Pauline Levigne1, François Peyron2, Martine Wallon1.
Abstract
When acquired during pregnancy toxoplasmosis can have devastating consequences on the fetus. As maternal infection is in the majority of cases subclinical, the diagnosis of toxoplasmosis relies on serological tests for the detection of IgG, IGM and the mesure of IgG avidity. We evaluated the performance of IDS-iSYS a new automatized instrument based on chemiluminescence for the diagnosis of the disease. Our study was based on non-selected samples received in our laboratory either for the determination of serological status or for distinguishing acute from chronic infection. Seven hundred eighty three samples were enrolled in the study. Compared with Architect IgG and IgM assays, the sensitivity and specificity were respectively 99% and 99% for IgG, and 75% and 97% for IgM. We observed higher remaining titers for IDS iSYS IgG, which could reduce the proportions of patients who have to be retested because of doubtful titers. IgM detection and avidity scored equivalent performance with both methods. This automate appears to be a reliable and easy-to-use tool for diagnosing toxoplasmosis in different clinical settings.Entities:
Keywords: Architect; IDS-iSYS; Serologic diagnosis; Toxoplasmosis
Mesh:
Substances:
Year: 2016 PMID: 27450223 DOI: 10.1016/j.diagmicrobio.2016.07.006
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803