoscopes also have an elevator mechanism; however, there are no recommendations or data regarding the risk of persistent bacterial contamination of echoendoscopes. Here we hoped to determine the yield of microbial growth with routine bacterial surveillance cultures of reprocessed CLA echoendoscopes. METHODS: Beginning in February 2015 to February 2016, CLA echoendoscopes at a single tertiary care center underwent prospective bacterial surveillance cultures after reprocessing. Any growth of gram-negative bacilli was considered to be critical. Echoendoscopes with a positive result underwent quarantine followed by repeat disinfection and culture. RESULTS: During the study period, 540 cultures were obtained; 521 (96.5%) were primary cultures obtained from 18 CLA echoendoscopes. Twenty-two primary cultures (4.2%) were positive for gram-negative bacilli after high-level disinfection reprocessing. Eleven different bacteria were isolated: Klebsiella pneumoniae, Citrobacter freundii, Escherichia coli, Pseudomonas aeruginosa, Klebsiella oxytoca, Sphingomonas paucimobilis, Acinetobacter baumanii, Enterobacter cloacae, Hafnia alvei, Pseudomonas putida, and Stenotrophomonas maltophilia. Antibiotic sensitivity data on 19 of 24 bacteria (79.2%) isolated from positive primary cultures revealed no documented cases of carbapenem-resistant enterobacteriaceae, cephalosporin-resistant-Klebsiella, or multidrug-resistant Acinetobacter. There have been no documented cases of patient-to-patient transmission. CONCLUSIONS: After following standard high-level disinfection and reprocessing, CLA echoendoscopes can remain culture positive for high-concern organisms. Recommendations regarding infection risk should take into consideration elevator-containing echoendoscopes in addition to duodenoscopes to ensure patient safety and endoscope reprocessing efficacy.
BACKGROUND AND AIMS: The complex design of the elevator mechanism in duodenoscopes has been recognized as a challenge for disinfection and recently implicated as a potential source of persistent bacterial contamination. Curvilinear array (CLA) echoendoscopes also have an elevator mechanism; however, there are no recommendations or data regarding the risk of persistent bacterial contamination of echoendoscopes. Here we hoped to determine the yield of microbial growth with routine bacterial surveillance cultures of reprocessed CLA echoendoscopes. METHODS: Beginning in February 2015 to February 2016, CLA echoendoscopes at a single tertiary care center underwent prospective bacterial surveillance cultures after reprocessing. Any growth of gram-negative bacilli was considered to be critical. Echoendoscopes with a positive result underwent quarantine followed by repeat disinfection and culture. RESULTS: During the study period, 540 cultures were obtained; 521 (96.5%) were primary cultures obtained from 18 CLA echoendoscopes. Twenty-two primary cultures (4.2%) were positive for gram-negative bacilli after high-level disinfection reprocessing. Eleven different bacteria were isolated: Klebsiella pneumoniae, Citrobacter freundii, Escherichia coli, Pseudomonas aeruginosa, Klebsiella oxytoca, Sphingomonas paucimobilis, Acinetobacter baumanii, Enterobacter cloacae, Hafnia alvei, Pseudomonas putida, and Stenotrophomonas maltophilia. Antibiotic sensitivity data on 19 of 24 bacteria (79.2%) isolated from positive primary cultures revealed no documented cases of carbapenem-resistant enterobacteriaceae, cephalosporin-resistant-Klebsiella, or multidrug-resistant Acinetobacter. There have been no documented cases of patient-to-patient transmission. CONCLUSIONS: After following standard high-level disinfection and reprocessing, CLA echoendoscopes can remain culture positive for high-concern organisms. Recommendations regarding infection risk should take into consideration elevator-containing echoendoscopes in addition to duodenoscopes to ensure patient safety and endoscope reprocessing efficacy.
Authors: Gheorghe G Balan; Catalin Victor Sfarti; Stefan Andrei Chiriac; Carol Stanciu; Anca Trifan Journal: Eur J Clin Microbiol Infect Dis Date: 2019-09-03 Impact factor: 3.267
Authors: Hemant Goyal; Sara Larsen; Abhilash Perisetti; Nikolaj Birk Larsen; Lotte Klinten Ockert; Sven Adamsen; Benjamin Tharian; Nirav Thosani Journal: Endosc Int Open Date: 2022-06-10
Authors: Cori L Ofstead; Brandy L Buro; Krystina M Hopkins; John E Eiland; Harry P Wetzler; David R Lichtenstein Journal: Endosc Int Open Date: 2020-11-17
Authors: Judith A Kwakman; Arjan W Rauwers; Jolanda G Buijs; Woutrinus de Groot; Margreet C Vos; Marco J Bruno Journal: Endosc Int Open Date: 2022-09-14
Authors: Maarten Heuvelmans; Herman F Wunderink; Henny C van der Mei; Jan F Monkelbaan Journal: Antimicrob Resist Infect Control Date: 2021-12-23 Impact factor: 4.887