Literature DB >> 27455874

Sacroplasty procedural extravasation with high viscosity bone cement: comparing the intraoperative long-axis versus short-axis techniques in osteoporotic cadavers.

Jeffrey W Miller1, Art Diani2, Steve Docsa2, Kristi Ashton2, Michele Sciamanna3.   

Abstract

INTRODUCTION: Percutaneous sacroplasty involves image-guided injection of bone cement for sacral insufficiency fractures to alleviate pain and facilitate mobility. Correct sacral placement of the cement and the risk of cement extravasation present procedural challenges. This study compares the occurrence, number, location, and surface area of high viscosity radiopaque bone cement extravasation via biplane fluoroscopy with Dyna CT between the fluoroscopically-guided intraoperative long-axis and short-axis sacroplasty techniques in osteoporotic cadavers.
METHODS: Ten osteoporotic cadavers underwent bilateral percutaneous instillation of VertaPlex HV High Viscosity Radiopaque Bone Cement. Long- and short-axis sacroplasty techniques were randomly assigned to zone 1 of the left or right sacral ala of each cadaver. Cement extravasation data were summarized by technique (long-axis vs short-axis) and time period (15-min and 3-hour post-procedure syngo DynaCT scan) in the form of point and CI estimates for the true proportions of cement extravasation.
RESULTS: No procedural sacral extravasation differences were observed between the long-axis and short-axis sacroplasty techniques. There were no occurrences of intra-procedural or post-procedural cement extravasation at 15 min or 3 hours in association with either the long-axis sacroplasty technique or the short-axis sacroplasty technique.
CONCLUSIONS: The long- and short-axis sacroplasty techniques, using high viscosity cement with careful post-procedural positioning, result in no occurrence of cement extravasation in porous osteoporotic cadaver bone. Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/.

Entities:  

Keywords:  CT; Device; Technique

Mesh:

Substances:

Year:  2016        PMID: 27455874     DOI: 10.1136/neurintsurg-2016-012442

Source DB:  PubMed          Journal:  J Neurointerv Surg        ISSN: 1759-8478            Impact factor:   5.836


  2 in total

Review 1.  Percutaneous sacroplasty for the management of painful pathologic fracture in a multiple myeloma patient: Case report and review of the literature.

Authors:  A A Dmytriw; K Talla; R Smith
Journal:  Neuroradiol J       Date:  2016-11-25

2.  Percutaneous Sacroplasty for Painful Sacral Metastases Involving Multiple Sacral Vertebral Bodies: Initial Experience with an Interpedicular Approach.

Authors:  Qing Hua Tian; He Fei Liu; Tao Wang; Ying Sheng Cheng; Chun Gen Wu
Journal:  Korean J Radiol       Date:  2019-06       Impact factor: 3.500